RETRIEVAL OF OCCLUDING UNEXPANDED PALMAZ-SCHATZ STENT FROM A SAPHENOUS AORTOCORONARY VEIN GRAFT

被引:42
|
作者
ROZENMAN, Y
BURSTEIN, M
HASIN, Y
GOTSMAN, MS
机构
[1] Department of Cardiology, Hadassah University Hospital, Jerusalem
来源
CATHETERIZATION AND CARDIOVASCULAR DIAGNOSIS | 1995年 / 34卷 / 02期
关键词
PALMAZ-SCHATZ STENT; SAPHENOUS VEIN GRAFT; UNSUCCESSFUL DEPLOYMENT;
D O I
10.1002/ccd.1810340418
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The Palmaz-Schatz coronary stent is used frequently to reduce the rate of restenosis of balloon angioplasty in saphenous vein grafts. In many European centers, the stent, manually crimped on a balloon, is advanced across the stenosis without a protective sheath. This report describes a patient in whom an attempt to deploy a stent in the orifice of a saphenous vein graft was complicated by dislodgement of the unexpanded stent from the balloon. The unexpanded stent caused immediate occlusion of the vein graft and severe ischemic symptoms. The stent was retrieved by inflation of another balloon in the graft, distal to the stent, pulling the balloon toward the guiding catheter, and then withdrawing the whole system. (C) 1995 Wiley-Liss, Inc.
引用
收藏
页码:159 / 161
页数:3
相关论文
共 50 条
  • [31] DIVIDED PALMAZ-SCHATZ(TM) STENT FOR DISCRETE CORONARY STENOSIS
    MEDINA, A
    HERNANDEZ, E
    DELEZO, JS
    PAN, M
    MELIAN, F
    ROMERO, M
    MARRERO, J
    IRURITA, M
    PAVLOVIC, D
    JOURNAL OF INVASIVE CARDIOLOGY, 1992, 4 (08): : 389 - 392
  • [32] Coronary Palmaz-Schatz stent implantation in acute myocardial infarction
    Neumann, FJ
    Walter, H
    Richardt, G
    Schmitt, C
    Schomig, A
    HEART, 1996, 75 (02) : 121 - 126
  • [33] Palmaz-Schatz stent implantation with 10 days ticlopidine treatment
    Kataoka, K
    Sato, Y
    Ito, H
    Takatsu, Y
    CIRCULATION, 1996, 94 (08) : 3997 - 3997
  • [34] INFLUENCE OF EXPANDED BALLOON DIAMETER ON PALMAZ-SCHATZ STENT RECOIL
    RECHAVIA, E
    LITVACK, F
    MACKO, G
    EIGLER, NL
    CATHETERIZATION AND CARDIOVASCULAR DIAGNOSIS, 1995, 36 (01): : 11 - 16
  • [35] Comparison of women treated with Palmaz-Schatz Stent and with balloon angioplasty
    Kennard, ED
    Yeh, W
    Marinac-Dabic, D
    Bright, RA
    Torrence, ME
    Detre, KM
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1998, 31 (02) : 234A - 234A
  • [36] IS CORONARY STENT RECOIL DIFFERENT FOR GIANTURCO-ROUBIN AND PALMAZ-SCHATZ STENT
    FERNANDEZORTIZ, A
    GOICOLEA, J
    PEREZVIZCAYNO, MJ
    ALFONSO, F
    HERNANDEZ, R
    BANUELOS, C
    SEGOVIA, J
    DOMINGUEZ, J
    MARTINEZ, D
    MACAYA, C
    CIRCULATION, 1995, 92 (08) : 446 - 446
  • [37] Angiographic pattern of in-stent restenosis after Palmaz-Schatz stent implantation
    Sharma, SK
    Rajawat, Y
    Kakarala, V
    Marmur, JD
    Duvvuri, S
    Cocke, TP
    Ambrose, JA
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1997, 29 (02) : 1120 - 1120
  • [38] RESTENOSIS FOLLOWING SUCCESSFUL SINGLE PALMAZ-SCHATZ STENT IMPLANTATION
    FAJADET, JC
    MARCO, J
    CASSAGNEAU, BG
    ROBERT, GP
    JORDAN, CG
    LAURENT, JP
    CIRCULATION, 1990, 82 (04) : 314 - 314
  • [39] SUBOPTIMAL STENT GEOMETRY AFTER DEPLOYMENT WITH THE PALMAZ-SCHATZ STENT DELIVERY SYSTEM
    KIEMENEIJ, F
    LAARMAN, GJ
    SLAGBOOM, T
    CIRCULATION, 1993, 88 (04) : 641 - 641
  • [40] Effects of cilostazol on angiographic restenosis after the Palmaz-Schatz stent implantation
    Hara, K
    Ikari, Y
    Nakajima, H
    Kuroda, Y
    Tanabe, K
    Tsuji, T
    Nishimura, G
    Minatoya, Y
    Aoki, J
    Watanabe, T
    CORONARY ARTERY DISEASE: PREVENTION TO INTERVENTION, 2000, : 735 - 737