Estimation of central systolic blood pressure using an oscillometric blood pressure monitor

被引:0
|
作者
Hao-Min Cheng
Kang-Ling Wang
Ying-Hwa Chen
Shing-Jong Lin
Lung-Ching Chen
Shih-Hsien Sung
Philip Yu-An Ding
Wen-Chung Yu
Jaw-Wen Chen
Chen-Huan Chen
机构
[1] Taipei Veterans General Hospital,Department of Medical Research and Education
[2] National Yang-Ming University,Department of Medicine
[3] Taipei Veterans General Hospital,Department of Research and Education
[4] Cardiovascular Research Center,undefined
[5] National Yang-Ming University,undefined
来源
Hypertension Research | 2010年 / 33卷
关键词
central blood pressure; oscillometric signals; pressure wave reflection; pulse volume plethysmography;
D O I
暂无
中图分类号
学科分类号
摘要
Current noninvasive techniques for assessing central aortic pressure require the recording of an arterial pressure wave using a high-fidelity applanation tonometer. We therefore developed and validated a novel method to estimate the central aortic systolic pressure using an oscillometric blood pressure monitor alone. Invasive high-fidelity right brachial and central aortic pressure waves, and left-brachial pulse volume plethysmography from an oscillometric blood pressure monitor, were obtained at baseline and 3 min after administration of sublingual nitroglycerin in 100 patients during cardiac catheterization. In the initial 50 patients (Generation Group), Central systolic blood pressure was predicted by a multi-variate prediction model generated from the comprehensive analysis of the invasive brachial pressure wave, including brachial late-systolic shoulder pressure value and parameters related to wave reflection and arterial compliance. Another prediction model was similarly constructed from the noninvasively calibrated pulse volume plethysmography. Both models were validated in the subsequent 50 patients (Validation Group) with results: r=0.98 (P<0.001) and mean difference=0.5±4.5 (95% confidence interval −8.3 to 9.3) mm Hg for the invasive model, and r=0.93 (P<0.001) and mean difference=−0.1±7.6 (95% confidence interval −15.0 to 14.8) mm Hg for the noninvasive model. Thus, our results indicate that central aortic systolic blood pressure could be estimated by analysis of the noninvasive brachial pressure wave alone from an oscillometric blood pressure monitor.
引用
收藏
页码:592 / 599
页数:7
相关论文
共 50 条
  • [41] Systolic Blood Pressure Estimation Using Android Smart Phones
    Li, Han
    Zhao, Hai
    PROCEEDINGS OF THE 2013 6TH INTERNATIONAL CONFERENCE ON BIOMEDICAL ENGINEERING AND INFORMATICS (BMEI 2013), VOLS 1 AND 2, 2013, : 260 - 264
  • [42] Diagnostic accuracy of an oscillometric blood pressure monitor for atrial fibrillation screening
    Li, Linyi
    Bing, Sen
    Liu, Kui
    Jiang, Jun
    Zhang, Jieqiong
    Chen, Changsheng
    Wan, Yi
    BLOOD PRESSURE MONITORING, 2023, 28 (03) : 144 - 148
  • [43] Evaluation of a new oscillometric blood pressure monitor by cuff inflation method
    Nakajima, M
    Komatsu, T
    Kimura, T
    Fujiwara, Y
    Shimada, Y
    ANESTHESIOLOGY, 1996, 85 (3A) : A401 - A401
  • [44] Comparison of noninvasive estimations of central blood pressure by tonometric radial pressure waveform calibrated with oscillometric brachial blood pressure
    Hirata, K.
    Sagawa, Y.
    Kojima, I.
    Yamazaki, M.
    Shinomiya, N.
    Miyawaki, Y.
    Momomura, S. -I.
    JOURNAL OF HYPERTENSION, 2008, 26 : S490 - S490
  • [45] Validation of the Welch Allyn ‘Vital Signs’ oscillometric blood pressure monitor
    CR Jones
    K Taylor
    L Poston
    AH Shennan
    Journal of Human Hypertension, 2001, 15 : 191 - 195
  • [46] The Impact of Central Blood Pressure Levels on the Relationship Between Oscillometric and Central Blood Pressure Measurements: A Multicenter Invasive Study
    Alihanoglu, Yusuf I.
    Kayrak, Mehmet
    Ulgen, Mehmet S.
    Yazici, Mehmet
    Yazici, Mehmet
    Yilmaz, Remzi
    Demir, Kenan
    Dogan, Yildiz
    Sizer, Murat
    Ozhan, Hakan
    Koc, Fatih
    Bodur, Sait
    JOURNAL OF CLINICAL HYPERTENSION, 2013, 15 (09): : 681 - 686
  • [47] Oscillometric Blood Pressure Estimation Using Principal Component Analysis and Neural Networks
    Forouzanfar, Mohamad
    Dajani, Hilmi R.
    Groza, Voicu Z.
    Bolic, Miodrag
    Rajan, Sreeraman
    IEEE TIC-STH 09: 2009 IEEE TORONTO INTERNATIONAL CONFERENCE: SCIENCE AND TECHNOLOGY FOR HUMANITY, 2009, : 981 - 986
  • [48] P2.14 Influence of Calibration of Peripheral Pressure on the Estimation of Central Systolic Blood Pressure
    A. Guilcher
    B. Clapp
    S. Brett
    P. Chowienczyk
    Artery Research, 2008, 2 (3) : 109 - 110
  • [49] Use of a smartphone camera to estimate systolic blood pressure compared to standard oscillometric blood pressure measurement a prospective trial
    Eckstein, J.
    Burkardt, T.
    Winterhalder, C.
    Leonardi, L.
    Thommen, E.
    Weber, S.
    Birkemeyer, R.
    Seeck, A.
    Koenig, N.
    Doerr, M.
    EUROPEAN HEART JOURNAL, 2017, 38 : 61 - 61
  • [50] Confidence Interval Estimation for Oscillometric Blood Pressure Measurements Using Bootstrap Approaches
    Lee, Soojeong
    Bolic, Miodrag
    Groza, Voicu Z.
    Dajani, Hilmi R.
    Rajan, Sreeraman
    IEEE TRANSACTIONS ON INSTRUMENTATION AND MEASUREMENT, 2011, 60 (10) : 3405 - 3415