A cryosystem for cooling samples in a secondary-ion mass spectrometer with a static analyzer

被引:0
|
作者
A. N. Pustovit
A. M. Ionov
机构
[1] Russian Academy of Sciences,Institute of Microelectronics Technology and High Purity Materials
[2] Russian Academy of Sciences,Institute of Solid State Physics
关键词
Cryogenic Temperature; Cooling Sample; Cryogenic System; GaAs Sample; Transport Line;
D O I
暂无
中图分类号
学科分类号
摘要
A cryogenic system for cooling samples in an IMS-4F secondary-ion mass spectrometer with a static analyzer is described. The cryogenic system allows maintenance of the sample temperature with an accuracy of ±1 K in a temperature range of 60–420 K by combining heating and cooling (by using liquid helium or nitrogen). The effect of the sample temperature (300 and 110 K) on the secondary-ion mass spectra of Si and GaAs samples is considered.
引用
收藏
页码:91 / 94
页数:3
相关论文
共 50 条
  • [21] SECONDARY-ION MASS-SPECTROMETRY OF GLYCOSYLATED PORPHYRINS
    SPIRO, M
    BLAIS, JC
    BOLBACH, G
    FOURNIER, F
    TABET, JC
    DRIAF, K
    GAUD, O
    GRANET, R
    KRAUSZ, P
    INTERNATIONAL JOURNAL OF MASS SPECTROMETRY AND ION PROCESSES, 1994, 134 (2-3): : 229 - 238
  • [22] STATIC AND DYNAMIC SURFACE ANALYSIS WITH A SECONDARY ION MASS-SPECTROMETER, A MOSSBAUER SPECTROMETER, AND A PURE GERMANIUM GAMMA SPECTROMETER
    HIGATSBERGER, MJ
    ACTA PHYSICA AUSTRIACA, 1978, 49 (3-4): : 181 - 191
  • [23] Secondary-ion mass spectrometry of photosensitive heterophase semiconductor
    Rokakh, AG
    Zhukov, AG
    Stetsura, SV
    Serdobintsev, AA
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION B-BEAM INTERACTIONS WITH MATERIALS AND ATOMS, 2004, 226 (04): : 595 - 600
  • [24] STUDIES OF ADHESION BY SECONDARY-ION MASS-SPECTROMETRY
    SPOOL, AM
    IBM JOURNAL OF RESEARCH AND DEVELOPMENT, 1994, 38 (04) : 391 - 411
  • [25] MICROANALYSIS USING SECONDARY-ION MASS-SPECTROMETRY
    LING, YC
    JOURNAL OF THE CHINESE CHEMICAL SOCIETY, 1994, 41 (03) : 329 - 333
  • [26] SURFACE CONCENTRATION OF CHAIN-ENDS IN POLYSTYRENE DETERMINED BY STATIC SECONDARY-ION MASS-SPECTROSCOPY
    AFFROSSMAN, S
    HARTSHORNE, M
    JEROME, R
    PETHRICK, RA
    PETITJEAN, S
    VILAR, MR
    MACROMOLECULES, 1993, 26 (23) : 6251 - 6254
  • [27] Secondary-Ion Mass Spectrometry of Genetically Encoded Targets
    Vreja, Ingrid C.
    Kabatas, Selda
    Saka, Sinem K.
    Kroehnert, Katharina
    Hoeschen, Carmen
    Opazo, Felipe
    Diederichsen, Ulf
    Rizzoli, Silvio O.
    ANGEWANDTE CHEMIE-INTERNATIONAL EDITION, 2015, 54 (19) : 5784 - 5788
  • [28] QUANTIFICATION OF ENDGROUP FUNCTIONALIZATION USING TIME-OF-FLIGHT STATIC SECONDARY-ION MASS-SPECTROMETRY
    BELU, AM
    HUNT, MO
    DESIMONE, JM
    LINTON, RW
    ABSTRACTS OF PAPERS OF THE AMERICAN CHEMICAL SOCIETY, 1993, 206 : 45 - POLY
  • [29] Improvement of signal intensities in static secondary-ion mass spectrometry using halide additives and substrate modification
    Gusev, AI
    Choi, BK
    Hercules, DM
    JOURNAL OF MASS SPECTROMETRY, 1998, 33 (05): : 480 - 485
  • [30] SIMULTANEOUS ION-SCATTERING AND SECONDARY-ION MASS-SPECTROMETRY
    VASILE, MJ
    MALM, DL
    INTERNATIONAL JOURNAL OF MASS SPECTROMETRY AND ION PROCESSES, 1976, 21 (1-2): : 145 - 157