Midterm results of aortic valve replacement with Carpentier-Edwards porcine bioprosthesis

被引:0
|
作者
Sakata J. [1 ]
Tsukamoto M. [1 ]
Kuwaki K. [1 ]
Komatsu K. [1 ]
Harada N. [1 ]
Abe T. [1 ]
机构
[1] Department of Cardiovascular Surgery, Sapporo Med. Univ. Sch. of Medicine, Sapporo 060-8543, South 1, West 16, Chuo-ku
关键词
Bioprosthesis; Carpentier-Edwards valve; Midterm results;
D O I
10.1007/s100470200046
中图分类号
学科分类号
摘要
Since 1987 we have been using Carpentier-Edwards porcine bioprostheses for aortic valve replacements in patients of advanced age and those with a contraindication for anticoagulation therapy. The purpose of this study was to provide insight into the midterm results of our series of patients with Carpentier-Edwards porcine bioprostheses in the aortic position. From September 1987 through June 1999, 31 patients underwent aortic valve replacement with Capentier-Edwards porcine bioprostheses. The age of the patients ranged from 21 to 89 (average, 71 ± 11 years old), and 21 patients (67.7%) were 70 years old or older. There were no operative deaths. Three hospital deaths (9.7%) and one nonhospital death (3.2%) took place. The actuarial survival rate of 31 patients was 90.3% at 5 years and 81.3% at 7 years. Valve-related complications were seen in six patients (19.4%). The postoperative Doppler peak gradient of aortic prostheses was less than 30 mm Hg in all groups, irrespective of valve size. In conclusion, the Carpentier-Edwards porcine bioprosthesis affords good durability with a low rate of valve-related complications and reoperation. Furthermore, the postoperative function of the prosthesis was reasonable in the Doppler echo studies. © The Japanese Society for Artificial Organs 2002.
引用
收藏
页码:246 / 250
页数:4
相关论文
共 50 条
  • [1] Carpentier-Edwards supraannular porcine bioprosthesis: Second-generation prosthesis in aortic valve replacement
    Jamieson, WRE
    Janusz, MT
    Burr, LH
    Ling, H
    Miyagishima, RT
    Germann, E
    ANNALS OF THORACIC SURGERY, 2001, 71 (05): : S224 - S227
  • [2] LATE RESULTS WITH CARPENTIER-EDWARDS PORCINE BIOPROSTHESIS
    AKINS, CW
    CARROLL, DL
    BUCKLEY, MJ
    DAGGETT, WM
    HILGENBERG, AD
    AUSTEN, WG
    CIRCULATION, 1990, 82 (05) : 65 - 74
  • [3] Aortic and mitral valve replacement with the Carpentier-Edwards pericardial bioprosthesis: 10-year results
    Murakami, T
    Eishi, K
    Nakano, S
    Kobayashi, J
    Sasako, Y
    Isobe, F
    Kosakai, Y
    Kito, Y
    Kawashima, Y
    JOURNAL OF HEART VALVE DISEASE, 1996, 5 (01): : 45 - 49
  • [4] HEMODYNAMICS OF THE HANCOCK PERICARDIAL AND THE CARPENTIER-EDWARDS AORTIC PORCINE BIOPROSTHESIS
    LEROY, G
    DESNOS, M
    BELFANTE, M
    GAIDE, L
    GAY, J
    DELOCHE, A
    GERBAUX, A
    CIRCULATION, 1983, 68 (04) : 343 - 343
  • [5] INTERMEDIATE CLINICAL-RESULTS AFTER AORTIC-VALVE REPLACEMENT WITH THE CARPENTIER-EDWARDS PERICARDIAL BIOPROSTHESIS
    TORKA, MC
    SALEFSKY, BE
    HACKER, RW
    ANNALS OF THORACIC SURGERY, 1995, 60 (02): : S311 - S315
  • [6] RISK OF THROMBOEMBOLISM EARLY FOLLOWING AORTIC-VALVE REPLACEMENT WITH THE CARPENTIER-EDWARDS BIOPROSTHESIS
    ORSZULAK, TA
    SCHAFF, HV
    MULLANY, CJ
    PUGA, FJ
    DANIELSON, GK
    CIRCULATION, 1992, 86 (04) : 497 - 497
  • [7] EVALUATION OF A NEW AORTIC BIOPROSTHESIS - THE CARPENTIER-EDWARDS PERICARDIAL VALVE
    PELLETIER, LC
    CARTIER, R
    BONAN, R
    CREPEAU, J
    DYRDA, I
    CLINICAL AND INVESTIGATIVE MEDICINE-MEDECINE CLINIQUE ET EXPERIMENTALE, 1986, 9 (03): : B36 - B36
  • [8] CLINICAL AND HEMODYNAMIC-RESULTS WITH THE CARPENTIER-EDWARDS PORCINE BIOPROSTHESIS
    PELLETIER, C
    CHAITMAN, BR
    BAILLOT, R
    VAL, PG
    BONAN, R
    DYRDA, I
    ANNALS OF THORACIC SURGERY, 1982, 34 (06): : 612 - 624
  • [9] CARPENTIER-EDWARDS SUPRAANNULAR PORCINE BIOPROSTHESIS - NEW GENERATION BIOLOGICAL VALVE
    JAMIESON, WRE
    GEREIN, AN
    TYERS, GFO
    ALLEN, P
    MUNRO, AI
    MIYAGISHIMA, RT
    JANUSZ, MT
    BURR, LH
    ZEITSCHRIFT FUR KARDIOLOGIE, 1986, 75 : 245 - 250
  • [10] DURABILITY OF THE CARPENTIER-EDWARDS PORCINE BIOPROSTHESIS - ROLE OF AGE AND VALVE POSITION
    BERNAL, JM
    RABASA, JM
    LOPEZ, R
    NISTAL, JF
    MUNIZ, R
    REVUELTA, JM
    ANNALS OF THORACIC SURGERY, 1995, 60 (02): : S248 - S252