Fatigue strength of common tibial intramedullary nail distal locking screws

被引:21
|
作者
Griffin L.V. [1 ]
Harris R.M. [2 ]
Zubak J.J. [3 ]
机构
[1] California Polytechnic State University, Biomedical and General Engineering, San Luis Obispo, CA
[2] Orthopaedic Trauma, Holston Valley Medical Center, Wellmont Health System, Kingsport, TN
[3] Shannon Clinic Southwest, San Angelo, TX
关键词
Fatigue Life; Intramedullary Nail; Full Weight Bearing; Hardware Failure; High Cycle Fatigue;
D O I
10.1186/1749-799X-4-11
中图分类号
学科分类号
摘要
Background. Premature failure of either the nail and/or locking screws with unstable fracture patterns may lead to angulation, shortening, malunion, and IM nail migration. Up to thirty percent of all unreamed nail locking screws can break after initial weight bearing is allowed at 8-10 weeks if union has not occurred. The primary problem this presents is hardware removal during revision surgery. The purposes of our study was to evaluate the relative fatigue resistance of distal locking screws and bolts from representative manufacturers of tibial IM nail systems, and develop a relative risk assessment of screws and materials used. Evaluations included quantitative and qualitative measures of the relative performance of these screws. Methods. Fatigue tests were conducted to simulate a comminuted fracture that was treated by IM nailing assuming that all load was carried by the screws. Each screw type was tested ten times in a single screw configuration. One screw type was tested an additional ten times in a two-screw parallel configuration. Fatigue tests were performed using a servohydraulic materials testing system and custom fixturing that simulated screws placed in the distal region of an appropriately sized tibial IM nail. Fatigue loads were estimated based on a seventy-five kilogram individual at full weight bearing. The test duration was one million cycles (roughly one year), or screw fracture, whichever occurred first. Failure analysis of a representative sample of titanium alloy and stainless steel screws included scanning electron microscopy (SEM) and quantitative metallography. Results. The average fatigue life of a single screw with a diameter of 4.0 mm was 1200 cycles, which would correspond roughly to half a day of full weight bearing. Single screws with a diameter of 4.5 mm or larger have approximately a 50 percent probability of withstanding a week of weight bearing, whereas a single 5.0 mm diameter screw has greater than 90 percent probability of withstanding more than a week of weight bearing. If two small diameter screws are used, our tests showed that the probability of withstanding a week of weight bearing increases from zero to about 20 percent, which is similar to having a single 4.5 mm diameter screw providing fixation. Conclusion. Our results show that selecting the system that uses the largest distal locking screws would offer the best fatigue resistance for an unstable fracture pattern subjected to full weight bearing. Furthermore, using multiple screws will substantially reduce the risk of premature hardware failure.
引用
收藏
相关论文
共 50 条
  • [21] Bending strength and holding power of tibial locking screws
    Lin, J
    Lin, SJ
    Chiang, HS
    Hou, SM
    CLINICAL ORTHOPAEDICS AND RELATED RESEARCH, 2001, (385) : 199 - 206
  • [22] A COMPARATIVE STUDY OF INTRAMEDULLARY INTERLOCKING NAIL AND LOCKING PLATE FIXATION IN THE MANAGEMENT OF EXTRA ARTICULAR DISTAL TIBIAL FRACTURES
    Pawar, Eknath D.
    Agrawal, Saurabh R.
    Patil, Atul W.
    Choudhary, Sandeep
    Asadi, Ghaniuzzoha
    JOURNAL OF EVOLUTION OF MEDICAL AND DENTAL SCIENCES-JEMDS, 2014, 3 (24): : 6812 - 6826
  • [23] Insertion of distal locking screws of tibial intramedullary nails: A comparison between the free-hand technique and the SURESHOT™ Distal Targeting System
    Moreschini, O.
    Petrucci, V.
    Cannata, R.
    INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 2014, 45 (02): : 405 - 407
  • [24] The risk of neurovascular injury with distal locking screws of humeral intramedullary nails
    Rupp, RE
    Chrissos, MG
    Ebraheim, NA
    ORTHOPEDICS, 1996, 19 (07) : 593 - 595
  • [25] The Insertion of Intramedullary Nail Locking Screws Without Fluoroscopy: A Faster and Safer Technique
    Chan, Daniel S.
    Burris, Richard B.
    Erdogan, Murat
    Sagi, H. Claude
    JOURNAL OF ORTHOPAEDIC TRAUMA, 2013, 27 (07) : 363 - 366
  • [26] A SIMPLE JIG TO EASE THE INSERTION OF DISTAL SCREWS IN INTRAMEDULLARY LOCKING NAILS
    HASHEMINEJAD, A
    GARLICK, N
    GODDARD, NJ
    INJURY-INTERNATIONAL JOURNAL OF THE CARE OF THE INJURED, 1994, 25 (06): : 407 - 408
  • [27] Anatomical study of the placement of proximal oblique locking screws in intramedullary tibial nailing
    Jones, B. G.
    Mehin, R.
    Young, D.
    JOURNAL OF BONE AND JOINT SURGERY-BRITISH VOLUME, 2007, 89B (11): : 1495 - 1497
  • [28] Treatment of bilateral distal tibial nonunions.: Use of an intramedullary compression nail and a modified locking compression plate osteosynthesis
    Mueckley, T.
    Wilharm, A.
    Marintschev, I.
    Hofmann, G. O.
    Buehren, V.
    UNFALLCHIRURG, 2007, 110 (03): : 264 - 267
  • [29] Retrograde tibial intramedullary nail versus minimally invasive locking plate for extra-articular distal tibial fractures: a comparative and retrospective study
    Liu, Hui
    Xu, Weizhen
    Xiong, Yuanfei
    Zhang, Jinhui
    Xu, Zunying
    Wu, Jin
    JOURNAL OF ORTHOPAEDIC SURGERY AND RESEARCH, 2024, 19 (01):
  • [30] Distal locking screws
    Blasier, RB
    ORTHOPEDICS, 1996, 19 (11) : 926 - 926