Comparison of the hypothetical Co-57 brachytherapy source with the Ir-192 source

被引:3
|
作者
Toossi, Mohammad Taghi Bahreyni [1 ]
Ghorbani, Mahdi [1 ]
Rostami, Atefeh [1 ]
Khosroabadi, Mohsen [2 ]
Khademi, Sara [3 ]
Knaup, Courtney [4 ]
机构
[1] Shahid Beheshti Univ Med Sci, Biomed Engn & Med Phys Dept, Fac Med, Tehran, Iran
[2] North Khorasan Univ Med Sci, Fac Med, Bojnurd, Iran
[3] Univ Tehran Med Sci, Med Phys & Med Engn Dept, Fac Med, Tehran, Iran
[4] Comprehens Canc Ctr Nevada, Las Vegas, NV USA
来源
关键词
hypothetical source; Co-57; Ir-192; TG-43 dosimetric parameters; Monte Carlo simulation;
D O I
10.5114/wo.2016.61854
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Aim of the study: The Co-57 radioisotope has recently been proposed as a hypothetical brachytherapy source due to its high specific activity, appropriate half-life (272 days) and medium energy photons (114.17 keV on average). In this study, Task Group No. 43 dosimetric parameters were calculated and reported for a hypothetical Co-57 source. Material and methods: A hypothetical Co-57 source was simulated in MCNPX, consisting of an active cylinder with 3.5 mm length and 0.6 mm radius encapsulated in a stainless steel capsule. Three photon energies were utilized (136 keV [10.68%], 122 keV [85.60%], 14 keV [9.16%]) for the Co-57 source. Air kerma strength, dose rate constant, radial dose function, anisotropy function, and isodose curves for the source were calculated and compared to the corresponding data for a Ir-192 source. Results: The results are presented as tables and figures. Air kerma strength per 1 mCi activity for the Co-57 source was 0.46 cGyh(-1) cm 2 mCi(-1). The dose rate constant for the Co-57 source was determined to be 1.215 cGyh(-1)U(-1). The radial dose function for the Co-57 source has an increasing trend due to multiple scattering of low energy photons. The anisotropy function for the Co-57 source at various distances from the source is more isotropic than the Ir-192 source. Conclusions: The Co-57 source has advantages over Ir-192 due to its lower energy photons, longer half-life, higher dose rate constant and more isotropic anisotropic function. However, the Ir-192 source has a higher initial air kerma strength and more uniform radial dose function. These properties make Co-57 a suitable source for use in brachytherapy applications.
引用
收藏
页码:327 / 334
页数:8
相关论文
共 50 条
  • [31] Measurement of dosimetric parameters for a new HDR Ir-192 source
    Muller-Runkel, R
    MEDICAL PHYSICS, 2004, 31 (06) : 1900 - 1900
  • [32] IR-192 INTERSTITIAL BRACHYTHERAPY OF EQUINE SARCOID
    TURREL, JM
    STOVER, SM
    GYORGYFALVY, J
    VETERINARY RADIOLOGY, 1985, 26 (01): : 20 - 24
  • [33] IR-192 BRACHYTHERAPY IN THE TREATMENT OF CANCER OF THE PROSTATE
    DONNELLY, BJ
    PEDERSEN, JE
    PORTER, AT
    MCPHEE, MS
    UROLOGIC CLINICS OF NORTH AMERICA, 1991, 18 (03) : 481 - 483
  • [34] IR-192 BRACHYTHERAPY SOURCES IN BELGIAN HOSPITALS
    SCHAEKEN, B
    VANNESTE, F
    BOUILLER, A
    HOORNAERT, MT
    VANDENBROECK, S
    HERMANS, J
    PIRON, A
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1992, 312 (1-2): : 251 - 256
  • [35] Effect of inhomogeneities and source position on dose distribution of nucletron high dose rate Ir-192 brachytherapy source by Monte Carlo simulation
    Chandola, R. M.
    Tiwari, S.
    Kowar, M. K.
    Choudhary, V.
    JOURNAL OF CANCER RESEARCH AND THERAPEUTICS, 2010, 6 (01) : 54 - 57
  • [36] Dosimetric comparison between the microSelectron HDR Ir-192 v2 source and the BEBIG Co-60 source for HDR brachytherapy using the EGSnrc Monte Carlo transport code
    Islam, M. Anwarul
    Akramuzzaman, M. M.
    Zakaria, G. A.
    JOURNAL OF MEDICAL PHYSICS, 2012, 37 (04) : 219 - 225
  • [37] Trials of transmission imaging using clinically used Ir-192 source for high-dose-rate brachytherapy
    Nagata, J.
    Yamamoto, S.
    Nakanishi, K.
    Noguchi, Y.
    Okudaira, K.
    Kataoka, J.
    JOURNAL OF INSTRUMENTATION, 2022, 17 (06)
  • [38] The non-uniformity correction factor for the cylindrical ionization chambers in dosimetry of an HDR Ir-192 brachytherapy source
    Majumdar, Bishnu
    Patel, Narayan Prasad
    Vijayan, V.
    JOURNAL OF MEDICAL PHYSICS, 2006, 31 (02) : 83 - 88
  • [39] Imaging of Ir-192 source using a high energy gamma camera for high-dose-rate brachytherapy
    Yamamoto, Seiichi
    Yogo, Katsunori
    Noguchi, Yumiko
    Nakaya, Takayoshi
    Okudaira, Kuniyasu
    RADIATION MEASUREMENTS, 2019, 126
  • [40] QUALITY ASSURANCE FOR CLINICAL PRACTICE IN CERVICAL CANCER BRACHYTHERAPY WITH IR-192 SOURCE: IN VIVO DOSIMETRY WITH QED DIODES
    Mare, Valerio
    Carlotti, Daniele
    Fiore, Michele
    Vicenzi, Lisa
    Guarnaccia, Roberta
    Matteucci, Paolo
    Iurato, Aurelia
    Petrianni, Gian Marco
    Cirnigliaro, Vera
    Moraschi, Marta
    Palmese, Valentina Pirozzi
    Martucci, Paola
    Alaimo, Rita
    Ippolito, Edy
    Ramella, Sara
    Insero, Teresa
    INTERNATIONAL JOURNAL OF GYNECOLOGICAL CANCER, 2023, 33 : A102 - A102