An empirical expression for the full energy peak efficiency of an N-type high purity germanium detector

被引:0
|
作者
E. K. Osae
B. J. B. Nyarko
Y. Serfor-Armah
E. O. Darko
机构
[1] Ghana Atomic Energy Commission,Neutron Activation Analysis Laboratory, National Nuclear Research Institute
[2] Ghana Atomic Energy Commission,Radiation Protection Board
来源
Journal of Radioanalytical and Nuclear Chemistry | 1999年 / 242卷
关键词
Standard Deviation; Physical Chemistry; Inorganic Chemistry; Radionuclide; Energy Range;
D O I
暂无
中图分类号
学科分类号
摘要
An empirical expression for the full energy peak efficiency (ε) in terms of the gamma-ray energy (E) and the vertical distance from the detector surface (z) (i.e.,ε=ε(z,E)) has been obtained for an N-type high purity germanium (HPGE) detector using an extended mixed standard radionuclide solution. Comparison of the calculated efficiencies and the experimentally measured values for the energy range of 59.5–1332.5 keV and a source-to-detector distance of 1.2–7.2 cm showed that the theoretical values agree with that of the experiment within 1.5–3.7% standard deviations. This shows a good agreement between the theory and the experiment.
引用
收藏
页码:617 / 622
页数:5
相关论文
共 50 条
  • [21] Methods and software for predicting germanium detector absolute full-energy peak efficiencies
    Jackman, K. R.
    Biegalski, S. R.
    JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY, 2009, 279 (01) : 355 - 360
  • [22] An empirical formula to calculate the full energy peak efficiency of scintillation detectors
    Badawi, Mohamed S.
    Abd-Elzaher, Mohamed
    Thabet, Abouzeid A.
    El-khatib, Ahmed M.
    APPLIED RADIATION AND ISOTOPES, 2013, 74 : 46 - 49
  • [23] Optical deep level transient spectroscopy of minority carrier traps in n-type high-purity germanium
    Blondeel, A
    Clauws, P
    Vyncke, D
    JOURNAL OF APPLIED PHYSICS, 1997, 81 (10) : 6767 - 6772
  • [24] ENERGY LEVELS IN ELECTRON IRRADIATED N-TYPE GERMANIUM.
    Mooney, P.M.
    Cherki, M.
    Bourgoin, J.C.
    1979, 40 (02):
  • [25] Development of program DETSIM to simulate detector's full energy peak efficiency
    Thanh, T. T.
    Tao, C. V.
    Au, B. H.
    Chuong, H. D.
    KERNTECHNIK, 2012, 77 (06) : 449 - 452
  • [26] Characteristic absolute efficiency response curves of a high purity germanium detector in the energy range 50-1500 keV
    Rajput, MU
    Ahmad, M
    Ahmad, W
    JOURNAL OF RADIOANALYTICAL AND NUCLEAR CHEMISTRY, 2002, 251 (03) : 457 - 462
  • [27] Monte Carlo efficiency transfer method for full energy peak efficiency calibration of three type HPGe detectors: A coaxial N-type, a coaxial P-type and four BEGe detectors
    Liu Liye
    Ma Jizeng
    Franck, Didier
    de Carlan, Loic
    Zhang Binquan
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 2006, 564 (01): : 608 - 613
  • [28] ACCEPTOR LUMINESCENCE IN HIGH-PURITY N-TYPE GAAS
    ROSSI, JA
    WOLFE, CM
    DIMMOCK, JO
    PHYSICAL REVIEW LETTERS, 1970, 25 (23) : 1614 - &
  • [29] HYDROGEN PASSIVATION OF HIGH-PURITY N-TYPE INP
    CROOKES, CG
    LANCEFIELD, D
    WATERHOUSE, K
    ADAMS, AR
    GREENE, PD
    GLEW, RW
    ELECTRONICS LETTERS, 1990, 26 (06) : 369 - 371
  • [30] EVIDENCE FOR DELOCALIZATION IN HIGH-PURITY N-TYPE INP
    BENZAQUEN, M
    WALSH, D
    SOLID STATE COMMUNICATIONS, 1994, 89 (12) : 1033 - 1036