Comparison of the hypothetical Co-57 brachytherapy source with the Ir-192 source

被引:3
|
作者
Toossi, Mohammad Taghi Bahreyni [1 ]
Ghorbani, Mahdi [1 ]
Rostami, Atefeh [1 ]
Khosroabadi, Mohsen [2 ]
Khademi, Sara [3 ]
Knaup, Courtney [4 ]
机构
[1] Shahid Beheshti Univ Med Sci, Biomed Engn & Med Phys Dept, Fac Med, Tehran, Iran
[2] North Khorasan Univ Med Sci, Fac Med, Bojnurd, Iran
[3] Univ Tehran Med Sci, Med Phys & Med Engn Dept, Fac Med, Tehran, Iran
[4] Comprehens Canc Ctr Nevada, Las Vegas, NV USA
来源
关键词
hypothetical source; Co-57; Ir-192; TG-43 dosimetric parameters; Monte Carlo simulation;
D O I
10.5114/wo.2016.61854
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Aim of the study: The Co-57 radioisotope has recently been proposed as a hypothetical brachytherapy source due to its high specific activity, appropriate half-life (272 days) and medium energy photons (114.17 keV on average). In this study, Task Group No. 43 dosimetric parameters were calculated and reported for a hypothetical Co-57 source. Material and methods: A hypothetical Co-57 source was simulated in MCNPX, consisting of an active cylinder with 3.5 mm length and 0.6 mm radius encapsulated in a stainless steel capsule. Three photon energies were utilized (136 keV [10.68%], 122 keV [85.60%], 14 keV [9.16%]) for the Co-57 source. Air kerma strength, dose rate constant, radial dose function, anisotropy function, and isodose curves for the source were calculated and compared to the corresponding data for a Ir-192 source. Results: The results are presented as tables and figures. Air kerma strength per 1 mCi activity for the Co-57 source was 0.46 cGyh(-1) cm 2 mCi(-1). The dose rate constant for the Co-57 source was determined to be 1.215 cGyh(-1)U(-1). The radial dose function for the Co-57 source has an increasing trend due to multiple scattering of low energy photons. The anisotropy function for the Co-57 source at various distances from the source is more isotropic than the Ir-192 source. Conclusions: The Co-57 source has advantages over Ir-192 due to its lower energy photons, longer half-life, higher dose rate constant and more isotropic anisotropic function. However, the Ir-192 source has a higher initial air kerma strength and more uniform radial dose function. These properties make Co-57 a suitable source for use in brachytherapy applications.
引用
收藏
页码:327 / 334
页数:8
相关论文
共 50 条
  • [21] Comparative dosimetry of GammaMed Plus high-dose rate Ir-192 brachytherapy source
    Patel, N. P.
    Majumdar, B.
    Vijayan, V.
    JOURNAL OF MEDICAL PHYSICS, 2010, 35 (03) : 137 - 143
  • [22] PTFE AS A SOURCE GUIDE IN IR-192 BREAST IMPLANTS
    HUTCHINGS, RE
    BRITISH JOURNAL OF RADIOLOGY, 1985, 58 (689): : 456 - 458
  • [23] In-Phantom Calibration of High Dose Rate Ir-192 Brachytherapy Source - A Feasibility Study
    Mandiram, Kuppuswamy
    Senthil, G.
    Sundaram, Visalatchi
    Muthulingam, S.
    Karthik, S.
    MEDICAL PHYSICS, 2009, 36 (06)
  • [24] Dose calculation in brachytherapy for a Ir-192 source using a primary and scatter dose separation technique
    Russell, KR
    Ahnesjo, A
    PHYSICS IN MEDICINE AND BIOLOGY, 1996, 41 (06): : 1007 - 1024
  • [25] A simple program for calculated implant time verification in Ir-192 line source interstitial brachytherapy
    Bejarano, MAB
    Castillo, GA
    Ceballos, JLO
    de la Higuera, JABJ
    RADIOTHERAPY AND ONCOLOGY, 2003, 68 : S85 - S85
  • [26] Comparison of brachytherapy sources of endometrial cancer: Electronic brachytherapy source and 192Ir
    Lozares, S.
    Font Gomez, J. A.
    Gandia Martinez, A.
    Miranda Burgos, A.
    Mendez Villamon, A.
    RADIOTHERAPY AND ONCOLOGY, 2017, 123 : S519 - S520
  • [27] Determination of Changes in Dosimetric Parameters with Co-60 and Ir-192 Source in High Dose Rate Brachytherapy of Cervical Carcinoma
    Srivastava, S.
    Moirangthem, N.
    Singh, N.
    MEDICAL PHYSICS, 2021, 48 (06)
  • [28] Impact of different Ir-192 source models on dose calculations in high-dose-rate brachytherapy
    Otani, Yuki
    Shimamoto, Shigetoshi
    Sumida, Iori
    Takahashi, Yutaka
    Tani, Shoji
    Oshima, Tetsuya
    Onosaka, Satoshi
    Isohashi, Fumiaki
    Tamari, Keisuke
    Ogawa, Kazuhiko
    PHYSICS & IMAGING IN RADIATION ONCOLOGY, 2018, 7 : 23 - 26
  • [29] TLD, diode and monte carlo dosimetry of an Ir-192 source for high dose-rate brachytherapy
    Kirov, AS
    Williamson, JF
    Meigooni, AS
    Zhu, Y
    PHYSICS IN MEDICINE AND BIOLOGY, 1995, 40 (12): : 2015 - 2036
  • [30] Radiochromic film measurement of anisotropy function for high-dose-rate Ir-192 brachytherapy source
    Sharma, SD
    Bianchi, C
    Conte, L
    Novario, R
    Bhatt, BC
    PHYSICS IN MEDICINE AND BIOLOGY, 2004, 49 (17): : 4065 - 4072