Simultaneous Determination of Choline and Acetylcholine Based on a Trienzyme Chemiluminometric Biosensor in a Single Line Flow Injection System

被引:0
|
作者
Nobutoshi Kiba
Seiji Ito
Masaki Tachibana
Kazue Tani
Hitoshi Koizumi
机构
[1] University of Yamanashi,Department of Applied Chemistry and Biotechnology, Interdisciplinary Graduate School of Medicine and Engineering
来源
Analytical Sciences | 2003年 / 19卷
关键词
D O I
暂无
中图分类号
学科分类号
摘要
A detector for the simultaneous determination of choline (Ch) and acetylcholine (ACh) based on a sensitive trienzyme chemiluminometric biosensor in a single line flow injection (FI) system is described. Immobilized choline oxidase (ChOx), immobilized peroxidase (POx), immobilized acetylcholinesterase, and coimmobilized ChOx/POx were packed, in turn, in a transparent ETFE tube (1 mm i.d., 75 cm) and the tube was placed in front of a photomultipier tube as a flow cell. Two-peak response was obtained by one injection of the sample solution. The first and second peaks were dependent on the concentrations of Ch and ACh, respectively. The influence of some experimental parameters such as flow rate, amounts of immobilized enzymes on the behavior of the sensor was studied in order to optimize the sensitivity, sample throughput and resolution. Calibration curves were linear at 1 − 1000 nM for Ch and 3 − 3000 nM for ACh. The sample throughput was 25/h without carryover. The FI system was applied to the simultaneous determination of Ch and ACh in rabbit brain tissue homogenates.
引用
收藏
页码:1647 / 1651
页数:4
相关论文
共 50 条
  • [21] Simultaneous Chemiluminometric Determination of Levodopa and Benserazide in a Multi-pumping Flow System with Multivariate Calibration
    Karine L. Marques
    João L. M. Santos
    João A. Lopes
    José L. F. C. Lima
    Analytical Sciences, 2008, 24 : 985 - 991
  • [22] An electrochemiluminescence-based fibre optic biosensor for choline flow injection analysis
    Tsafack, VC
    Marquette, CA
    Leca, B
    Blum, LJ
    ANALYST, 2000, 125 (01) : 151 - 155
  • [23] Simultaneous chemiluminometric determination of levodopa and benserazide in a multi-pumping flow system with multivariate calibration
    Marques, Karine L.
    Santos, Joao L. M.
    Lopes, Joao A.
    Lima, Jose L. F. C.
    ANALYTICAL SCIENCES, 2008, 24 (08) : 985 - 991
  • [24] A MULTIFUNCTIONAL FLOW-INJECTION BIOSENSOR FOR THE SIMULTANEOUS DETERMINATION OF AMMONIA, CREATININE, AND UREA
    RUI, CS
    OGAWA, HI
    SONOMOTO, K
    KATO, Y
    ENZYME ENGINEERING XII, 1995, 750 : 30 - 38
  • [25] FLOW-INJECTION CHEMILUMINOMETRIC DETERMINATION OF ASCORBIC-ACID BASED ON ITS SENSITIZED PHOTOOXIDATION
    PEREZRUIZ, T
    MARTINEZLOZANO, C
    SANZ, A
    ANALYTICA CHIMICA ACTA, 1995, 308 (1-3) : 299 - 307
  • [26] Construction and Application of Flow Amperometric Biosensor Based on Enzymatic Reactor for Determination of Choline
    Tvorynska, Sofiia
    Barek, Jiri
    Josypcuk, Bohdan
    PROCEEDINGS OF INTERNATIONAL CONFERENCE MODERN ELECTROCHEMICAL METHODS XXXIX, 2019, : 218 - 222
  • [27] A biosensor for the determination of ammonium ion using flow injection amperometric system
    Preeyanut Butmee
    Sarawut Prasertsri
    Saichol Pimmongkol
    Gamolwan Tumcharern
    Albana Veseli
    Eda Mehmeti
    Kurt Kalcher
    Anchalee Samphao
    Monatshefte für Chemie - Chemical Monthly, 2017, 148 : 635 - 644
  • [28] An acetylcholinesterase/choline oxidase-based amperometric biosensor as a liquid chromatography detector for acetylcholine and choline determination in brain tissue homogenates
    Guerrieri, A
    Palmisano, F
    ANALYTICAL CHEMISTRY, 2001, 73 (13) : 2875 - 2882
  • [29] A biosensor for the determination of ammonium ion using flow injection amperometric system
    Butmee, Preeyanut
    Prasertsri, Sarawut
    Pimmongkol, Saichol
    Tumcharern, Gamolwan
    Veseli, Albana
    Mehmeti, Eda
    Kalcher, Kurt
    Samphao, Anchalee
    MONATSHEFTE FUR CHEMIE, 2017, 148 (04): : 635 - 644
  • [30] IMMOBILIZED ENZYME FLOW-INJECTION CONDUCTIMETRIC SYSTEM FOR THE DETERMINATION OF ACETYLCHOLINE
    GODINHO, OES
    RODRIGUES, TC
    TUBINO, M
    NETO, GDO
    ALEIXO, LM
    ANALYTICAL PROCEEDINGS, 1995, 32 (08): : 333 - 335