DIAGNOSTIC PERFORMANCE - A COMPARATIVE-STUDY OF THE LEUKOCYTE DIFFERENTIAL COUNT ON 4 AUTOMATED HEMATOLOGY ANALYZERS

被引:0
|
作者
BUTTARELLO, M
LORENZ, C
GADOTTI, M
TOFFALORI, E
VALENTINI, A
RIZZOTTI, P
机构
[1] OSPED GERIATR,ANAL CHIMICOCLIN & MICROBIOL LAB,VIA VENDRAMINI 7,I-35100 PADUA,ITALY
[2] OSPED SANTA CHIARA,BIOCHIM & EMATOL LAB,TRENT,ITALY
[3] OSPED SANTA CHIARA,SERV FIS SANIT,TRENT,ITALY
关键词
D O I
暂无
中图分类号
Q5 [生物化学]; Q7 [分子生物学];
学科分类号
071010 ; 081704 ;
摘要
The diagnostic performance of the haematology analysers, Technicon H1, Sysmex NE 8000, Coulter STKS and Sequoia Cell Dyn 3000 was determined by comparing their results for automated differential counting with the result obtained for 400 cells by the manual microscopic method. These comparative studies were performed on a group of adults, containing both normal individuals (n = 150) and those affected by various pathologies (n = 113). The number of morphologically false negatives is low for all the systems (from 1.8% for the Cell Dyn 3000 to 6.2% for the NE 8000), while the number of distributionally false negatives was slightly higher (from 4.4% for the Cell Dyn 3000 to 7% for the H1 and NE 8000). The morphological flags, though useful for improving the diagnostic performance of the instruments, show a rather modest sensitivity when taken individually: immature granulocytes/bands from 71% for the STKS to 43% for the NE 8000; blasts from 66.6% for the H1 to 53.3% for the STKS, atypical lymphocytes from 59% for the H1 to 13.6% for the NE 8000; erythroblasts from 42.8% for the STKS to 7% for the NE 8000.
引用
收藏
页码:251 / 258
页数:8
相关论文
共 50 条
  • [21] EVALUATION OF THE AUTOMATED LEUKOCYTE DIFFERENTIAL COUNT IN EMERGENCY DEPARTMENT PATIENTS
    RIMARENKO, S
    CASTELLA, A
    SALZBERG, MR
    STRAND, CL
    AMERICAN JOURNAL OF EMERGENCY MEDICINE, 1987, 5 (03): : 187 - 189
  • [22] COMPARATIVE-STUDY OF 4 AUTOMATED DONOR PLASMAPHERESIS SYSTEMS
    COFFE, C
    COUTERET, Y
    MASSE, M
    LAMY, B
    HERVE, P
    KIEFFER, Y
    PETERS, A
    ARTIFICIAL ORGANS, 1988, 12 (03) : 267 - 268
  • [23] Performance evaluation of the automated cell image analyzer DI-60 for leukocyte differential count
    Chang, S.
    Kim, E.
    Kim, Y.
    Won, D.
    CLINICAL CHEMISTRY, 2024, 70 : I143 - I143
  • [24] Performance Evaluation of the Automated Cell Image Analyzer DI-60 for Leukocyte Differential Count
    Chang, Soon Hee
    CLINICAL LABORATORY, 2025, 71 (03) : 538 - 544
  • [25] Performance of the Pentra DX120 automated haematology analyzer in evaluating the leukocyte differential count
    Nagai, N.
    Maegawa, Y.
    Yamamoto, Y.
    Matuo, S.
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2008, 30 : 73 - 74
  • [26] Diagnostic usefulness of pleural fluid total leukocyte count and differential
    Porcel, J. M.
    REVISTA CLINICA ESPANOLA, 2017, 217 (03): : 149 - 150
  • [27] TOWARD A REFERENCE METHOD FOR LEUKOCYTE DIFFERENTIAL COUNTS IN BLOOD FOR EVALUATION OF AUTOMATED HEMATOLOGY ANALYZERS: VALIDATION OF FLOWCYTOMETRIC JS']JSLH DIFF METHOD AND SOP
    Kawai, Yohko
    Kondo, Hiroshi
    Yoneyama, Akiko
    Ikemoto, Toshiyuki
    Tsuruda, Kazuto
    Shimizu, Nobuko
    Shirakami, Atsushi
    Banzai, Mariko
    Ikeda, Hisataka
    Kondo, Yasunori
    Sugiyama, Yoko
    Yonekawa, Osamu
    Ogawa, Etsuko
    Nagai, Yutaka
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2014, 36 : 99 - 100
  • [28] A COMPARATIVE-STUDY OF HEMATOLOGY OF 3 SPECIES OF RANA
    SAMANTARAY, K
    JOURNAL OF ADVANCED ZOOLOGY, 1984, 5 (01): : 42 - 47
  • [29] COMPARATIVE-STUDY OF 2 COMPUTERIZED SEMEN MOTILITY ANALYZERS
    GILL, HS
    VANARSDALEN, K
    HYPOLITE, J
    LEVIN, RM
    RUZICH, JV
    ANDROLOGIA, 1988, 20 (05) : 433 - 440
  • [30] COMPARATIVE-STUDY OF 4 NEW AUTOMATED DONOR PLASMAPHERESIS SYSTEMS
    COFFE, C
    COUDURIER, N
    LEVY, G
    PUJOL, JP
    COUTERET, Y
    MASSE, M
    LAMY, B
    KIEFFER, Y
    HERVE, P
    PETERS, A
    PLASMA THERAPY & TRANSFUSION TECHNOLOGY, 1986, 7 (01): : 49 - 55