Task-based quantification of image quality using a model observer in abdominal CT: a multicentre study

被引:19
|
作者
Racine, Damien [1 ]
Ryckx, Nick [1 ]
Ba, Alexandre [1 ]
Becce, Fabio [2 ]
Viry, Anais [1 ]
Verdun, Francis R. [1 ]
Schmidt, Sabine [2 ]
机构
[1] Lausanne Univ Hosp, Inst Radiat Phys, Rue Grand Pre 1, CH-1007 Lausanne, Switzerland
[2] Lausanne Univ Hosp, Dept Diagnost & Intervent Radiol, Rue Bugnon 46, CH-1011 Lausanne, Switzerland
关键词
Abdominal computed tomography; Image quality; Model observer; Standardisation; Task-based assessment; STATISTICAL ITERATIVE RECONSTRUCTION; LOW-CONTRAST DETECTABILITY; FILTERED BACK-PROJECTION; LOW-TUBE-VOLTAGE; NOISE; OPTIMIZATION; PERFORMANCE; RESOLUTION; ALGORITHM; PHANTOM;
D O I
10.1007/s00330-018-5518-8
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
ObjectiveWe investigated the variability in diagnostic information inherent in computed tomography (CT) images acquired at 68 different CT units, with the selected acquisition protocols aiming to answer the same clinical question.MethodsAn anthropomorphic abdominal phantom with two optional rings was scanned on 68 CT systems from 62 centres using the local clinical acquisition parameters of the portal venous phase for the detection of focal liver lesions. Low-contrast detectability (LCD) was assessed objectively with channelised Hotelling observer (CHO) using the receiver operating characteristic (ROC) paradigm. For each lesion size, the area under the ROC curve (AUC) was calculated and considered as a figure of merit. The volume computed tomography dose index (CTDIvol) was used to indicate radiation dose exposure.ResultsThe median CTDIvol used was 5.8 mGy, 10.5 mGy and 16.3 mGy for the small, medium and large phantoms, respectively. The median AUC obtained from clinical CT protocols was 0.96, 0.90 and 0.83 for the small, medium and large phantoms, respectively.ConclusionsOur study used a model observer to highlight the difference in image quality levels when dealing with the same clinical question. This difference was important and increased with growing phantom size, which generated large variations in patient exposure. In the end, a standardisation initiative may be launched to ensure comparable diagnostic information for well-defined clinical questions. The image quality requirements, related to the clinical question to be answered, should be the starting point of patient dose optimisation.Key Points center dot Model observers enable to assess image quality objectively based on clinical tasks.center dot Objective image quality assessment should always include several patient sizes.center dot Clinical diagnostic image quality should be the starting point for patient dose optimisation.center dot Dose optimisation by applying DRLs only is insufficient for ensuring clinical requirements.
引用
收藏
页码:5203 / 5210
页数:8
相关论文
共 50 条
  • [1] Task-based quantification of image quality using a model observer in abdominal CT: a multicentre study
    Damien Racine
    Nick Ryckx
    Alexandre Ba
    Fabio Becce
    Anais Viry
    Francis R. Verdun
    Sabine Schmidt
    European Radiology, 2018, 28 : 5203 - 5210
  • [2] A Computer Simulation Model Observer for Task-Based Image Quality Assessment in Radiation Therapy
    Dolly, S.
    Yu, L.
    Mutic, S.
    Anastasio, M.
    Li, H.
    MEDICAL PHYSICS, 2016, 43 (06) : 3449 - 3449
  • [3] Automated exposure control for CT using a task-based image quality metric
    Khobragade, P.
    Fan, Jiahua
    Rupcich, Franco
    Crotty, Dominic J.
    Schmidt, Taly Gilat
    MEDICAL IMAGING 2018: PHYSICS OF MEDICAL IMAGING, 2018, 10573
  • [4] Assessment of task-based image quality for abdominal CT protocols linked with national diagnostic reference levels
    Anaïs Viry
    Christoph Aberle
    Thiago Lima
    Reto Treier
    Sebastian T. Schindera
    Francis R. Verdun
    Damien Racine
    European Radiology, 2022, 32 : 1227 - 1237
  • [5] Assessment of task-based image quality for abdominal CT protocols linked with national diagnostic reference levels
    Viry, Anais
    Aberle, Christoph
    Lima, Thiago
    Treier, Reto
    Schindera, Sebastian T.
    Verdun, Francis R.
    Racine, Damien
    EUROPEAN RADIOLOGY, 2022, 32 (02) : 1227 - 1237
  • [6] CT iterative reconstruction algorithms: a task-based image quality assessment
    Greffier, J.
    Frandon, J.
    Larbi, A.
    Beregi, J. P.
    Pereira, F.
    EUROPEAN RADIOLOGY, 2020, 30 (01) : 487 - 500
  • [7] Efficient and Quantitative Measurement of Task-Based Image Quality Metrics in CT
    Dirks, S.
    Yu, L.
    Vrieze, T.
    Favazza, C.
    Ferrero, A.
    Leng, S.
    Weaver, J.
    McCollough, C.
    MEDICAL PHYSICS, 2017, 44 (06) : 3315 - 3315
  • [8] CT iterative reconstruction algorithms: a task-based image quality assessment
    J. Greffier
    J. Frandon
    A. Larbi
    J. P. Beregi
    F. Pereira
    European Radiology, 2020, 30 : 487 - 500
  • [9] Assessment of temporal resolution and detectability of moving objects in CT: A task-based image quality study
    Monnin, P.
    Rotzinger, D.
    Viry, A.
    Vitzthum, V.
    Racine, D.
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2024, 120
  • [10] Influence of deep learning reconstruction on task-based model observer performance in CT: an anthropomorphic head phantom study
    Hernandez-Giron, Irene
    Kaasalainen, Touko
    Makela, Teemu
    Peltonen, Juha
    Kortesniemi, Mika
    MEDICAL IMAGING 2022: IMAGE PERCEPTION, OBSERVER PERFORMANCE, AND TECHNOLOGY ASSESSMENT, 2022, 12035