DIAGNOSTIC PERFORMANCE - A COMPARATIVE-STUDY OF THE LEUKOCYTE DIFFERENTIAL COUNT ON 4 AUTOMATED HEMATOLOGY ANALYZERS

被引:0
|
作者
BUTTARELLO, M
LORENZ, C
GADOTTI, M
TOFFALORI, E
VALENTINI, A
RIZZOTTI, P
机构
[1] OSPED GERIATR,ANAL CHIMICOCLIN & MICROBIOL LAB,VIA VENDRAMINI 7,I-35100 PADUA,ITALY
[2] OSPED SANTA CHIARA,BIOCHIM & EMATOL LAB,TRENT,ITALY
[3] OSPED SANTA CHIARA,SERV FIS SANIT,TRENT,ITALY
关键词
D O I
暂无
中图分类号
Q5 [生物化学]; Q7 [分子生物学];
学科分类号
071010 ; 081704 ;
摘要
The diagnostic performance of the haematology analysers, Technicon H1, Sysmex NE 8000, Coulter STKS and Sequoia Cell Dyn 3000 was determined by comparing their results for automated differential counting with the result obtained for 400 cells by the manual microscopic method. These comparative studies were performed on a group of adults, containing both normal individuals (n = 150) and those affected by various pathologies (n = 113). The number of morphologically false negatives is low for all the systems (from 1.8% for the Cell Dyn 3000 to 6.2% for the NE 8000), while the number of distributionally false negatives was slightly higher (from 4.4% for the Cell Dyn 3000 to 7% for the H1 and NE 8000). The morphological flags, though useful for improving the diagnostic performance of the instruments, show a rather modest sensitivity when taken individually: immature granulocytes/bands from 71% for the STKS to 43% for the NE 8000; blasts from 66.6% for the H1 to 53.3% for the STKS, atypical lymphocytes from 59% for the H1 to 13.6% for the NE 8000; erythroblasts from 42.8% for the STKS to 7% for the NE 8000.
引用
收藏
页码:251 / 258
页数:8
相关论文
共 50 条
  • [41] WITHIN-DAY PHYSIOLOGIC VARIATION OF LEUKOCYTE TYPES IN HEALTHY-SUBJECTS AS ASSAYED BY 2 AUTOMATED LEUKOCYTE DIFFERENTIAL ANALYZERS
    WINKEL, P
    STATLAND, BE
    SAUNDERS, AM
    OSBORN, H
    KUPPERMAN, H
    AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 1981, 75 (05) : 693 - 700
  • [42] AUTOMATED CYTOCHEMICAL DIFFERENTIAL LEUKOCYTE COUNT IN PATIENTS HOSPITALIZED WITH ACUTE BACTERIAL-INFECTIONS
    MYHRE, EB
    BRACONIER, JH
    SJOGREN, U
    SCANDINAVIAN JOURNAL OF INFECTIOUS DISEASES, 1985, 17 (02) : 201 - 208
  • [43] QUALITY TEAM-APPROACH IN EVALUATING 3 AUTOMATED HEMATOLOGY ANALYZERS WITH 5-PART DIFFERENTIAL CAPABILITY
    JONES, RG
    FAUST, AM
    MATTHEWS, RA
    AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 1995, 103 (02) : 159 - 166
  • [44] Performance evaluation of the automated nucleated red blood cell count of five commercial hematological analyzers
    Da Rin, G.
    Vidali, M.
    Balboni, F.
    Benegiamo, A.
    Borin, M.
    Ciardelli, M. L.
    Dima, F.
    Di Fabio, A.
    Fanelli, A.
    Fiorini, F.
    Francione, S.
    Germagnoli, L.
    Gioia, M.
    Lari, T.
    Lorubbio, M.
    Marini, A.
    Papa, A.
    Seghezzi, M.
    Solarino, L.
    Pipitone, S.
    Tilocca, E.
    Buoro, S.
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2017, 39 (06) : 663 - 670
  • [45] COMPARATIVE-STUDY OF 4 ANTACIDS
    JACYNA, MR
    BOYD, EJS
    WORMSLEY, KG
    POSTGRADUATE MEDICAL JOURNAL, 1984, 60 (707) : 592 - 596
  • [46] LACTASE DEFICIENCY - COMPARATIVE-STUDY OF DIAGNOSTIC METHODS
    ARVANITAKIS, C
    CHEN, GH
    FOLSCROFT, J
    KLOTZ, AP
    AMERICAN JOURNAL OF CLINICAL NUTRITION, 1977, 30 (10): : 1597 - 1602
  • [47] Performance evaluation of a novel platelet count parameter, hybrid platelet count, on the BC-780 automated hematology analyzer
    Prompetchara, Eakachai
    Parnsamut, Chalisa
    Chirapanuruk, Angkana
    Ketloy, Chutitorn
    CLINICAL CHEMISTRY AND LABORATORY MEDICINE, 2024, 62 (04) : 690 - 697
  • [48] Changes in automated complete blood cell count and differential leukocyte count results induced by storage of blood at room temperature
    Gulati, GL
    Hyland, LJ
    Kocher, W
    Schwarting, R
    ARCHIVES OF PATHOLOGY & LABORATORY MEDICINE, 2002, 126 (03) : 336 - 342
  • [49] Use of an automated hematology analyzer and flow cytometry to assess bone marrow cellularity and differential cell count
    Kim, N
    Kim, J
    Lim, J
    Kim, Y
    Han, K
    Kang, CS
    ANNALS OF CLINICAL AND LABORATORY SCIENCE, 2004, 34 (03): : 307 - 313
  • [50] A COMPARATIVE-STUDY OF VARIOUS CELL SEPARATORS FOR THE PRODUCTION OF LEUKOCYTE CONCENTRATES
    AUBERT, C
    TOUZET, C
    DUEDARI, N
    ANNALES DE MEDECINE INTERNE, 1984, 135 : 3 - 3