RETRIEVAL OF OCCLUDING UNEXPANDED PALMAZ-SCHATZ STENT FROM A SAPHENOUS AORTOCORONARY VEIN GRAFT

被引:42
|
作者
ROZENMAN, Y
BURSTEIN, M
HASIN, Y
GOTSMAN, MS
机构
[1] Department of Cardiology, Hadassah University Hospital, Jerusalem
来源
关键词
PALMAZ-SCHATZ STENT; SAPHENOUS VEIN GRAFT; UNSUCCESSFUL DEPLOYMENT;
D O I
10.1002/ccd.1810340418
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
The Palmaz-Schatz coronary stent is used frequently to reduce the rate of restenosis of balloon angioplasty in saphenous vein grafts. In many European centers, the stent, manually crimped on a balloon, is advanced across the stenosis without a protective sheath. This report describes a patient in whom an attempt to deploy a stent in the orifice of a saphenous vein graft was complicated by dislodgement of the unexpanded stent from the balloon. The unexpanded stent caused immediate occlusion of the vein graft and severe ischemic symptoms. The stent was retrieved by inflation of another balloon in the graft, distal to the stent, pulling the balloon toward the guiding catheter, and then withdrawing the whole system. (C) 1995 Wiley-Liss, Inc.
引用
收藏
页码:159 / 161
页数:3
相关论文
共 50 条
  • [1] PALMAZ-SCHATZ STENT IMPLANTATION IN THE TREATMENT OF SAPHENOUS-VEIN GRAFT LESIONS
    WONG, SC
    LEON, MB
    JOURNAL OF INVASIVE CARDIOLOGY, 1995, 7 : A23 - A29
  • [2] Successful retrieval of unexpanded Palmaz-Schatz stent from left main coronary artery
    Kobayashi, Y
    Nonogi, H
    Miyazaki, S
    Daikoku, S
    Yamamoto, Y
    Takamiya, M
    CATHETERIZATION AND CARDIOVASCULAR DIAGNOSIS, 1996, 38 (04): : 402 - 404
  • [3] Deployment of a previously embolized, unexpanded, and disarticulated Palmaz-Schatz stent
    Kirk, MM
    Herzog, WR
    CATHETERIZATION AND CARDIOVASCULAR DIAGNOSIS, 1997, 42 (03): : 331 - 334
  • [4] A SOLUTION TO THE PROBLEM OF AN UNEXPANDED PALMAZ-SCHATZ STENT FOLLOWING BALLOON RUPTURE
    PITNEY, MR
    CUMPSTON, N
    CATHETERIZATION AND CARDIOVASCULAR DIAGNOSIS, 1991, 24 (04): : 246 - 247
  • [5] INVESTIGATIONAL USE OF THE PALMAZ-SCHATZ BILIARY STENT IN LARGE SAPHENOUS-VEIN GRAFTS
    FRIEDRICH, SP
    DAVIS, SF
    KUNTZ, RE
    CARROZZA, JP
    BAIM, DS
    AMERICAN JOURNAL OF CARDIOLOGY, 1993, 71 (05): : 439 - 441
  • [6] FAVORABLE RESULTS OF TREATMENT OF NARROWED SAPHENOUS-VEIN GRAFTS WITH PALMAZ-SCHATZ STENT IMPLANTATION
    MAIELLO, L
    COLOMBO, A
    GIANROSSI, R
    GOLDENBERG, S
    MARTINI, G
    FINCI, L
    EUROPEAN HEART JOURNAL, 1994, 15 (09) : 1212 - 1216
  • [7] INFLUENCE OF EXPANDED BALLOON DIAMETER ON PALMAZ-SCHATZ STENT RECOIL IN SAPHENOUS-VEIN GRAFTS
    RECHAVIA, E
    MACKO, G
    LITVACK, F
    EIGLER, N
    CIRCULATION, 1994, 90 (04) : 279 - 279
  • [8] PALMAZ-SCHATZ STENT (PSS), DIRECTIONAL CORONARY ATHERECTOMY (DCA), AND BALLOON ANGIOPLASTY FOR SAPHENOUS-VEIN GRAFT STENOSES (SVG)
    WAKSMAN, R
    THOMAS, C
    GHAZZAL, ZMB
    WEINTRAUB, WS
    SHEN, Y
    DOUGLAS, JS
    CIRCULATION, 1994, 90 (04) : 334 - 334
  • [9] IMMEDIATE RESULTS AND LATE OUTCOMES AFTER STENT IMPLANTATION IN SAPHENOUS-VEIN GRAFT LESIONS - THE MULTICENTER US PALMAZ-SCHATZ STENT EXPERIENCE
    WONG, SC
    BAIM, DS
    SCHATZ, RA
    TEIRSTEIN, PS
    KING, SB
    CURRY, RC
    HEUSER, RR
    ELLIS, SG
    CLEMAN, MW
    OVERLIE, P
    HIRSHFELD, JW
    WALKER, CM
    LITVACK, F
    FISH, D
    BRINKER, JA
    BUCHBINDER, M
    GOLDBERG, S
    CHUANG, YC
    LEON, MB
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1995, 26 (03) : 704 - 712
  • [10] Dressing up the Palmaz-Schatz stent
    Williams, DO
    CIRCULATION, 1996, 93 (03) : 400 - 402