Task-based assessment for radiation dose optimization in CT abdominal examinations: A phantom study

被引:0
|
作者
Halim, Esnu [1 ]
Kechik, Mohd Mustafa Awang [1 ]
Ibahim, Mohamad Johari [2 ]
Harun, Hanif Haspi [1 ]
Shaffiq, Said Mohd [3 ]
Yusof, Aimi Adibah [1 ]
Karim, Muhammad Khalis Abdul [1 ]
机构
[1] Univ Putra Malaysia, Fac Sci, Dept Phys, Serdang 43400, Selangor, Malaysia
[2] Univ Teknol MARA, Fac Med, Sungai Buloh, Selangor, Malaysia
[3] Natl Canc Inst, Putrajaya 62250, Malaysia
关键词
Abdominal CT; Optimization; Radiation dose; Signal-to-noise ratio; Figure-of-merit; IMAGE QUALITY; COMBINATION;
D O I
10.1016/j.radphyschem.2024.111966
中图分类号
O64 [物理化学(理论化学)、化学物理学];
学科分类号
070304 ; 081704 ;
摘要
Recent advancements in Computed Tomography (CT) imaging technology have significantly enhanced healthcare delivery. However, public concern regarding the safety and health risks associated with increased exposure to ionizing radiation has escalated. In response, this study performs a quantitative analysis comparing the image quality of the standard CT abdomen protocol (P1 protocols) with alternative optimization techniques. Adjustments to standard CT scan parameters, including tube potential (P2 protocols), effective mAs value (P3), and pitch factor (P4, P5, P6 protocols), were made to evaluate the effectiveness of these optimization strategies. A 64slice CT scanner (Canon, Japan) and a polymethylmethacrylate (PMMA) textural water phantom was utilized in this study. CT dose descriptors, volume weighted CT Dose Index (CTDI vol ) and dose-length product (DLP) alongside evaluation parameters such as Hounsfield Unit (HU) number, noise levels, Signal-to-Noise Ratio (SNR), and Contrast-to-Noise Ratio (CNR) were measured and analysed. Furthermore, this research introduced a Figure of Merit (FOM) based on the ratio of image quality parameters (SNR and CNR) to the CTDI vol , establishing a quantitative measure for optimization. A notable reduction in radiation dose -34.85% below the standard protocol (P1) -was observed in protocol P6, attributed to an increased pitch factor. Conversely, increase the effective mAs value has improved SNR and CNR values. The findings reveal that protocols P1 and P2 deliver the highest FOM for soft tissue and fat imaging, respectively. Protocol P1, with its higher attenuation factor, is optimal for abdominal examinations targeting soft tissues. Meanwhile, protocol P2 's increase effective mAs value enhances beam intensity, improving CT numbers for fat tissues with their high-density composition. The study also found a strong correlation between FOM based on SNR and CNR values. In conclusion, adjustments to primary CT parameters significantly influence image quality. The FOM serves as an effective tool for measuring the overall performance of image quality, facilitating post-optimization assessment and contributing to the development of safer, more efficient CT imaging protocols.
引用
收藏
页数:7
相关论文
共 50 条
  • [41] Task-based validation and application of a scanner-specific CT simulator using an anthropomorphic phantom
    Shankar, Sachin S.
    Felice, Nicholas
    Hoffman, Eric A.
    Atha, Jarron
    Sieren, Jessica C.
    Samei, Ehsan
    Abadi, Ehsan
    MEDICAL PHYSICS, 2022, 49 (12) : 7447 - 7457
  • [42] CT iterative reconstruction algorithms: a task-based image quality assessment
    J. Greffier
    J. Frandon
    A. Larbi
    J. P. Beregi
    F. Pereira
    European Radiology, 2020, 30 : 487 - 500
  • [43] Iterative Reconstruction Algorithm for Abdominal Multidetector CT at Different Tube Voltages: Assessment of Diagnostic Accuracy, Image Quality, and Radiation Dose in a Phantom Study
    Schindera, Sebastian T.
    Diedrichsen, Lars
    Mueller, Hubert C.
    Rusch, Oliver
    Marin, Daniele
    Schmidt, Bernhard
    Raupach, Rainer
    Vock, Peter
    Szucs-Farkas, Zsolt
    RADIOLOGY, 2011, 260 (02) : 454 - 462
  • [44] Deep learning image reconstruction algorithm for abdominal multidetector CT at different tube voltages: assessment of image quality and radiation dose in a phantom study
    Park, Hye Joo
    Choi, Seo-Youn
    Lee, Ji Eun
    Lim, Sanghyeok
    Lee, Min Hee
    Yi, Boem Ha
    Cha, Jang Gyu
    Min, Ji Hye
    Lee, Bora
    Jung, Yunsub
    EUROPEAN RADIOLOGY, 2022, 32 (06) : 3974 - 3984
  • [45] Deep learning image reconstruction algorithm for abdominal multidetector CT at different tube voltages: assessment of image quality and radiation dose in a phantom study
    Hye Joo Park
    Seo-Youn Choi
    Ji Eun Lee
    Sanghyeok Lim
    Min Hee Lee
    Boem Ha Yi
    Jang Gyu Cha
    Ji Hye Min
    Bora Lee
    Yunsub Jung
    European Radiology, 2022, 32 : 3974 - 3984
  • [46] Abdominal CT during pregnancy: a phantom study on the effect of patient centring on conceptus radiation dose and image quality
    Solomou, G.
    Papadakis, A. E.
    Damilakis, J.
    EUROPEAN RADIOLOGY, 2015, 25 (04) : 911 - 921
  • [47] Abdominal CT during pregnancy: a phantom study on the effect of patient centring on conceptus radiation dose and image quality
    G. Solomou
    A. E. Papadakis
    J. Damilakis
    European Radiology, 2015, 25 : 911 - 921
  • [48] Task-based characterization of a deep learning image reconstruction and comparison with filtered back-projection and a partial model-based iterative reconstruction in abdominal CT: A phantom study
    Racine, Damien
    Becce, Fabio
    Viry, Anais
    Monnin, Pascal
    Thomsen, Brian
    Verdun, Francis R.
    Rotzinger, David C.
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2020, 76 : 28 - 37
  • [49] Assessment of temporal resolution and detectability of moving objects in CT: A task-based image quality study
    Monnin, P.
    Rotzinger, D.
    Viry, A.
    Vitzthum, V.
    Racine, D.
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2024, 120
  • [50] Task-based measures of image quality and their relation to radiation dose and patient risk
    Barrett, Harrison H.
    Myers, Kyle J.
    Hoeschen, Christoph
    Kupinski, Matthew A.
    Little, Mark P.
    PHYSICS IN MEDICINE AND BIOLOGY, 2015, 60 (02): : R1 - R75