Toward Next-Generation Molecular Imaging with a Clinical Low-Field (0.064 T) Point-of-Care MRI Scanner

被引:2
|
作者
Iqbal, Nadiya [1 ]
Brittin, Drew O. [1 ]
Daluwathumullagamage, Praveen J. [1 ]
Alam, Md Shahabuddin [1 ]
Senanayake, Ishani M. [1 ]
Gafar, A. Tobi [1 ]
Siraj, Zahid [1 ]
Petrilla, Anthony [1 ]
Pugh, Margaret [1 ]
Tonazzi, Brockton [2 ]
Ragunathan, Sudarshan [3 ]
Poorman, Megan E. [3 ]
Sacolick, Laura [3 ]
Theis, Thomas [4 ]
Rosen, Matthew S. [5 ]
Chekmenev, Eduard Y. [6 ]
Goodson, Boyd M. [1 ]
机构
[1] Southern Illinois Univ, Sch Chem & Biomol Sci, Carbondale, IL 62901 USA
[2] Southern Illinois Univ, Sch Med, Carbondale, IL 62901 USA
[3] Hyperfine Inc, Guilford, CT 06437 USA
[4] North Carolina State Univ, Dept Chem, Raleigh, NC 27695 USA
[5] Harvard Univ, Sch Med, AA Martinos Ctr Biomed Imaging, Massachusetts Gen Hosp,Dept Radiol, Boston, MA 02129 USA
[6] Wayne State Univ, Karmanos Canc Inst KCI, Dept Chem, Integrat Biosci IBio, Detroit, MI 48202 USA
关键词
HETEROCYCLIC CARBENE COMPLEXES; PARA-HYDROGEN; REVERSIBLE EXCHANGE; IN-SITU; SABRE; PARAHYDROGEN; POLARIZATION; SPECTROSCOPY;
D O I
10.1021/acs.analchem.4c01299
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Low-field (LF) MRI promises soft-tissue imaging without the expensive, immobile magnets of clinical scanners but generally suffers from limited detection sensitivity and contrast. The sensitivity boost provided by hyperpolarization can thus be highly synergistic with LF MRI. Initial efforts to integrate a continuous-bubbling SABRE (signal amplification by reversible exchange) hyperpolarization setup with a portable, point-of-care 64 mT clinical MRI scanner are reported. Results from H-1 SABRE MRI of pyrazine and nicotinamide are compared with those of benchtop NMR spectroscopy. Comparison with MRI signals from samples with known H2O/D2O ratios allowed quantification of the SABRE enhancements of imaged samples with various substrate concentrations (down to 3 mM). Respective limits of detection and quantification of 3.3 and 10.1 mM were determined with pyrazine( 1)H polarization (P H) enhancements of similar to 1900 (P- H similar to 0.04%), supporting ongoing and envisioned efforts to realize SABRE-enabled MRI-based molecular imaging.
引用
收藏
页码:10348 / 10355
页数:8
相关论文
共 50 条
  • [21] Potential of Stroke Imaging Using a New Prototype of Low-Field MRI: A Prospective Direct 0.55 T/1.5 T Scanner Comparison
    Rusche, Thilo
    Breit, Hanns-Christian
    Bach, Michael
    Wasserthal, Jakob
    Gehweiler, Julian
    Manneck, Sebastian
    Lieb, Johanna Maria
    De Marchis, Gian Marco
    Psychogios, Marios Nikos
    Sporns, Peter B.
    JOURNAL OF CLINICAL MEDICINE, 2022, 11 (10)
  • [22] Intracerebral Hemorrhage and Clinical Outcome on Low Field, Portable, Point-of-Care Magnetic Resonance Imaging
    Yuen, Matthew M.
    Cahn, Bradley A.
    Mazurek, Mercy H.
    Shah, Jill T.
    Welch, E. Brian
    By, Samantha
    Hu, Houchun Harry
    Sacolick, Laura I.
    Ward, Adrienne
    Timario, Nona
    Falcone, Guido J.
    Hwang, David Y.
    Kim, Jennifer A.
    Jasne, Adam
    Gordon-Kundu, Barbara
    Gobeske, Kevin
    Sharma, Richa
    Hebert, Ryan M.
    Kaddouh, Firas
    Gilmore, Emily J.
    Schindler, Joseph
    Matouk, Charles
    Wira, Charles
    Sze, Gordon
    Rosen, Matthew S.
    Kimberly, W. Taylor
    Sheth, Kevin N.
    ANNALS OF NEUROLOGY, 2020, 88 : S209 - S209
  • [23] Advocating the Development of Next-Generation, Advanced-Design Low-Field Magnetic Resonance Systems
    Runge, Val M.
    Heverhagen, Johannes T.
    INVESTIGATIVE RADIOLOGY, 2020, 55 (12) : 747 - 753
  • [24] Next-Generation Intelligent MXene-Based Electrochemical Aptasensors for Point-of-Care Cancer Diagnostics
    Parihar, Arpana
    Singhal, Ayushi
    Kumar, Neeraj
    Khan, Raju
    Khan, Mohd Akram
    Srivastava, Avanish K.
    NANO-MICRO LETTERS, 2022, 14 (01)
  • [25] Next-Generation Intelligent MXene-Based Electrochemical Aptasensors for Point-of-Care Cancer Diagnostics
    Arpana Parihar
    Ayushi Singhal
    Neeraj Kumar
    Raju Khan
    Mohd. Akram Khan
    Avanish K. Srivastava
    Nano-Micro Letters, 2022, 14
  • [26] Next-generation microfluidic lab-on-a-chip platforms for point-of-care diagnostics and systems biology
    Ducree, Jens
    PROCEEDINGS OF THE EUROSENSORS XXIII CONFERENCE, 2009, 1 (01): : 517 - 520
  • [27] Next-Generation Intelligent MXene-Based Electrochemical Aptasensors for Point-of-Care Cancer Diagnostics
    Arpana Parihar
    Ayushi Singhal
    Neeraj Kumar
    Raju Khan
    Mohd.Akram Khan
    Avanish K.Srivastava
    Nano-Micro Letters, 2022, 14 (06) : 254 - 287
  • [28] Low-Field (64 mT) Portable MRI for Rapid Point-of-Care Diagnosis of Dissemination in Space in Patients Presenting with Optic Neuritis
    Lim, Timothy Reynold
    Suthiphosuwan, Suradech
    Micieli, Jonathan
    Vosoughi, Reza
    Schneider, Raphael
    Lin, Amy W.
    Chen, Yingming Amy
    Muccilli, Alexandra
    Marriott, James J.
    Selchen, Daniel
    Mathur, Shobhit
    Oh, Jiwon
    Bharatha, Aditya
    AMERICAN JOURNAL OF NEURORADIOLOGY, 2024, 45 (01) : 1819 - 1825
  • [29] Low-Field Magnetic Resonance Imaging A New Generation of Breakthrough Technology in Clinical Imaging
    Heiss, Rafael
    Nagel, Armin M.
    Lain, Frederik B.
    Uder, Michael
    Bickelhaupt, Sebastian
    INVESTIGATIVE RADIOLOGY, 2021, 56 (11) : 726 - 733
  • [30] Point-of-Care Low-Field-Strength MRI Is Moving Beyond the Hype