Calibration of buried NaI(Tl) scintillator detectors for natural radionuclide measurement based on Monte Carlo modelling

被引:0
|
作者
Chambon, A. [1 ]
Klinkby, E. B. [1 ]
Bu, M. [1 ]
Murray, A. S. [2 ,3 ]
Kook, M. [1 ]
Olesen, H. [1 ]
Nielsen, K. B. [1 ]
Lauritzen, B. [1 ]
机构
[1] Tech Univ Denmark, Dept Phys, DTU Riso Campus, DK-4000 Roskilde, Denmark
[2] Aarhus Univ, Dept Geosci, Nord Lab Luminescence Dating, DTU Riso Campus, DK-4000 Roskilde, Denmark
[3] DTU Phys, DTU Riso Campus, DK-4000 Roskilde, Denmark
关键词
NaI(Tl) detector; Field measurement; Calibration; MCNP; RESPONSE FUNCTIONS; GAMMA;
D O I
10.1016/j.radphyschem.2024.111803
中图分类号
O64 [物理化学(理论化学)、化学物理学];
学科分类号
070304 ; 081704 ;
摘要
Measurements of naturally occurring concentrations of 40K, and of the decay series of 238U and 232Th, are of interest in the earth sciences in general, and in particular, scintillator-based gamma spectrometers can be used for the low cost determination of burial dose rates in natural geological samples. We are currently developing a robust, portable, wireless detector specifically intended for field measurement of natural radionuclide concentrations and hence, the calculation of dose rates. One of the challenges in developing and applying such an instrument is reliable calibration. Most calibrations of field instruments depend on access to non-finite matrices of known K, U, Th activity concentrations, in either a 4 pi or 2 pi geometry; these are only available at a few facilities around the world. Here we investigate an alternative approach, based on the measurement of small samples containing well-known activity concentrations of only K or U or Th, and Monte-Carlo radiation transport modelling to convert the observed spectra into those expected from specific activity concentrations in a nonfinite 4 pi geometry. We first validate our modelling procedure by simulating these observed spectra. The nonfinite matrix calibration spectra are then predicted, and least-squares fitted to the spectrum observed at the centre of a 1 m3 of granite chips; the resulting predicted U, Th and K activity concentrations are compared with independently known values.
引用
收藏
页数:8
相关论文
共 46 条
  • [21] Efficiency calibration for a NaI scintillation detector based on Monte-Carlo process and preliminary measurements of bremsstrahlung
    Huang Jian-Wei
    Wang Nai-Yan
    ACTA PHYSICA SINICA, 2014, 63 (18)
  • [22] Monte Carlo based calibration of scintillation detectors for laboratory and in situ gamma ray measurements
    van der Graaf, E. R.
    Limburg, J.
    Koomans, R. L.
    Tijs, M.
    JOURNAL OF ENVIRONMENTAL RADIOACTIVITY, 2011, 102 (03) : 270 - 282
  • [24] An inversion decomposition test based on Monte Carlo response matrix on the γ-ray spectra from NaI(Tl) scintillation detector
    He, Jian-Feng
    Wu, Qi-Fan
    Cheng, Jian-Ping
    Fang, Fang
    Yang, Yao-Zong
    Qu, Jin-Hui
    NUCLEAR SCIENCE AND TECHNIQUES, 2016, 27 (04)
  • [25] An inversion decomposition test based on Monte Carlo response matrix on the γ-ray spectra from NaI(Tl) scintillation detector
    Jian-Feng He
    Qi-Fan Wu
    Jian-Ping Cheng
    Fang Fang
    Yao-Zong Yang
    Jin-Hui Qu
    Nuclear Science and Techniques, 2016, 27 (04) : 76 - 87
  • [26] An inversion decomposition test based on Monte Carlo response matrix on the γ-ray spectra from NaI(Tl) scintillation detector
    Jian-Feng He
    Qi-Fan Wu
    Jian-Ping Cheng
    Fang Fang
    Yao-Zong Yang
    Jin-Hui Qu
    Nuclear Science and Techniques, 2016, 27
  • [27] Precise Monte Carlo Simulations of NaI(Tl) Spectra for Measurement of Actinium-225 Targeted Alpha Particle Bio-Distribution
    Tichacek, Christopher J.
    Budzevich, Mikalai M.
    Martinez, Gary V.
    Morse, David L.
    Moros, Eduardo G.
    MEDICAL PHYSICS, 2017, 44 (06) : 3227 - 3228
  • [28] A new NaI(Tl) four-detector layout for field contamination assessment using artificial neural networks and the Monte Carlo method for system calibration
    Moreira, M. C. F.
    Conti, C. C.
    Schirru, R.
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2010, 621 (1-3): : 302 - 309
  • [29] Laser triangulation measurement system with Scheimpflug calibration based on the Monte Carlo optimization strategy
    Chen, Ruiming
    Li, Yiming
    Xue, Gaopeng
    Tao, Yihao
    Li, Xinghui
    OPTICS EXPRESS, 2022, 30 (14) : 25290 - 25307
  • [30] Monte Carlo modelling of Germanium detectors for the measurement of low energy photons in internal dosimetry: Results of an international comparison
    Gomez-Ros, J. M.
    de Carlan, L.
    Franck, D.
    Gualdrini, G.
    Lis, M.
    Lopez, M. A.
    Moraleda, M.
    Zankl, M.
    Badal, A.
    Capello, K.
    Cowan, P.
    Ferrari, P.
    Heide, B.
    Henniger, J.
    Hooley, V.
    Hunt, J.
    Kinase, S.
    Kramer, G. H.
    Loehnert, D.
    Lucas, S.
    Nuttens, V.
    Packer, L. W.
    Reichelt, U.
    Vrba, T.
    Sempau, J.
    Zhang, B.
    RADIATION MEASUREMENTS, 2008, 43 (2-6) : 510 - 515