Processing calibration data of low-temperature thermometer based on clustering algorithm

被引:0
|
作者
Liao, Yi [1 ]
Zhang, Yu [2 ,3 ]
Zha, Kuifan [2 ]
Liu, Xuming [2 ,3 ]
Pan, Changzhao [2 ,3 ]
机构
[1] Southern Univ Sci & Technol, Dept Phys, Shenzhen 518055, Peoples R China
[2] Int Quantum Acad, Shenzhen 518048, Peoples R China
[3] Southern Univ Sci & Technol, Shenzhen Inst Quantum Sci & Engn, Shenzhen 518055, Peoples R China
来源
REVIEW OF SCIENTIFIC INSTRUMENTS | 2024年 / 95卷 / 08期
基金
国家重点研发计划;
关键词
Data curation - Gluing - Network security;
D O I
10.1063/5.0216712
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The scarcity of cryogenic thermometers often stems from their high cost and lengthy lead times for calibration. Establishing an in-lab temperature calibration system is necessary to quickly make use of uncalibrated sensors or self-made sensors. This paper introduces a straightforward and high-accuracy thermometer calibration system. By employing copper screws as thermal links between the sensor platform and the cryogen-free refrigerator, temperature oscillation on the sensor platform is suppressed to a few millikelvins. In addition, this paper presents a data processing model based on clustering algorithms. These algorithms sort and group data based on distance, which is similar to human visual judgment of data. This paper discusses the parameter optimization process of the clustering algorithm to interpret the automated data process. The cryogenic temperature sensors calibrated by this system exhibited high accuracy, with relative errors of less than 1% compared to standard thermometers. Moreover, automatically processing calibration data from two uncalibrated thermometers takes just over 10 min, highlighting the effectiveness of this calibration system.
引用
收藏
页数:10
相关论文
共 50 条
  • [11] CARBONACEOUS MICROCOMPOSITE LOW-TEMPERATURE RESISTANCE THERMOMETER
    DULNEV, GN
    MESHKOVSKII, IK
    NOVIKOV, AF
    SHAVKUNOVA, VA
    CHERNOPOLSKII, AD
    INSTRUMENTS AND EXPERIMENTAL TECHNIQUES, 1983, 26 (01) : 230 - 232
  • [12] Low-temperature relaxations in glasses and the thermometer effect
    Roling, B
    Martiny, C
    GLASS SCIENCE AND TECHNOLOGY-GLASTECHNISCHE BERICHTE, 2000, 73 : 410 - 419
  • [13] A LOW-TEMPERATURE LIQUID-IN-GLASS THERMOMETER
    HOCHANADEL, C
    REVIEW OF SCIENTIFIC INSTRUMENTS, 1954, 25 (05): : 524 - 524
  • [14] CARBON RESISTOR AS LOW-TEMPERATURE THERMOMETER AND SENSOR
    BATRA, VK
    RESEARCH AND INDUSTRY, 1985, 30 (03): : 313 - 315
  • [15] DEVELOPMENT OF RADIATION THERMOMETER FOR LOW-TEMPERATURE RANGE
    MANABE, C
    AKAMATSU, M
    YANAI, S
    SUZUKI, N
    TERAI, K
    FUJIMOTO, S
    TETSU TO HAGANE-JOURNAL OF THE IRON AND STEEL INSTITUTE OF JAPAN, 1993, 79 (07): : 759 - 764
  • [16] A SQUID-based primary noise thermometer for low-temperature metrology
    Kirste, A.
    Engert, J.
    PHILOSOPHICAL TRANSACTIONS OF THE ROYAL SOCIETY A-MATHEMATICAL PHYSICAL AND ENGINEERING SCIENCES, 2016, 374 (2064):
  • [17] LOW-TEMPERATURE PROCESSING
    PANCHOLY, R
    SOLID STATE TECHNOLOGY, 1985, 28 (06) : 117 - 117
  • [18] Realization of PLTS-2000 for Low-Temperature Resistance Thermometer Calibration Services Below 650 mK
    Hisashi Nakagawa
    International Journal of Thermophysics, 2016, 37
  • [19] Analysis of Preparation Conditions of Low-Temperature Curing Powder Coatings Based on Local Clustering Algorithm
    Jiang Rong
    Bi Haipeng
    Zheng Ronghui
    MATHEMATICAL PROBLEMS IN ENGINEERING, 2022, 2022
  • [20] LOW-TEMPERATURE SEMICONDUCTOR RESISTANCE THERMOMETER WITH HEAT CONDUCTOR
    LOGVINENKO, SP
    ROSSOSHANSKII, OA
    CRYOGENICS, 1985, 25 (05) : 249 - 250