DEPTH-DOSE FLATTENING OF ELECTRON-BEAMS USING A WIRE MESH BOLUS

被引:9
|
作者
ALASTI, H
GALBRAITH, DM
机构
[1] Princess Margaret Hospital, Toronto, M4X 1K9
关键词
D O I
10.1118/1.597432
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
A common problem with low-energy electron beams (< 15 MeV) is their low surface dose when the incident electrons are monodirectional. This makes it difficult to deliver a uniform dose to tumor with any precision, limiting the clinical usefulness of such beams. A practical method is presented for greatly increasing the tissue depth enclosed by the 95% isodose region, while delivering the entire dose in a single uninterrupted treatment. Beam modification is achieved by placing a wire mesh of high atomic number (Z) on the treatment surface throughout the treatment. Electron beams of energies 6, 9, and 13 MeV are modified to produce a near uniform dose from the surface to the original depth of maximum, approximately doubling the depth enclosed by the 95% isodose. These beams have a step-function-like depth dose and can be arranged to deliver a constant dose to tumor located at varying depths while simultaneously sparing deeper tissues which are also located at varying depths. A single mesh design was found to be suitable for all energies.
引用
收藏
页码:1675 / 1683
页数:9
相关论文
共 50 条
  • [1] OPTIMIZATION OF FIELD FLATNESS AND DEPTH-DOSE FOR THERAPY ELECTRON-BEAMS
    MARBACH, JR
    ALMOND, PR
    PHYSICS IN MEDICINE AND BIOLOGY, 1981, 26 (03): : 435 - 443
  • [2] MONTE-CARLO CALCULATION OF DEPTH-DOSE PROFILES OF PULSED ELECTRON-BEAMS IN INP
    LAUGIER, A
    CHEMISTRY, G
    BARBIER, D
    VICARIO, E
    JOURNAL DE MICROSCOPIE ET DE SPECTROSCOPIE ELECTRONIQUES, 1984, 9 (04): : 259 - 265
  • [3] PARAMETRIZATION OF DEPTH DOSE FOR ELECTRON-BEAMS
    MEIGOONI, AS
    DAS, IJ
    PHYSICS IN MEDICINE AND BIOLOGY, 1987, 32 (06): : 761 - 768
  • [4] DEPENDENCE OF DEPTH-DOSE CURVES ON THE ENERGY-SPECTRUM OF 5 TO 13 MEV ELECTRON-BEAMS
    ZAGORSKI, ZP
    RADIATION PHYSICS AND CHEMISTRY, 1983, 22 (3-5): : 409 - 418
  • [5] AN EVALUATION OF THE RECOMMENDATIONS OF THE TG-25 PROTOCOL FOR DETERMINATION OF DEPTH-DOSE CURVES FOR ELECTRON-BEAMS USING IONIZATION CHAMBERS
    HUQ, MS
    AGOSTINELLI, AG
    NATH, R
    MEDICAL PHYSICS, 1995, 22 (08) : 1333 - 1337
  • [6] EXPERIMENTAL AND THEORETICAL DEPTH DOSE DISTRIBUTION FOR ELECTRON-BEAMS
    CECATTI, ER
    MALAMUT, C
    DEALMEIDA, CE
    MEDICAL PHYSICS, 1984, 11 (03) : 401 - 401
  • [7] ANALYTICAL APPROXIMATION OF DEPTH DOSE CURVES FOR ELECTRON-BEAMS
    KOVAR, I
    PHYSICS IN MEDICINE AND BIOLOGY, 1989, 34 (07): : 939 - 948
  • [8] CENTRAL AXIS DEPTH DOSE CURVE FOR ELECTRON-BEAMS
    STRYDOM, WJ
    MEDICAL PHYSICS, 1991, 18 (06) : 1254 - 1255
  • [9] EFFECT OF ANGULATION ON THE PERCENT DEPTH DOSE OF ELECTRON-BEAMS
    BIGGS, PJ
    MEDICAL PHYSICS, 1983, 10 (04) : 528 - 528
  • [10] Determination of percentage depth-dose curves for electron beams using different types of detectors
    Ding, GX
    Yu, CW
    MEDICAL PHYSICS, 2001, 28 (03) : 298 - 302