Development and optimization of a backscatter X-ray detection system based on pencil beam scanning

被引:0
|
作者
Park, Jaehyun [1 ]
An, Geunyoung [2 ]
Song, Gyohyeok [1 ]
Kim, Wonku [1 ]
Kim, Junhyeok [1 ]
Lim, Kyung Taek [3 ]
Hwang, Jisung [1 ]
Cho, Gyuseong [1 ]
机构
[1] Korea Adv Inst Sci & Technol, Dept Nucl & Quantum Engn, 291 Daehak Ro, Daejeon 34141, South Korea
[2] Jeonbuk Natl Univ, Dept Appl Plasma & Quantum Beam Engn, 567 Baekje Do, Jeonju, South Korea
[3] Sejong Univ, Dept Quantum & Nucl Engn, Seoul 05006, South Korea
基金
新加坡国家研究基金会;
关键词
Backscatter X-ray; Backscatter X-ray image; PVT; Compton scattering; DAQ board; DETECTION EFFICIENCY;
D O I
10.1016/j.nima.2024.169277
中图分类号
TH7 [仪器、仪表];
学科分类号
0804 ; 080401 ; 081102 ;
摘要
The backscatter X-ray system makes an image by detecting scattered particles from an object in the opposite direction by Compton scattering. This system has the advantage of detecting drugs or explosives composed of organic substances. In this study, we developed a backscatter X-ray detection system. The 6-ch DAQ system is able to store acquired data for 50 mu s as 8 bytes and the 6-ch high voltage power supply board provides adjustable power from 0 to 1600 V. Using this detection system, we varied the tube voltage from 80 to 160 kV and observed backscatter X-ray images of two plastic phantoms at 160 kV. The results indicate a need to optimize the speed of the collimator and data acquisition time of the DAQ system to improve image quality. The backscatter X-ray imaging technology exhibits high scalability, making it applicable not only to security screening but also for medical and industrial purposes.
引用
收藏
页数:4
相关论文
共 50 条
  • [21] Monte-Carlo simulation for the optimization of a fan-beam X-ray backscatter system for the identification of human remains
    Choi, Sunghoon
    Park, Sora
    Kang, Jun-Tae
    Yeon, Ji-Hwan
    Jeong, Jin-Woo
    RADIATION PHYSICS AND CHEMISTRY, 2025, 229
  • [22] Status of the XMIS x-ray backscatter radiography land mine detection system
    Dugan, ET
    Jacobs, AM
    Su, Z
    Houssay, L
    Ekdahl, D
    DETECTION AND REMEDIATION TECHNOLOGIES FOR MINES AND MINELIKE TARGETS VIII, PTS 1 AND 2, 2003, 5089 : 34 - 44
  • [23] Detection of buried landmine with X-ray backscatter technique
    Yuk, Sunwoo
    Kim, Kwang Hyun
    Yi, Yun
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2006, 568 (01): : 388 - 392
  • [24] Detection of buried landmines with X-ray backscatter technology
    Niemann, Wilhelm
    Olesinski, S.
    Thiele, T.
    Martens, G.
    Carlsen, I.-C.
    Insight: Non-Destructive Testing and Condition Monitoring, 2002, 44 (10): : 634 - 637
  • [25] Detection of buried landmines with X-ray backscatter technology
    Niemann, W
    Olesinski, S
    Thiele, T
    Martens, G
    Carlsen, IC
    INSIGHT, 2002, 44 (10) : 634 - 637
  • [26] Field tests of x-ray backscatter mine detection
    Lockwood, GJ
    Shope, SL
    Wehlburg, JC
    Selph, MM
    Jojola, JM
    Turman, BN
    Jacobs, JA
    SECOND INTERNATIONAL CONFERENCE ON THE DETECTION OF ABANDONED LAND MINES, 1998, (458): : 160 - 163
  • [27] Corrosion detection in aircraft by X-ray backscatter methods
    Dunn, WL
    Yacout, AM
    APPLIED RADIATION AND ISOTOPES, 2000, 53 (4-5) : 625 - 632
  • [28] Scanning-beam digital x-ray system significantly reduces x-ray exposure for cardiac angiography
    Van Lysel, MS
    Wolff, MR
    Tanke, TE
    Speidel, MA
    Betts, TD
    Heanue, JA
    Wilfley, BP
    Pretti, J
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 2000, 35 (02) : 54A - 54A
  • [29] X-ray Fluorescence Computed Tomography (XFCT) Imaging with a Superfine Pencil Beam X-ray Source
    Romero, Ignacio O.
    Fang, Yile
    Lun, Michael
    Li, Changqing
    PHOTONICS, 2021, 8 (07)
  • [30] Development of a GPU-Based Dose Delivery System for Adaptive Pencil Beam Scanning
    Giordanengo, S.
    Russo, G.
    Marchetto, F.
    Monaco, V.
    Pella, A.
    Varasteh, M.
    Attili, A.
    MEDICAL PHYSICS, 2014, 41 (06) : 246 - 247