BIOCOMPATIBILITY OF EXTRACORPOREAL-CIRCULATION - INVITRO COMPARISON OF HEPARIN-COATED AND UNCOATED OXYGENATOR CIRCUITS

被引:0
|
作者
VIDEM, V
MOLLNES, TE
GARRED, P
SVENNEVIG, JL
机构
[1] UNIV OSLO,NATL HOSP,INST IMMUNOL & RHEUMATOL,OSLO 3,NORWAY
[2] UNIV OSLO,ULLEVAL HOSP,DEPT SURG,OSLO 1,NORWAY
来源
关键词
D O I
暂无
中图分类号
R5 [内科学];
学科分类号
1002 ; 100201 ;
摘要
Oxygenator/tubing sets coated with endpoint-attached heparin were compared to uncoated sets in a dynamic model of extracorporeal circulation. Biocompatibility was assessed by evaluations of complement activation and platelet loss. The median concentration of C3 activation products increased gradually from 12 AU/ml at baseline to 65 AU/ml after 120 minutes in the uncoated sets and from 12 AU/ml to 19 AU/ml after 120 minutes in the coated sets (p < 0.002). The median concentration of the terminal complement complex in the uncoated sets increased gradually from 3.1 AU/ml at baseline to 18.2 AU/ml after 120 minutes. In the coated sets the terminal complement complex reached a peak of 12.0 AU/ml at 15 minutes and returned to baseline values at 60 minutes. Median platelet loss at 120 minutes was 166 X 10(9) in the uncoated and 21 X 10(9) in the coated sets (p < 0.01). Heparin coating thus improved biocompatibility by reducing both complement activation and platelet loss.
引用
收藏
页码:654 / 660
页数:7
相关论文
共 50 条
  • [21] Activation of the clotting system:: heparin-coated versus non coated systems for extracorporeal circulation
    Urlesberger, B
    Zobel, G
    Rödl, S
    Dacar, D
    Friehs, I
    Leschnik, B
    Muntean, W
    INTERNATIONAL JOURNAL OF ARTIFICIAL ORGANS, 1997, 20 (12): : 708 - 712
  • [22] NEUTRALIZATION OF HEPARIN AFTER EXTRACORPOREAL-CIRCULATION
    BUTTNER, W
    POPOVCENIC, S
    MULLER, N
    KLADETZKY, R
    ANAESTHESIST, 1975, 24 (07): : 299 - 305
  • [23] FULL HEMODILUTION EXTRACORPOREAL-CIRCULATION WITH MEMBRANE-OXYGENATOR
    BAILLE, Y
    SICARDDESNUELLE, MP
    SIGWALT, M
    VAILLANT, A
    VARNET, B
    DESLIONS, C
    SCHEERS, F
    WIRAMUS, S
    NOUVELLE PRESSE MEDICALE, 1981, 10 (27): : 2291 - 2292
  • [24] USE OF A DOG AS OXYGENATOR DURING EXPERIMENTAL EXTRACORPOREAL-CIRCULATION
    LOYGUE, L
    CABROL, C
    MATTEI, MF
    ANNALES DE CHIRURGIE, 1973, 27 (10): : 433 - 438
  • [25] Heparin-coated equipment reduces the risk of oxygenator failure
    Wahba, A
    Philipp, A
    Behr, R
    Birnbaum, DE
    ANNALS OF THORACIC SURGERY, 1998, 65 (05): : 1310 - 1312
  • [26] Heparin-coated extracorporeal circulation with full and low dose heparinization:: Comparison of thrombin related coagulatory effects
    Mirow, N
    Brinkmann, T
    Minami, K
    Tenderich, G
    Kleesiek, K
    Körfer, R
    ARTIFICIAL ORGANS, 2001, 25 (06) : 480 - 485
  • [27] Complement consumption during cardiopulmonary bypass: comparison of Duraflo II heparin-coated and uncoated circuits in fully heparinized patients
    Hamulu, Ahmet
    Discigil, Berent
    Ozbaran, Mustafa
    Calkavur, Tanzer
    Kara, Erkan
    Kokuludag, Ali
    Buket, Suat
    Bilkay, Onol
    PERFUSION-UK, 1996, 11 (04): : 333 - 337
  • [28] IMPROVED BIOCOMPATIBILITY OF EXTRACORPOREAL-CIRCULATION BY THE USE OF AUTO-OXYGENATION INSTEAD OF ARTIFICIAL OXYGENATOR IN PERFUSION CIRCUIT
    BOCHENEK, A
    RELIGA, Z
    KUSTOSZ, R
    WNUKWOJNAR, AM
    WNUK, R
    WOJNAR, J
    WOS, S
    WITES, M
    ZEMBALA, M
    SPYT, T
    KOKOT, F
    NEPHROLOGY DIALYSIS TRANSPLANTATION, 1991, 6 : 75 - 80
  • [29] MEMBRANE OXYGENATOR OR BUBBLE OXYGENATOR DURING EXTRACORPOREAL-CIRCULATION FOR CARDIAC-SURGERY
    LOISANCE, D
    NOUVELLE PRESSE MEDICALE, 1978, 7 (15): : 1304 - 1304
  • [30] Activity of anticoagulant proteins on the polymer-coated and heparin-coated membranes in an extracorporeal circulation circuit
    Tagaya, Masashi
    Murataka, Takuo
    Okano, Shinya
    Handa, Hiroki
    Tanaka, Yuri
    Takahashi, Shunsuke
    PERFUSION-UK, 2023, 38 (03): : 600 - 608