DIAGNOSTIC PERFORMANCE - A COMPARATIVE-STUDY OF THE LEUKOCYTE DIFFERENTIAL COUNT ON 4 AUTOMATED HEMATOLOGY ANALYZERS

被引:0
|
作者
BUTTARELLO, M
LORENZ, C
GADOTTI, M
TOFFALORI, E
VALENTINI, A
RIZZOTTI, P
机构
[1] OSPED GERIATR,ANAL CHIMICOCLIN & MICROBIOL LAB,VIA VENDRAMINI 7,I-35100 PADUA,ITALY
[2] OSPED SANTA CHIARA,BIOCHIM & EMATOL LAB,TRENT,ITALY
[3] OSPED SANTA CHIARA,SERV FIS SANIT,TRENT,ITALY
关键词
D O I
暂无
中图分类号
Q5 [生物化学]; Q7 [分子生物学];
学科分类号
071010 ; 081704 ;
摘要
The diagnostic performance of the haematology analysers, Technicon H1, Sysmex NE 8000, Coulter STKS and Sequoia Cell Dyn 3000 was determined by comparing their results for automated differential counting with the result obtained for 400 cells by the manual microscopic method. These comparative studies were performed on a group of adults, containing both normal individuals (n = 150) and those affected by various pathologies (n = 113). The number of morphologically false negatives is low for all the systems (from 1.8% for the Cell Dyn 3000 to 6.2% for the NE 8000), while the number of distributionally false negatives was slightly higher (from 4.4% for the Cell Dyn 3000 to 7% for the H1 and NE 8000). The morphological flags, though useful for improving the diagnostic performance of the instruments, show a rather modest sensitivity when taken individually: immature granulocytes/bands from 71% for the STKS to 43% for the NE 8000; blasts from 66.6% for the H1 to 53.3% for the STKS, atypical lymphocytes from 59% for the H1 to 13.6% for the NE 8000; erythroblasts from 42.8% for the STKS to 7% for the NE 8000.
引用
收藏
页码:251 / 258
页数:8
相关论文
共 50 条
  • [1] EVALUATION OF 4 AUTOMATED HEMATOLOGY ANALYZERS - A COMPARATIVE-STUDY OF DIFFERENTIAL COUNTS (IMPRECISION AND INACCURACY)
    BUTTARELLO, M
    GADOTTI, M
    LORENZ, C
    TOFFALORI, E
    CESCHINI, N
    VALENTINI, A
    RIZZOTTI, P
    AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 1992, 97 (03) : 345 - 352
  • [2] LEUKOCYTE DIFFERENTIAL IN HEMATOLOGY ANALYZERS ACCORDING TO WHITE BLOOD CELL COUNT.
    Bielawska, Lena
    Kasprzak, Grazyna
    Urbanska, Natalia
    Lewandowski, Krzysztof
    Haertle, Danuta
    Wysocka, Ewa
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2018, 40 : 18 - 18
  • [3] A comparative study of blood cell count in four automated hematology analyzers: An evaluation of the impact of preanalytical factors
    Astrand, Annika
    Wingren, Cecilia
    Walton, Claire
    Mattsson, Johan
    Agrawal, Komal
    Lindqvist, Madelene
    Odqvist, Lina
    Burmeister, Bjorn
    Eck, Steven
    Hughes, Glen
    Luporini Saraiva, Gabriela
    Schantz, Anna
    Psallidas, Ioannis
    McCrae, Christopher
    PLOS ONE, 2024, 19 (05):
  • [4] Automated hematology analysers and spurious counts Part 2. Leukocyte count and differential
    Genevieve, Franck
    Godon, Alban
    Marteau-Tessier, Anne
    Zandecki, Marc
    ANNALES DE BIOLOGIE CLINIQUE, 2012, 70 (02) : 141 - 154
  • [5] A PARALLEL EVALUATION OF 4 AUTOMATED HEMATOLOGY ANALYZERS
    BENTLEY, SA
    JOHNSON, A
    BISHOP, CA
    AMERICAN JOURNAL OF CLINICAL PATHOLOGY, 1993, 100 (06) : 626 - 632
  • [6] COMPARISON BETWEEN 4 AUTOMATED SYSTEMS FOR LEUKOCYTE DIFFERENTIAL COUNT
    JOU, JM
    VIVESCORRONS, JL
    AGUILAR, JL
    INSA, MJ
    ESTER, A
    DELOLMO, J
    COLOMER, E
    JOURNAL OF AUTOMATIC CHEMISTRY, 1982, 4 (02): : 100 - 101
  • [7] COMPARATIVE STUDY OF FOUR HEMATOLOGY ANALYZERS
    Dave, Deepshikha
    Pandya, Amrish N.
    JOURNAL OF EVOLUTION OF MEDICAL AND DENTAL SCIENCES-JEMDS, 2014, 3 (12): : 3107 - 3113
  • [8] Lamellar body count in amniotic fluid: A comparative study of four different hematology analyzers
    Szallasi, A
    Gronowski, AM
    Eby, CS
    CLINICAL CHEMISTRY, 2003, 49 (06) : 994 - 997
  • [9] AUTOMATED HEMATOLOGY ANALYZERS - A 4-WAY COMPARISON
    WARNER, BA
    REARDON, DM
    MARSHALL, DP
    MEDICAL LABORATORY SCIENCES, 1990, 47 (04): : 285 - 296
  • [10] Leukocyte fragments may interfere in the fluorescent platelet count of Sysmex XN hematology analyzers
    Hummel, Kornelia
    van Dun, Ludi P. J. M.
    Sachse, Michaela
    Beeger, Cord
    Hoffmann, Johannes J. M. L.
    INTERNATIONAL JOURNAL OF LABORATORY HEMATOLOGY, 2020, 42 (04) : E167 - E169