Comparison of the hypothetical Co-57 brachytherapy source with the Ir-192 source

被引:3
|
作者
Toossi, Mohammad Taghi Bahreyni [1 ]
Ghorbani, Mahdi [1 ]
Rostami, Atefeh [1 ]
Khosroabadi, Mohsen [2 ]
Khademi, Sara [3 ]
Knaup, Courtney [4 ]
机构
[1] Shahid Beheshti Univ Med Sci, Biomed Engn & Med Phys Dept, Fac Med, Tehran, Iran
[2] North Khorasan Univ Med Sci, Fac Med, Bojnurd, Iran
[3] Univ Tehran Med Sci, Med Phys & Med Engn Dept, Fac Med, Tehran, Iran
[4] Comprehens Canc Ctr Nevada, Las Vegas, NV USA
来源
关键词
hypothetical source; Co-57; Ir-192; TG-43 dosimetric parameters; Monte Carlo simulation;
D O I
10.5114/wo.2016.61854
中图分类号
R73 [肿瘤学];
学科分类号
100214 ;
摘要
Aim of the study: The Co-57 radioisotope has recently been proposed as a hypothetical brachytherapy source due to its high specific activity, appropriate half-life (272 days) and medium energy photons (114.17 keV on average). In this study, Task Group No. 43 dosimetric parameters were calculated and reported for a hypothetical Co-57 source. Material and methods: A hypothetical Co-57 source was simulated in MCNPX, consisting of an active cylinder with 3.5 mm length and 0.6 mm radius encapsulated in a stainless steel capsule. Three photon energies were utilized (136 keV [10.68%], 122 keV [85.60%], 14 keV [9.16%]) for the Co-57 source. Air kerma strength, dose rate constant, radial dose function, anisotropy function, and isodose curves for the source were calculated and compared to the corresponding data for a Ir-192 source. Results: The results are presented as tables and figures. Air kerma strength per 1 mCi activity for the Co-57 source was 0.46 cGyh(-1) cm 2 mCi(-1). The dose rate constant for the Co-57 source was determined to be 1.215 cGyh(-1)U(-1). The radial dose function for the Co-57 source has an increasing trend due to multiple scattering of low energy photons. The anisotropy function for the Co-57 source at various distances from the source is more isotropic than the Ir-192 source. Conclusions: The Co-57 source has advantages over Ir-192 due to its lower energy photons, longer half-life, higher dose rate constant and more isotropic anisotropic function. However, the Ir-192 source has a higher initial air kerma strength and more uniform radial dose function. These properties make Co-57 a suitable source for use in brachytherapy applications.
引用
收藏
页码:327 / 334
页数:8
相关论文
共 50 条
  • [1] Comparison of a Commercial Electronic Brachytherapy Source and Ir-192/Co-60 HDR Source for Endometrial and Cervical Cancers
    Mobit, P.
    Nguyen, A.
    Yang, C.
    MEDICAL PHYSICS, 2013, 40 (06)
  • [2] A DOSIMETRIC STUDY OF THE MICROSELECTRON HDR IR-192 BRACHYTHERAPY SOURCE
    Granero, D.
    Rivard, M.
    Ballester, F.
    Perez-Calatayud, J.
    Vijande, J.
    Otal, A.
    RADIOTHERAPY AND ONCOLOGY, 2010, 96 : S479 - S479
  • [3] A Dosimetric Study of the MicroSelectron HDR Ir-192 Brachytherapy Source
    Granero, D.
    Rivard, M. J.
    Ballester, F.
    Perez-Calatayud, J.
    Vijande, J.
    Otal, A.
    MEDICAL PHYSICS, 2010, 37 (06)
  • [4] Dosimetric Comparison of Co-60 and Ir-192 HDR Source Used in Brachytherapy Treatment of Cervical Cancer
    Shukla, A.
    Singh, N.
    Kumar, S.
    Rana, B.
    Sandhu, I.
    MEDICAL PHYSICS, 2017, 44 (06)
  • [5] Air Cavity Effect on Dose Distribution in Ir-192 Brachytherapy Source
    Osman, Alexander F. I.
    Maalej, Nabil
    Ur-Rehman, Khateeb
    Abdel-Rahman, Wamied
    2015 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2015,
  • [6] Determination of contributions of scatter and distance error to the source strength of Ir-192 HDR brachytherapy source
    BONDEL, Shwetha
    RAVIKUMAR, Manickam
    POLISH JOURNAL OF MEDICAL PHYSICS AND ENGINEERING, 2016, 22 (03) : 55 - 59
  • [7] Monte Carlo characterization of the GammaMed HDR Plus Ir-192 brachytherapy source
    Reyes, E.
    Sosa, M.
    Gil-Villegas, A.
    Monzon, E.
    BIOMEDICAL PHYSICS & ENGINEERING EXPRESS, 2016, 2 (01):
  • [8] Dosimetry of a two new IR-192 source models for HDR and PDR brachytherapy
    Granero, D.
    Perez-Calatayud, J.
    Gimeno, J.
    Gonzalez, V.
    Ballester, F.
    Casal, E.
    Pujades, M.
    RADIOTHERAPY AND ONCOLOGY, 2007, 84 : S147 - S147
  • [9] Transit dose of an Ir-192 high dose rate brachytherapy stepping source
    Wong, TPY
    Fernando, W
    Johnston, PN
    Bubb, IF
    PHYSICS IN MEDICINE AND BIOLOGY, 2001, 46 (02): : 323 - 331
  • [10] CALIBRATION OF A MICROSELECTRON HDR IR-192 SOURCE
    FLYNN, A
    WORKMAN, G
    BRITISH JOURNAL OF RADIOLOGY, 1991, 64 (764): : 734 - 739