Calibration of buried NaI(Tl) scintillator detectors for natural radionuclide measurement based on Monte Carlo modelling

被引:0
|
作者
Chambon, A. [1 ]
Klinkby, E. B. [1 ]
Bu, M. [1 ]
Murray, A. S. [2 ,3 ]
Kook, M. [1 ]
Olesen, H. [1 ]
Nielsen, K. B. [1 ]
Lauritzen, B. [1 ]
机构
[1] Tech Univ Denmark, Dept Phys, DTU Riso Campus, DK-4000 Roskilde, Denmark
[2] Aarhus Univ, Dept Geosci, Nord Lab Luminescence Dating, DTU Riso Campus, DK-4000 Roskilde, Denmark
[3] DTU Phys, DTU Riso Campus, DK-4000 Roskilde, Denmark
关键词
NaI(Tl) detector; Field measurement; Calibration; MCNP; RESPONSE FUNCTIONS; GAMMA;
D O I
10.1016/j.radphyschem.2024.111803
中图分类号
O64 [物理化学(理论化学)、化学物理学];
学科分类号
070304 ; 081704 ;
摘要
Measurements of naturally occurring concentrations of 40K, and of the decay series of 238U and 232Th, are of interest in the earth sciences in general, and in particular, scintillator-based gamma spectrometers can be used for the low cost determination of burial dose rates in natural geological samples. We are currently developing a robust, portable, wireless detector specifically intended for field measurement of natural radionuclide concentrations and hence, the calculation of dose rates. One of the challenges in developing and applying such an instrument is reliable calibration. Most calibrations of field instruments depend on access to non-finite matrices of known K, U, Th activity concentrations, in either a 4 pi or 2 pi geometry; these are only available at a few facilities around the world. Here we investigate an alternative approach, based on the measurement of small samples containing well-known activity concentrations of only K or U or Th, and Monte-Carlo radiation transport modelling to convert the observed spectra into those expected from specific activity concentrations in a nonfinite 4 pi geometry. We first validate our modelling procedure by simulating these observed spectra. The nonfinite matrix calibration spectra are then predicted, and least-squares fitted to the spectrum observed at the centre of a 1 m3 of granite chips; the resulting predicted U, Th and K activity concentrations are compared with independently known values.
引用
收藏
页数:8
相关论文
共 46 条
  • [41] Comparison of the X-ray tube spectrum measurement using BGO, NaI, LYSO, and HPGe detectors in a preclinical mini-CT scanner: Monte Carlo simulation and practical experiment
    Lohrabian, Vahid
    Kamali-Asl, Alireza
    Harvani, Hossein Ghadiri
    Hosseini Aghdam, Seyed Rashid
    Arabi, Hossein
    Zaidi, Habib
    Radiation Physics and Chemistry, 2021, 189
  • [42] Comparison of the X-ray tube spectrum measurement using BGO, NaI, LYSO, and HPGe detectors in a preclinical mini-CT scanner: Monte Carlo simulation and practical experiment
    Lohrabian, Vahid
    Kamali-Asl, Alireza
    Harvani, Hossein Ghadiri
    Aghdam, Seyed Rashid Hosseini
    Arabi, Hossein
    Zaidi, Habib
    RADIATION PHYSICS AND CHEMISTRY, 2021, 189
  • [43] Small field output factor measurement and verification for CyberKnife robotic radiotherapy and radiosurgery system using 3D polymer gel, ionization chamber, diode, diamond and scintillator detectors, Gafchromic film and Monte Carlo simulation
    Kafi, Md Abdullah Al
    Arib, Mehenna
    Al Moussa, Akram
    Alzorkany, Faisal
    Shehadeh, Mamoun
    Yusof, Mohd Fahmi Mohd
    Moftah, Belal
    APPLIED RADIATION AND ISOTOPES, 2023, 192
  • [44] The elusive ettringite under the high-vacuum SEM - a reflection based on natural samples, the use of Monte Carlo modelling of EDS analyses and an extension to the ettringite group minerals
    Thiery, Vincent
    Trincal, Vincent
    Davy, Catherine A.
    JOURNAL OF MICROSCOPY, 2017, 268 (01) : 84 - 93
  • [45] Depth sensitivity analysis of functional near-infrared spectroscopy measurement using three-dimensional Monte Carlo modelling-based magnetic resonance imaging
    Chemseddine Mansouri
    Jean-Pierre L’Huillier
    Nasser H. Kashou
    Anne Humeau
    Lasers in Medical Science, 2010, 25 : 431 - 438
  • [46] Depth sensitivity analysis of functional near-infrared spectroscopy measurement using three-dimensional Monte Carlo modelling-based magnetic resonance imaging
    Mansouri, Chemseddine
    L'Huillier, Jean-Pierre
    Kashou, Nasser H.
    Humeau, Anne
    LASERS IN MEDICAL SCIENCE, 2010, 25 (03) : 431 - 438