CLINICAL AND HEMODYNAMIC PERFORMANCE OF THE 19-MM CARPENTIER-EDWARDS PORCINE BIOPROSTHESIS

被引:18
|
作者
BOJAR, RM
RASTEGAR, H
PAYNE, DD
MACK, CA
SCHWARTZ, SL
机构
[1] TUFTS UNIV NEW ENGLAND MED CTR,DEPT CARDIOL,BOSTON,MA 02111
[2] TUFTS UNIV,SCH MED,BOSTON,MA 02111
来源
ANNALS OF THORACIC SURGERY | 1993年 / 56卷 / 05期
关键词
D O I
10.1016/0003-4975(95)90032-2
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Because of concerns about the hemodynamic performance of 19-mm porcine valves, we retrospectively reviewed the clinical results and echocardiographic studies of 52 consecutive patients who received a 19-mm Carpentier-Edwards porcine bioprosthesis (model 2625) for aortic valve replacement from 1986 through 1991. Nearly 87% of the patients were women, the mean age was 69 years, and the mean body surface area was 1.63 +/- 0.27 m2. Seventy-three percent of the patients had pure aortic stenosis, 96% were in New York Heart Association classes III and IV, and 56% underwent urgent or emergent operation. Overall hospital mortality was 7.7% with a late mortality of 8.3% at a mean follow-up of 25 +/- 18 months. No patient experienced a valve-related complication, and 95% of surviving patients were in New York Heart Association classes I and II. Two-dimensional and Doppler echocardiography performed during the first postoperative week revealed a maximal instantaneous gradient of 44.7 +/- 13.0 mm Hg. In 43 patients for whom additional data were available, the mean gradient was 26.4 +/- 8.2 mm Hg with an effective orifice area of 0.85 +/- 0.18 cm2. This study defines the normal range of postoperative gradients across the 19-mm Carpentier-Edwards porcine valve and demonstrates that patients receiving this valve can achieve significant clinical improvement despite the presence of high transvalvular gradients measured by echocardiography.
引用
收藏
页码:1141 / 1147
页数:7
相关论文
共 50 条
  • [31] Carpentier-Edwards supraannular porcine bioprosthesis evaluation over 15 years
    Jamieson, WRE
    Ling, HT
    Burr, LH
    Fradet, GJ
    Miyagishima, RT
    Janusz, MT
    Lichtenstein, SV
    ANNALS OF THORACIC SURGERY, 1998, 66 (06): : S49 - S52
  • [32] Midterm results of aortic valve replacement with Carpentier-Edwards porcine bioprosthesis
    Sakata J.
    Tsukamoto M.
    Kuwaki K.
    Komatsu K.
    Harada N.
    Abe T.
    Journal of Artificial Organs, 2002, 5 (4) : 246 - 250
  • [33] CARPENTIER-EDWARDS SUPRAANNULAR PORCINE BIOPROSTHESIS - NEW GENERATION BIOLOGICAL VALVE
    JAMIESON, WRE
    GEREIN, AN
    TYERS, GFO
    ALLEN, P
    MUNRO, AI
    MIYAGISHIMA, RT
    JANUSZ, MT
    BURR, LH
    ZEITSCHRIFT FUR KARDIOLOGIE, 1986, 75 : 245 - 250
  • [34] DURABILITY OF THE CARPENTIER-EDWARDS PORCINE BIOPROSTHESIS - ROLE OF AGE AND VALVE POSITION
    BERNAL, JM
    RABASA, JM
    LOPEZ, R
    NISTAL, JF
    MUNIZ, R
    REVUELTA, JM
    ANNALS OF THORACIC SURGERY, 1995, 60 (02): : S248 - S252
  • [35] Carpentier-Edwards standard porcine bioprosthesis: A 21-year experience
    Jamieson, WRE
    Burr, LH
    Munro, AI
    Miyagishima, RT
    ANNALS OF THORACIC SURGERY, 1998, 66 (06): : S40 - S43
  • [36] HEMODYNAMIC AND CLINICAL EVALUATION OF CARPENTIER-EDWARDS PORCINE HETEROGRAFT AORTIC-VALVE
    ISHIMORI, T
    DEBOER, LW
    RAIZNER, AE
    GUINN, GA
    REA, JE
    JAMAL, NM
    BURNETT, WC
    NELSON, DB
    CHAHINE, RA
    MILLER, RR
    CLINICAL RESEARCH, 1978, 26 (03): : A240 - A240
  • [37] USE OF THE CARPENTIER-EDWARDS PORCINE BIOPROSTHESIS - ASSESSMENT OF A PATIENT SELECTION POLICY
    LOUAGIE, Y
    NOIRHOMME, P
    ARANGUIS, E
    EUCHER, P
    VANRUYSSEVELT, P
    BUCHE, M
    DION, R
    JAUMIN, P
    SCHOEVAERDTS, JC
    CHALANT, CH
    JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 1992, 104 (04): : 1013 - 1024
  • [38] THE CARPENTIER-EDWARDS SUPRAANNULAR PORCINE BIOPROSTHESIS - A NEW GENERATION TISSUE VALVE WITH EXCELLENT INTERMEDIATE CLINICAL-PERFORMANCE
    JAMIESON, WRE
    MUNRO, AI
    MIYAGISHIMA, RT
    BURR, LH
    GEREIN, AN
    JANUSZ, MT
    TYERS, GFO
    ALLEN, P
    JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 1988, 96 (04): : 652 - 666
  • [39] Late hemodynamic and clinical outcomes of aortic valve replacement with the Carpentier-Edwards Perimount pericardial bioprosthesis
    Dellgren, G
    David, TE
    Raanani, E
    Armstrong, S
    Ivanov, J
    Rakowski, H
    JOURNAL OF THORACIC AND CARDIOVASCULAR SURGERY, 2002, 124 (01): : 146 - 154
  • [40] Clinical experience with the Carpentier-Edwards supra-annular porcine bioprosthesis implanted in the aortic position
    Hurle, A
    Meseguer, J
    Llamas, P
    Casillas, JA
    JOURNAL OF HEART VALVE DISEASE, 1998, 7 (03): : 331 - 335