Fatigue strength of common tibial intramedullary nail distal locking screws

被引:21
|
作者
Griffin L.V. [1 ]
Harris R.M. [2 ]
Zubak J.J. [3 ]
机构
[1] California Polytechnic State University, Biomedical and General Engineering, San Luis Obispo, CA
[2] Orthopaedic Trauma, Holston Valley Medical Center, Wellmont Health System, Kingsport, TN
[3] Shannon Clinic Southwest, San Angelo, TX
关键词
Fatigue Life; Intramedullary Nail; Full Weight Bearing; Hardware Failure; High Cycle Fatigue;
D O I
10.1186/1749-799X-4-11
中图分类号
学科分类号
摘要
Background. Premature failure of either the nail and/or locking screws with unstable fracture patterns may lead to angulation, shortening, malunion, and IM nail migration. Up to thirty percent of all unreamed nail locking screws can break after initial weight bearing is allowed at 8-10 weeks if union has not occurred. The primary problem this presents is hardware removal during revision surgery. The purposes of our study was to evaluate the relative fatigue resistance of distal locking screws and bolts from representative manufacturers of tibial IM nail systems, and develop a relative risk assessment of screws and materials used. Evaluations included quantitative and qualitative measures of the relative performance of these screws. Methods. Fatigue tests were conducted to simulate a comminuted fracture that was treated by IM nailing assuming that all load was carried by the screws. Each screw type was tested ten times in a single screw configuration. One screw type was tested an additional ten times in a two-screw parallel configuration. Fatigue tests were performed using a servohydraulic materials testing system and custom fixturing that simulated screws placed in the distal region of an appropriately sized tibial IM nail. Fatigue loads were estimated based on a seventy-five kilogram individual at full weight bearing. The test duration was one million cycles (roughly one year), or screw fracture, whichever occurred first. Failure analysis of a representative sample of titanium alloy and stainless steel screws included scanning electron microscopy (SEM) and quantitative metallography. Results. The average fatigue life of a single screw with a diameter of 4.0 mm was 1200 cycles, which would correspond roughly to half a day of full weight bearing. Single screws with a diameter of 4.5 mm or larger have approximately a 50 percent probability of withstanding a week of weight bearing, whereas a single 5.0 mm diameter screw has greater than 90 percent probability of withstanding more than a week of weight bearing. If two small diameter screws are used, our tests showed that the probability of withstanding a week of weight bearing increases from zero to about 20 percent, which is similar to having a single 4.5 mm diameter screw providing fixation. Conclusion. Our results show that selecting the system that uses the largest distal locking screws would offer the best fatigue resistance for an unstable fracture pattern subjected to full weight bearing. Furthermore, using multiple screws will substantially reduce the risk of premature hardware failure.
引用
收藏
相关论文
共 50 条
  • [31] Peroneal nerve damage associated with the proximal locking screws of the AIM tibial nail
    Hems, TEJ
    Jones, BG
    INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2005, 36 (05): : 651 - 654
  • [32] Fatigue strength of locking screws and prototypes used in small-diameter tibial nails:: A biomechanical study
    Gaebler, C
    Stanzl-Tschegg, S
    Heinze, G
    Holper, B
    Milne, T
    Berger, G
    Vécsei, V
    JOURNAL OF TRAUMA-INJURY INFECTION AND CRITICAL CARE, 1999, 47 (02): : 379 - 384
  • [33] Endoscopy-assisted distal locking of an intramedullary nail: A new experimental technique to reduce radiation exposure during distal locking of the intramedullary nails
    Davut, Serkan
    Dogramaci, Yunus
    JOINT DISEASES AND RELATED SURGERY, 2021, 32 (03): : 642 - 648
  • [34] Intramedullary Tibial Nail Fixation of Simple Intraarticular Distal Tibia Fractures
    Scolaro, John A.
    Broghammer, Francis H.
    Donegan, Derek J.
    JOURNAL OF ORTHOPAEDIC TRAUMA, 2016, 30 : S12 - S16
  • [35] Distal Tibial Reconstruction with Use of a Circular External Fixator and an Intramedullary Nail
    Eralp, Levent
    Kocaoglu, Mehmet
    JOURNAL OF BONE AND JOINT SURGERY-AMERICAN VOLUME, 2008, 90A : 181 - 194
  • [36] Mechanical strength, fatigue life, and failure analysis of two prototypes and five conventional tibial locking screws
    Hou, SM
    Wang, JL
    Lin, J
    JOURNAL OF ORTHOPAEDIC TRAUMA, 2002, 16 (10) : 701 - 708
  • [37] The Role of Biplanar Distal Locking in Intramedullary Nailing of Tibial Shaft Fractures
    Alemdaroglu, Kadir Bahadir
    Iltar, Serkan
    Ozturk, Alper
    Gultac, Emre
    Yucens, Mehmet
    Aydogan, Nevres Hurriyet
    ARCHIVES OF BONE AND JOINT SURGERY-ABJS, 2019, 7 (01): : 33 - 37
  • [38] How to measure distal locking screws for intramedullary nails without a depth gauge
    Harrison, W. D.
    Rourke, K. H.
    Narayan, B.
    ANNALS OF THE ROYAL COLLEGE OF SURGEONS OF ENGLAND, 2015, 97 (03) : 239 - 239
  • [39] Literature review of current techniques for the insertion of distal screws into intramedullary locking nails
    Whatling, GM
    Nokes, LDM
    INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2006, 37 (02): : 109 - 119
  • [40] For intramedullary nailing of femoral shaft fractures, talon fixation is helpful to cope with the troublesome distal locking, but conventional distal locking with screws offers a more stable construct: Talon femoral nail versus conventional femoral nail
    Yapici, Furkan
    Gur, Volkan
    Onac, Osman
    Alpay, Yakup
    Tardus, Ismail
    Ucpunar, Hanifi
    Camurcu, Yalkin
    ULUSAL TRAVMA VE ACIL CERRAHI DERGISI-TURKISH JOURNAL OF TRAUMA & EMERGENCY SURGERY, 2022, 28 (04): : 513 - 522