This is a theoretical study of the inherent spatial resolution of the latent image on the surface of an amorphous selenium (a-Se) plate used for diagnostic x-ray imaging. The following effects are considered: (A) ranges of primary photoelectrons; (B) reabsorption of K fluorescence; (C) reabsorption of Compton scattered photons; (D) diffusion; (E) the geometric effect due to oblique incidence of x rays; (F) electrostatic effect; and (G) the space charge effect. The modulation transfer function of a-Se in the diagnostic x-ray energy range has been estimated. In conclusion, (A) and (E) are the main factors limiting the resolution, and for diagnostic x rays, the inherent spatial resolution of 2-Se plates is much better than that of Csl layers used in x-ray image intensifiers. © 1995, American Association of Physicists in Medicine. All rights reserved.
引用
收藏
页码:365 / 374
页数:10
相关论文
共 50 条
  • [21] THE SPATIAL-RESOLUTION OF X-RAY-MICROANALYSIS EXPERIMENTAL ASPECTS
    ZELECHOWER, M
    INSTITUTE OF PHYSICS CONFERENCE SERIES, 1993, (130): : 157 - 160
  • [22] HARD X-RAY MICROSCOPE WITH SUBMICROMETER SPATIAL-RESOLUTION
    KURIYAMA, M
    DOBBYN, RC
    SPAL, RD
    BURDETTE, HE
    BLACK, DR
    JOURNAL OF RESEARCH OF THE NATIONAL INSTITUTE OF STANDARDS AND TECHNOLOGY, 1990, 95 (05) : 559 - 574
  • [23] HIGH SPATIAL-RESOLUTION IN STEM X-RAY-MICROANALYSIS
    THOMAS, LE
    ULTRAMICROSCOPY, 1982, 9 (03) : 311 - 318
  • [24] DIGITAL X-RAY-IMAGING
    MOORES, BM
    IEE PROCEEDINGS-A-SCIENCE MEASUREMENT AND TECHNOLOGY, 1987, 134 (02): : 115 - 125
  • [25] HIGH-SPATIAL-RESOLUTION SCINTILLATOR SCREENS COUPLED TO CCD DETECTORS FOR X-RAY-IMAGING APPLICATIONS
    CASTELLI, CM
    ALLINSON, NM
    MOON, KJ
    WATSON, DL
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1994, 348 (2-3): : 649 - 653
  • [26] X-RAY-IMAGING TUBES
    SCHAGEN, P
    NDT INTERNATIONAL, 1981, 14 (01): : 9 - 14
  • [27] CONVENTIONAL X-RAY-IMAGING
    GARRETT, JA
    SMITHSON, PH
    IEE PROCEEDINGS-A-SCIENCE MEASUREMENT AND TECHNOLOGY, 1987, 134 (02): : 107 - 114
  • [28] IMPROVEMENT OF SPATIAL-RESOLUTION OF MONOCHROMATIC X-RAY CT USING SYNCHROTRON RADIATION
    SAKAMOTO, K
    SUZUKI, Y
    HIRANO, T
    USAMI, K
    JAPANESE JOURNAL OF APPLIED PHYSICS PART 1-REGULAR PAPERS SHORT NOTES & REVIEW PAPERS, 1988, 27 (01): : 127 - 132
  • [29] CHARACTERIZATION OF NOVA PLASMAS USING AN X-RAY SPECTROMETER WITH TEMPORAL-RESOLUTION AND SPATIAL-RESOLUTION
    BACK, CA
    KAUFFMAN, RL
    BELL, PM
    KILKENNY, JD
    REVIEW OF SCIENTIFIC INSTRUMENTS, 1995, 66 (01): : 764 - 766
  • [30] X-RAY MICROSCOPY AND X-RAY-IMAGING
    BURGE, RE
    MICHETTE, AG
    DUKE, PJ
    SCANNING MICROSCOPY, 1987, 1 (03) : 891 - 900