TIDAL VOLUME MEASUREMENTS IN NEWBORNS USING RESPIRATORY INDUCTIVE PLETHYSMOGRAPHY

被引:60
|
作者
ADAMS, JA
ZABALETA, IA
STROH, D
JOHNSON, P
SACKNER, MA
机构
[1] MT SINAI MED CTR,DIV PULM DIS,MIAMI BEACH,FL 33140
[2] FLORIDA INT UNIV,DEPT STAT,MIAMI,FL 33199
来源
AMERICAN REVIEW OF RESPIRATORY DISEASE | 1993年 / 148卷 / 03期
关键词
D O I
10.1164/ajrccm/148.3.585
中图分类号
R56 [呼吸系及胸部疾病];
学科分类号
摘要
Respiratory inductive plethysmography (RIP) is a well-accepted noninvasive technology for monitoring breathing patterns in adults. Prior attempts to calibrate this device in babies have been fraught with technical difficulties, thereby limiting applications in this population. Recently, a new method, qualitative diagnostic calibration (QDC), has been shown to provide accurate calibration of tidal volume in adults. The QDC method is based upon principles of the isovolume maneuver and carried out during natural breathing without specialized respiratory maneuvers or postural changes. We calibrated RIP with QDC in the supine posture and compared tidal volume (VT) measured with RIP to VT by a face mask-pneumotachograph (PNT) in 21 healthy full-term newborns in supine and prone postures. Eleven of the babies were calibrated during active sleep and 10 in quiet sleep. The mean VT in the supine and prone postures were 19 and 25 ml, respectively. In the supine and prone postures, weighted mean difference between RIP (VT) and PNT (VT) and 95% confidence intervals were -0.05 ml (-0.27, 0.18) and -0.32 ml (-0.08, 0.55), respectively. There was no difference in the accuracy of RIP relative to PNT calibrated during active sleep when thoracoabdominal incoordination was present or quiet sleep when it was not in either the supine or the prone postures. Therefore, in full-term infants, RIP calibrated with QDC solely in the supine posture provides clinically acceptable measurements of VT in both supine and prone postures.
引用
收藏
页码:585 / 588
页数:4
相关论文
共 50 条
  • [31] Lung volume recruitment in lambs during high-frequency oscillatory ventilation using respiratory inductive plethysmography
    Göthberg, S
    Parker, TA
    Griebel, J
    Abman, SH
    Kinsella, JP
    PEDIATRIC RESEARCH, 2001, 49 (01) : 38 - 44
  • [32] Tidal Volume Measurements in Infants: Opto-Electronic Plethysmography Versus Pneumotachograph
    Albuquerque Reinaux, Cyda Maria
    Aliverti, Andrea
    Melo da Silva, Livia Gabriely
    da Silva, Rafael Justino
    Goncalves, Juliane Neves
    Noronha, Jessica Brito
    Cabral Filho, Jose Eulalio
    de Andrade, Armele Dornelas
    de Amorim Britto, Murilo Carlos
    PEDIATRIC PULMONOLOGY, 2016, 51 (08) : 850 - 857
  • [33] ACCURACY OF RESPIRATORY INDUCTIVE PLETHYSMOGRAPH IN MEASURING TIDAL VOLUME DURING SLEEP
    WHYTE, KF
    GUGGER, M
    GOULD, GA
    MOLLOY, J
    WRAITH, PK
    DOUGLAS, NJ
    JOURNAL OF APPLIED PHYSIOLOGY, 1991, 71 (05) : 1866 - 1871
  • [34] Chest wall motion in preterm infants using respiratory inductive plethysmography
    Warren, RH
    Horan, SM
    Robertson, PK
    EUROPEAN RESPIRATORY JOURNAL, 1997, 10 (10) : 2295 - 2300
  • [35] Validation of respiratory inductive plethysmography using QDC calibration in anaesthetized infants
    Aun, C
    Brown, K
    Jackson, E
    Mackersie, A
    Stocks, J
    Hatch, D
    ANESTHESIOLOGY, 1997, 87 (03) : A1041 - A1041
  • [36] Using respiratory inductive plethysmography for monitoring professional reed instrument performance
    Fuks, L
    Sandberg, J
    MEDICAL PROBLEMS OF PERFORMING ARTISTS, 1999, 14 (01) : 30 - 42
  • [37] NEW TRENDS IN THE USE OF RESPIRATORY INDUCTIVE PLETHYSMOGRAPHY
    DALLAVASANTUCCI, J
    BIOLOGICAL PSYCHOLOGY, 1993, 35 (03) : 269 - 269
  • [38] CALIBRATION OF RESPIRATORY INDUCTIVE PLETHYSMOGRAPHY (RESPICOMPTM) IN INFANTS
    WARREN, RH
    ALDERSON, SH
    CLINICAL RESEARCH, 1984, 32 (05): : A909 - A909
  • [39] CALIBRATION OF RESPIRATORY INDUCTIVE PLETHYSMOGRAPHY (RESPICOMP) IN LAMBS
    WARREN, RH
    FEWELL, JE
    ALDERSON, SH
    CLINICAL RESEARCH, 1983, 31 (05): : A917 - A917
  • [40] COMPARISON OF PIEZOELECTRODE BELTS TO RESPIRATORY INDUCTIVE PLETHYSMOGRAPHY
    Foreman, E. B.
    Underwood, W. W.
    Vaughn, B. V.
    SLEEP, 2009, 32 : A373 - A373