TRANSCATHETER OCCLUSION OF PATENT DUCTUS-ARTERIOSUS WITH GIANTURCO COILS

被引:203
|
作者
LLOYD, TR
FEDDERLY, R
MENDELSOHN, AM
SANDHU, SK
BEEKMAN, RH
机构
[1] CS MOTT CHILDRENS HOSP,DEPT PEDIAT,DIV PEDIAT CARDIOL,ANN ARBOR,MI
[2] UNIV MICHIGAN,ANN ARBOR,MI 48109
[3] UNIV ARIZONA,DEPT PEDIAT CARDIOL,TUCSON,AZ 85721
关键词
BRIEF COMMUNICATIONS; OCCLUSIONS; PATENT DUCTUS ARTERIOSUS;
D O I
10.1161/01.CIR.88.4.1412
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Background. Transcatheter occlusion with Gianturco coils has been attempted in a small number of patients with tiny (less-than-or-equal-to 1.5-mm diameter) patent ductus arteriosus, and preliminary results have been encouraging. The present study extends this method to larger ductus sizes and makes recommendations for proper coil size selection. Methods and Results. Coil occlusion was attempted in 24 consecutive patients with patent ductus arteriosus who did not require other cardiac surgery. Median patient age was 4.2 years (8 months to 30 years), and mean ductus diameter was 1.7+/-0.8 mm. Two instances of coil embolization occurred in the first 4 patients, with successful coil retrieval. Based on this experience, we proposed that the coil helical diameter should be twice or more the minimum ductus diameter, with coil length sufficient for three or more loops. With these recommendations, coils were successfully implanted in the subsequent 20 consecutive patients. Of the 22 patients with successful coil implantation, 15 (68%) had no residual shunting, and 7 had trace residual shunting by angiography. The continuous murmur was abolished in all 22 patients. No significant complications occurred, and all patients were discharged within 24 hours of successful coil implantation. No change in the systolic pressure gradient between main and left pulmonary artery or ascending and descending aorta was observed. Conclusions. Transcatheter occlusion of patent ductus arteriosus can be safely and effectively achieved in patients with ductus diameters up to 3.3 mm. Coil occlusion does not cause obstruction to flow in the left pulmonary artery or descending aorta. Coils should be selected to provide a helical diameter twice or more the minimum ductus diameter and a length sufficient for three or more loops.
引用
收藏
页码:1412 / 1420
页数:9
相关论文
共 50 条
  • [21] Percutaneous closure of the patent ductus arteriosus with Gianturco-coils
    Gildein, HP
    Germeroth, O
    Lilje, C
    Krüger, M
    Wildberg, A
    Mocellin, R
    ZEITSCHRIFT FUR KARDIOLOGIE, 1999, 88 (07): : 514 - 520
  • [22] TRANSCATHETER CLOSURE OF PATENT DUCTUS-ARTERIOSUS IN PIGLETS
    LOCK, JE
    BASS, JL
    LUND, G
    RYSAVY, JA
    LUCAS, RV
    AMERICAN JOURNAL OF CARDIOLOGY, 1985, 55 (06): : 826 - 829
  • [23] TRANSCATHETER CLOSURE OF THE PATENT DUCTUS-ARTERIOSUS IN ADULTS
    VERIN, V
    PANDE, AK
    MEIER, B
    JOURNAL OF INTERVENTIONAL CARDIOLOGY, 1992, 5 (03) : 209 - 217
  • [24] PATENT DUCTUS-ARTERIOSUS IN THE ADULT - CLINICAL MANIFESTATIONS BEFORE AND AFTER TRANSCATHETER OCCLUSION
    SCHENCK, M
    LUDOMIRSKY, A
    OLAUGHLIN, M
    MULLINS, C
    AMERICAN JOURNAL OF CARDIOLOGY, 1992, 70 (04): : 565 - 565
  • [25] BALLOON DILATATION OF THE VERY SMALL PATENT DUCTUS-ARTERIOSUS IN PREPARATION FOR TRANSCATHETER OCCLUSION
    BENSON, LN
    FREEDOM, RM
    CATHETERIZATION AND CARDIOVASCULAR DIAGNOSIS, 1989, 18 (01): : 48 - 49
  • [26] TRANSCATHETER OCCLUSION OF PATENT DUCTUS-ARTERIOSUS WITH AN ADJUSTABLE BUTTONED DEVICE - INTERNATIONAL EXPERIENCE
    RAO, PS
    HADDAD, J
    REY, C
    HAUSDORF, G
    WORMS, A
    ONORATO, E
    ARORA, R
    LOCHAN, R
    RAO, PS
    RAJU, BS
    SIDERIS, EB
    JOURNAL OF THE AMERICAN COLLEGE OF CARDIOLOGY, 1994, : A390 - A390
  • [27] CLINICAL EFFECTIVENESS AND MEDICAL ECONOMICS OF TRANSCATHETER COIL OCCLUSION OF PATENT DUCTUS-ARTERIOSUS
    SINGH, TP
    MORROW, WR
    HAKIMI, M
    WALTERS, HL
    VITALE, NA
    CIRCULATION, 1995, 92 (08) : 1471 - 1471
  • [28] CATHETER OCCLUSION OF THE PERSISTENTLY PATENT DUCTUS-ARTERIOSUS
    DYCK, JD
    BENSON, LN
    SMALLHORN, JF
    MCLAUGHLIN, PR
    FREEDOM, RM
    ROWE, RD
    AMERICAN JOURNAL OF CARDIOLOGY, 1988, 62 (16): : 1089 - 1092
  • [29] Transcatheter occlusion of a patent ductus arteriosus in a Newfoundland puppy using the Gianturco-Grifka vascular occlusion device
    Grifka, RG
    Miller, MW
    Frischmeyer, KJ
    Mullins, CE
    JOURNAL OF VETERINARY INTERNAL MEDICINE, 1996, 10 (01) : 42 - 44
  • [30] Transcatheter closure of patent ductus arteriosus in children weighing <10kg with Gianturco coils using the balloon occlusion technique
    Dalvi, B
    Nabar, A
    Goyal, V
    Naik, A
    Kulkarni, H
    Ramakanthan, R
    CATHETERIZATION AND CARDIOVASCULAR DIAGNOSIS, 1998, 44 (03): : 303 - 308