A feasibility study to predict 3D dose delivery accuracy for IMRT using DenseNet with log files

被引:0
|
作者
Huang, Ying [1 ,2 ,3 ]
Cai, Ruxin [3 ]
Pi, Yifei [4 ]
Ma, Kui [5 ]
Kong, Qing [1 ]
Zhuo, Weihai [2 ]
Kong, Yan [6 ]
机构
[1] Fudan Univ, Inst Modern Phys, Shanghai, Peoples R China
[2] Fudan Univ, Inst Radiat Med, Shanghai 200032, Peoples R China
[3] Shanghai Jiao Tong Univ, Shanghai Chest Hosp, Sch Med, Shanghai, Peoples R China
[4] hengzhou Univ, Affiliated Hosp 1, Dept Radiat Oncol, Zhengzhou, Henan, Peoples R China
[5] Varian Med Syst, Beijing, Peoples R China
[6] Jiangnan Univ, Dept Radiat Oncol, Affiliated Hosp, Wuxi 214122, Jiangsu, Peoples R China
基金
中国国家自然科学基金;
关键词
Three-dimensional gamma prediction; log files; patient-specific quality assurance; MLC POSITIONAL ERRORS; CLINICAL IMPLEMENTATION; QUALITY-ASSURANCE; SLIDING WINDOW; RADIOTHERAPY; VERIFICATION; EXPERIENCE;
D O I
10.3233/XST-230412
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
OBJECTIVE: This study aims to explore the feasibility of DenseNet in the establishment of a three-dimensional (3D) gamma prediction model of IMRT based on the actual parameters recorded in the log files during delivery. METHODS: A total of 55 IMRT plans (including 367 fields) were randomly selected. The gamma analysis was performed using gamma criteria of 3% /3 mm (Dose Difference/Distance to Agreement), 3% /2 mm, 2% /3 mm, and 2% /2 mm with a 10% dose threshold. In addition, the log files that recorded the gantry angle, monitor units (MU), multi-leaf collimator (MLC), and jaws position during delivery were collected. These log files were then converted to MU-weighted fluence maps as the input of DenseNet, gamma passing rates (GPRs) under four different gamma criteria as the output, and mean square errors (MSEs) as the loss function of this model. RESULTS: Under different gamma criteria, the accuracy of a 3D GPR prediction model decreased with the implementation of stricter gamma criteria. In the test set, the mean absolute error (MAE) of the prediction model under the gamma criteria of 3% /3 mm, 2% /3 mm, 3% /2 mm, and 2% /2 mm was 1.41, 1.44, 3.29, and 3.54, respectively; the root mean square error (RMSE) was 1.91, 1.85, 4.27, and 4.40, respectively; the Sr was 0.487, 0.554, 0.573, and 0.506, respectively. There was a correlation between predicted and measured GPRs (P < 0.01). Additionally, there was no significant difference in the accuracy between the validation set and the test set. The accuracy in the high GPR group was high, and the MAE in the high GPR group was smaller than that in the low GPR group under four different gamma criteria. CONCLUSIONS: In this study, a 3D GPR prediction model of patient-specific QA using DenseNet was established based on log files. As an auxiliary tool for 3D dose verification in IMRT, this model is expected to improve the accuracy and efficiency of dose validation.
引用
收藏
页码:1199 / 1208
页数:10
相关论文
共 50 条
  • [31] Evaluation of IMRT QA Using 3DVH, a 3D Patient Dose and Verification Analysis Software
    Jarry, G.
    Martin, D.
    Lemire, M.
    Ayles, M.
    Pater, P.
    MEDICAL PHYSICS, 2011, 38 (06)
  • [32] FROM 2D TO 3D DOSE EVALUATION FOR IN VIVO IMRT VERIFICATION
    Camargo, P.
    Rozendaal, R.
    Joep, S.
    Sonke, J. J.
    van Herk, M.
    Mijnheer, B.
    RADIOTHERAPY AND ONCOLOGY, 2011, 99 : S214 - S214
  • [33] Accuracy and feasibility of 3D virtual dynamic fit technology
    Kim, Youngsook
    Baytar, Fatma
    INTERNATIONAL JOURNAL OF CLOTHING SCIENCE AND TECHNOLOGY, 2024, 36 (03) : 499 - 515
  • [34] 3D Ultrasound Techniques for Accurate Dose Delivery
    Fraser, D.
    Cury, F.
    Koptenko, S.
    Scheipers, U.
    Verhaegen, F.
    MEDICAL PHYSICS, 2008, 35 (06)
  • [35] Evaluating VMAT Dose Accuracy Using 3D Detector Systems: A Comparative Study with Intentional MLC Errors
    Johnson, C.
    Varadhan, R.
    Higgins, P.
    MEDICAL PHYSICS, 2014, 41 (06) : 258 - 258
  • [36] The dosimetric impact of respiration motion on 3D, IMRT, and tomotherapy treatment delivery
    Byrne, T
    Ramsey, C
    Mahan, S
    Chase, D
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2005, 63 (02): : S496 - S496
  • [37] Impact of patient rotational errors on target and critical structure dose in IMRT: A 3D simulation study
    Arumugam, S.
    Xing, A.
    Vial, P.
    Scotti, A.
    Stirton, R.
    Goozee, G.
    Holloway, Lois
    7TH INTERNATIONAL CONFERENCE ON 3D RADIATION DOSIMETRY (IC3DDOSE), 2013, 444
  • [38] A Feasibility Study for Predicting 3D Radiotherapy Dose Distribution of Lung VMAT Patients
    Liu, Runxin
    Bai, Jingfeng
    Zhou, Jingjie
    Zhang, Kang
    Ni, Cheng
    2020 IEEE 32ND INTERNATIONAL CONFERENCE ON TOOLS WITH ARTIFICIAL INTELLIGENCE (ICTAI), 2020, : 1304 - 1308
  • [39] 3D Gamma Analysis of Dose Calculation Accuracy Using the Hypersight Imaging Solution
    Martin, P. R.
    Cherpak, A. J.
    MacDonald, L.
    Yashayaeva, A.
    McAloney, D.
    McMaster, N.
    Zhan, K.
    Cwajna, S.
    Patil, N.
    Dahn, H.
    Robar, J. L.
    MEDICAL PHYSICS, 2024, 51 (10) : 7725 - 7725
  • [40] 3D DenseNet for Improving Dose Prediction of Volumetric Modulated Arc Therapy in Prostate Cancer
    Fu, J.
    Singhrao, K.
    Wang, Z.
    Ruan, D.
    Low, D.
    Lewis, J.
    Qi, X.
    MEDICAL PHYSICS, 2021, 48 (06)