RELIABILITY OF INTRAOCULAR-PRESSURE MEASUREMENT WITH THE GOLDMANN APPLANATION TONOMETER IN EPIDEMIOLOGIC STUDIES

被引:153
|
作者
DIELEMANS, I
VINGERLING, JR
HOFMAN, A
GROBBEE, DE
DEJONG, PTVM
机构
[1] ERASMUS UNIV ROTTERDAM,SCH MED,DEPT OPHTALMOL,3000 DR ROTTERDAM,NETHERLANDS
[2] ERASMUS UNIV ROTTERDAM,SCH MED,DEPT EPIDEMIOL & BIOSTAT,3000 DR ROTTERDAM,NETHERLANDS
关键词
D O I
10.1007/BF00176782
中图分类号
R77 [眼科学];
学科分类号
100212 ;
摘要
The reproducibility of intraocular pressure (IOP) measurement with the Goldmann applanation tonometer was investigated as part of a population-based epidemiological study. Sixty-two subjects were examined in a first measurement session. The IOP was measured three times consecutively in both eyes according to a fixed protocol. The mean standard deviation (SD) of these measurements was 0.8 mmHg. The mean intra-observer variation for the first measurement was 1.64 (SD 2.07) mmHg. For the median of the three measurements the intra-observer variation was 1.50 (SD 1.96) mmHg. The mean inter-observer values were 1.79 (SD 2.41) mmHg for the first measurement and 1.60 (SD 2.15) mmHg for the median measurement. The correlation coefficient between the median values of the three measurements of both observers was 0.81. No systematic differences were found between the two observers. Using the median value of three consecutive measurements reduced the inter-observer variation by 11% and the intra-observer variation by 9% compared with a single measurement.
引用
收藏
页码:141 / 144
页数:4
相关论文
共 50 条
  • [11] Transpalpebral measurement of intraocular pressure using the Diaton tonometer versus standard Goldmann applanation tonometry
    Yuehua Li
    Jingming Shi
    Xuanchu Duan
    Fang Fan
    Graefe's Archive for Clinical and Experimental Ophthalmology, 2010, 248 : 1765 - 1770
  • [12] STUDIES OF INTRAOCULAR PRESSURE WITH GOLDMANS APPLANATION TONOMETER
    BECHRAKIS, E
    OPHTHALMOLOGICA, 1970, 161 (2-3) : 353 - +
  • [13] Level of disagreement between Proview™ phosphene tonometer and Goldmann applanation tonometer Intraocular pressure readings
    Rai, S
    Moster, MR
    Kesen, M
    Fontanarosa, J
    Spaeth, GL
    Steinmann, WC
    Wilson, R
    Myers, J
    Henderer, J
    JOURNAL OF GLAUCOMA, 2005, 14 (02) : 120 - 123
  • [14] Dynamic Contour Tonometer and Goldmann Applanation Tonometer Performance in a Developing World Setting: Intraocular Pressure Measurement Acquisition and Precision
    Cheng, Jason
    Salam, Tahrina
    Russell, Paul J.
    Heath, David G.
    Kotecha, Aachal
    JOURNAL OF GLAUCOMA, 2013, 22 (09) : 736 - 739
  • [15] Comparative evaluation of intraocular pressure with an air-puff tonometer versus a Goldmann applanation tonometer
    Farhood, Qasim K.
    CLINICAL OPHTHALMOLOGY, 2013, 7 : 23 - 27
  • [16] Influence of punctate superficial keratitis on intraocular pressure measured with Goldmann applanation tonometer and noncontact tonometer
    Shimizu, Daisuke
    Yamamoto, Shuichi
    INVESTIGATIVE OPHTHALMOLOGY & VISUAL SCIENCE, 2019, 60 (09)
  • [17] A comparison of intraocular pressure measurements with the XPERT noncontact tonometer and Goldmann applanation tonometry
    Popovich, KS
    Shields, MB
    JOURNAL OF GLAUCOMA, 1997, 6 (01) : 44 - 46
  • [18] Effect of corneal astigmatism on intraocular pressure measurement using ocular response analyzer and Goldmann applanation tonometer
    Mana Hagishima
    Kazutaka Kamiya
    Fusako Fujimura
    Tetsuya Morita
    Nobuyuki Shoji
    Kimiya Shimizu
    Graefe's Archive for Clinical and Experimental Ophthalmology, 2010, 248 : 257 - 262
  • [19] Effect of corneal astigmatism on intraocular pressure measurement using ocular response analyzer and Goldmann applanation tonometer
    Hagishima, Mana
    Kamiya, Kazutaka
    Fujimura, Fusako
    Morita, Tetsuya
    Shoji, Nobuyuki
    Shimizu, Kimiya
    GRAEFES ARCHIVE FOR CLINICAL AND EXPERIMENTAL OPHTHALMOLOGY, 2010, 248 (02) : 257 - 262
  • [20] Measurement of Goldmann Applanation Tonometer Calibration Error
    Choudhari, Nikhil S.
    George, Ronnie
    Baskaran, Mani
    Vijaya, Lingam
    Dudeja, Namrata
    OPHTHALMOLOGY, 2009, 116 (01) : 3 - 8