To develop an amperometric flow-biosensor for glucose, the stabilizing effect of methylene blue (MB) toward adsorbed glucose oxidase (GOx) on carbon felt (CF) was successfully applied to prepare the GOx-modified CF-based enzyme reactor combined with a horseradish peroxidase (HRP)-modified CF-based H2O2 detector. Upon mixing MB in the GOx-adsorption solution, the O2-dependent GOx-activity was significantly increased with increasing concentration of MB in the GOx-adsorption solution. The GOx-immobilization protocol on CF is very straightforward [i.e., adsorption of the GOx/MB mixed aqueous solution for 5 min under ultrasound (US)-irradiation]. Under the optimized operational conditions (i.e., applied potential, 0 vs. Ag/AgCl; carrier pH, 5.0; carrier flow rate, 4.0 mL min-1), the resulting GOx/MB-CF-reactor and HRP/TN-CF-detector combined amperometric flow-biosensor exhibited sensitive, selective, reproducible and stable cathodic peak current responses to glucose with the following analytical performances: sensitivity, 6.22 mu A mM-1; linear range, 0.01 to 1 mM; limit of detection, 9.6 mu M (S/N = 3, noise level, 20 nA); sample throughput, 46-96 samples per h for 10-0.1 mM glucose. The developed amperometric flow-biosensor allowed the determination of glucose in beverages and liquors, and the analytical results by the sensor were in fairly good agreement with those by conventional spectrophotometry. Based on the stabilizing effect of methylene blue toward glucose oxidase (GOx), amperometric flow-biosensor for glucose was developed by the combination of GOx-adsorbed carbon felt (CF)-reactor and peroxidase-adsorbed CF-based H2O2 detector.