Impact of scatter radiation on spectral quantification performance of first- and second-generation dual-layer spectral computed tomography

被引:0
|
作者
Salazar, Edgar [1 ,2 ]
Liu, Leening P. [1 ,2 ]
Perkins, Amy E. [3 ]
Halliburton, Sandra S. [3 ]
Shapira, Nadav [1 ]
Litt, Harold I. [1 ]
Noel, Peter B. [1 ]
机构
[1] Univ Penn, Dept Radiol, 3620 Hamilton Walk, Philadelphia, PA 19104 USA
[2] Univ Privada Boliviana, Dept Engn & Architecture, La Paz, Bolivia
[3] Philips Healthcare, Orange Village, OH USA
来源
基金
美国国家卫生研究院;
关键词
Compton scattering artifacts; dual-layer spectral CT; quantitative CT; spectral computed tomography;
D O I
10.1002/acm2.14383
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
ObjectiveTo assess the impact of scatter radiation on quantitative performance of first and second-generation dual-layer spectral computed tomography (DLCT) systems.MethodA phantom with two iodine inserts (1 and 2 mg/mL) configured to intentionally introduce high scattering conditions was scanned with a first- and second-generation DLCT. Collimation widths (maximum of 4 cm for first generation and 8 cm for second generation) and radiation dose levels were varied. To evaluate the performance of both systems, the mean CT numbers of virtual monoenergetic images (MonoEs) at different energies were calculated and compared to expected values. MonoEs at 50 versus 150 keV were plotted to assess material characterization of both DLCTs. Additionally, iodine concentrations were determined, plotted, and compared against expected values. For each experimental scenario, absolute errors were reported.ResultsAn experimental setup, including a phantom design, was successfully implemented to simulate high scatter radiation imaging conditions. Both CT scanners illustrated high spectral accuracy for small collimation widths (1 and 2 cm). With increased collimation (4 cm), the second-generation DLCT outperformed the earlier DLCT system. Further, the spectral performance of the second-generation DLCT at an 8 cm collimation width was comparable to a 4 cm collimation on the first-generation DLCT. A comparison of the absolute errors between both systems at lower energy MonoEs illustrates that, for the same acquisition parameters, the second-generation DLCT generated results with decreased errors. Similarly, the maximum error in iodine quantification was less with second-generation DLCT (0.45 and 0.33 mg/mL for the first and second-generation DLCT, respectively).ConclusionThe implementation of a two-dimensional anti-scatter grid in the second-generation DLCT improves the spectral quantification performance. In the clinical routine, this improvement may enable additional clinical benefits, for example, in lung imaging.
引用
收藏
页数:9
相关论文
共 50 条
  • [1] Performance of novel multiparametric second-generation dual-layer spectral detector CT in gouty arthritis
    Li, Meng
    Zhang, Hui
    Liu, Jia-ni
    Zhong, Fei
    Zheng, Song-yuan
    Zhang, Jing
    Chen, Shi-xian
    Lin, Rui-feng
    Zhang, Kang-yu
    Liu, Xiao-min
    Xu, Yi-kai
    Li, Juan
    EUROPEAN RADIOLOGY, 2024,
  • [2] Fat Quantification in Dual-Layer Detector Spectral Computed Tomography Experimental Development and First In-Patient Validation
    Molwitz, Isabel
    Campbell, Graeme Michael
    Yamamura, Jin
    Knopp, Tobias
    Toedter, Klaus
    Fischer, Roland
    Wang, Zhiyue Jerry
    Busch, Alina
    Ozga, Ann-Kathrin
    Zhang, Shuo
    Lindner, Thomas
    Sevecke, Florian
    Grosser, Mirco
    Adam, Gerhard
    Szwargulski, Patryk
    INVESTIGATIVE RADIOLOGY, 2022, 57 (07) : 463 - 469
  • [3] Early rectal neoplasm in dual-layer spectral detector computed tomography: dual-layer spectral computed tomography (CT) images improve tumor detection and staging
    Pan, Xuelin
    Wu, Zhihan
    Zhao, Jin
    Zhang, Xinyi
    Zhang, Xiaodi
    Tang, Li
    Yang, Jinlin
    Deng, Kai
    QUANTITATIVE IMAGING IN MEDICINE AND SURGERY, 2024, 14 (12) : 8260 - 8271
  • [4] Feasibility and accuracy of dual-layer spectral detector computed tomography for quantification of gadolinium: a phantom study
    Robbert W. van Hamersvelt
    Martin J. Willemink
    Pim A. de Jong
    Julien Milles
    Alain Vlassenbroek
    Arnold M. R. Schilham
    Tim Leiner
    European Radiology, 2017, 27 : 3677 - 3686
  • [5] Feasibility and accuracy of dual-layer spectral detector computed tomography for quantification of gadolinium: a phantom study
    van Hamersvelt, Robbert W.
    Willemink, Martin J.
    de Jong, Pim A.
    Milles, Julien
    Vlassenbroek, Alain
    Schilham, Arnold M. R.
    Leiner, Tim
    EUROPEAN RADIOLOGY, 2017, 27 (09) : 3677 - 3686
  • [6] Dual-layer spectral computed tomography: measuring relative electron density
    Mei K.
    Ehn S.
    Oechsner M.
    Kopp F.K.
    Pfeiffer D.
    Fingerle A.A.
    Pfeiffer F.
    Combs S.E.
    Wilkens J.J.
    Rummeny E.J.
    Noël P.B.
    European Radiology Experimental, 2 (1)
  • [7] Potential of a Second-Generation Dual-Layer Spectral CT for Dose Calculation in Particle Therapy Treatment Planning
    Longarino, Friderike K.
    Kowalewski, Antonia
    Tessonnier, Thomas
    Mein, Stewart
    Ackermann, Benjamin
    Debus, Juergen
    Mairani, Andrea
    Stiller, Wolfram
    FRONTIERS IN ONCOLOGY, 2022, 12
  • [8] The diagnostic value of second-generation dual-layer spectral detector CT for Stanford type A aortic dissection
    Tan, Cong
    Gui, Shen
    Zhong, Zhilin
    Chen, Xianfeng
    EUROPEAN HEART JOURNAL-CARDIOVASCULAR IMAGING, 2024, 25 (11) : e261 - e261
  • [9] Spectral imaging with dual-layer spectral detector computed tomography for the detection of perfusion defects in acute coronary syndrome
    Mochizuki, Junji
    Nakaura, Takeshi
    Yoshida, Naofumi
    Nagayama, Yasunori
    Kidoh, Masafumi
    Uetani, Hiroyuki
    Funama, Yoshinori
    Hata, Yoshiki
    Azuma, Minako
    Hirai, Toshinori
    HEART AND VESSELS, 2022, 37 (07) : 1115 - 1124
  • [10] Spectral imaging with dual-layer spectral detector computed tomography for the detection of perfusion defects in acute coronary syndrome
    Junji Mochizuki
    Takeshi Nakaura
    Naofumi Yoshida
    Yasunori Nagayama
    Masafumi Kidoh
    Hiroyuki Uetani
    Yoshinori Funama
    Yoshiki Hata
    Minako Azuma
    Toshinori Hirai
    Heart and Vessels, 2022, 37 : 1115 - 1124