Processing calibration data of low-temperature thermometer based on clustering algorithm

被引:0
|
作者
Liao, Yi [1 ]
Zhang, Yu [2 ,3 ]
Zha, Kuifan [2 ]
Liu, Xuming [2 ,3 ]
Pan, Changzhao [2 ,3 ]
机构
[1] Southern Univ Sci & Technol, Dept Phys, Shenzhen 518055, Peoples R China
[2] Int Quantum Acad, Shenzhen 518048, Peoples R China
[3] Southern Univ Sci & Technol, Shenzhen Inst Quantum Sci & Engn, Shenzhen 518055, Peoples R China
来源
REVIEW OF SCIENTIFIC INSTRUMENTS | 2024年 / 95卷 / 08期
基金
国家重点研发计划;
关键词
Data curation - Gluing - Network security;
D O I
10.1063/5.0216712
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The scarcity of cryogenic thermometers often stems from their high cost and lengthy lead times for calibration. Establishing an in-lab temperature calibration system is necessary to quickly make use of uncalibrated sensors or self-made sensors. This paper introduces a straightforward and high-accuracy thermometer calibration system. By employing copper screws as thermal links between the sensor platform and the cryogen-free refrigerator, temperature oscillation on the sensor platform is suppressed to a few millikelvins. In addition, this paper presents a data processing model based on clustering algorithms. These algorithms sort and group data based on distance, which is similar to human visual judgment of data. This paper discusses the parameter optimization process of the clustering algorithm to interpret the automated data process. The cryogenic temperature sensors calibrated by this system exhibited high accuracy, with relative errors of less than 1% compared to standard thermometers. Moreover, automatically processing calibration data from two uncalibrated thermometers takes just over 10 min, highlighting the effectiveness of this calibration system.
引用
收藏
页数:10
相关论文
共 50 条
  • [21] EU-151 LOW-TEMPERATURE MOSSBAUER THERMOMETER
    SHENOY, GK
    MALETTA, H
    ZEITSCHRIFT FUR PHYSIK, 1974, 269 (03): : 241 - 244
  • [22] THERMOMETER CLAMPS FOR LOW-TEMPERATURE THERMAL-CONDUCTIVITY
    GIEDD, RE
    REVIEW OF SCIENTIFIC INSTRUMENTS, 1972, 43 (12): : 1850 - &
  • [23] The N-MOS transistor as a low-temperature thermometer
    DelaHidalga, FJ
    Gutierrez, EA
    PROCEEDINGS OF THE FOURTH SYMPOSIUM ON LOW TEMPERATURE ELECTRONICS AND HIGH TEMPERATURE SUPERCONDUCTIVITY, 1997, 97 (02): : 387 - 394
  • [24] LOW-TEMPERATURE MAGNETORESISTANCE OF A SPEER CARBON RESISTANCE THERMOMETER
    THOMPSON, JR
    THOMSON, JO
    REVIEW OF SCIENTIFIC INSTRUMENTS, 1977, 48 (12): : 1713 - 1714
  • [25] TEMPERATURE CALIBRATION OF LOW-TEMPERATURE INFRARED CELLS
    FORD, TA
    SETO, PF
    FALK, M
    SPECTROCHIMICA ACTA PART A-MOLECULAR SPECTROSCOPY, 1969, A 25 (09): : 1650 - &
  • [27] LOW-TEMPERATURE RTD CALIBRATION SYSTEM
    Kozlowski, K.
    Malinowski, H.
    Dabrowski, D.
    Malinowski, K.
    Roslon, K.
    XI WORKSHOP ON PARTICLE CORRELATIONS AND FEMTOSCOPY AND NICA DAYS 2015, 2016, 9 (02): : 251 - 255
  • [28] Realization of PLTS-2000 for Low-Temperature Resistance Thermometer Calibration Services Below 650mK
    Nakagawa, Hisashi
    INTERNATIONAL JOURNAL OF THERMOPHYSICS, 2016, 37 (11)
  • [29] Carbon Clustering in Low-Temperature Bainite
    Rementeria, Rosalia
    Capdevila, Carlos
    Dominguez-Reyes, Ricardo
    Poplawsky, Jonathan D.
    Guo, Wei
    Urones-Garrote, Esteban
    Garcia-Mateo, Carlos
    Caballero, Francisca G.
    METALLURGICAL AND MATERIALS TRANSACTIONS A-PHYSICAL METALLURGY AND MATERIALS SCIENCE, 2018, 49A (11): : 5277 - 5287
  • [30] Carbon Clustering in Low-Temperature Bainite
    Rosalia Rementeria
    Carlos Capdevila
    Ricardo Domínguez-Reyes
    Jonathan D. Poplawsky
    Wei Guo
    Esteban Urones-Garrote
    Carlos Garcia-Mateo
    Francisca G. Caballero
    Metallurgical and Materials Transactions A, 2018, 49 : 5277 - 5287