Verification of patient specific quality assurance system for volumetric modulated arc therapy (VMAT)

被引:0
|
作者
Kwame Anokye Amoabeng
Anne Beate Langeland Marthinsen
Francis Hasford
Samuel Nii Adu Tagoe
Evelyn Anaafi
机构
[1] University of Ghana,Department of Medical Physics, Graduate School of Nuclear and Allied Sciences
[2] St. Olav’s University Hospital,Department of Radiotherapy
[3] Korlebu Teaching Hospital,National Centre for Radiotherapy and Nuclear Medicine
[4] Sweden – Ghana Medical Center,undefined
来源
Health and Technology | 2022年 / 12卷
关键词
VMAT; Gamma index; Quality assurance; Flattening filter free radiation; Optimization;
D O I
暂无
中图分类号
学科分类号
摘要
Volumetric Modulated Arc Therapy (VMAT) plans and its treatment are complex radiation therapy modalities and its clinical implementation requires very accurate acceptance testing and a comprehensive quality assurance program. Failure in their periodic applications can result in errors in treatment delivery. It is therefore needed to compare the calculated dose distribution to the measured dose distribution of treatment plans to ensure accuracy and efficiency to minimize errors in planned radiation dose delivery to the patient. VMAT plans for fifteen patients with cervical cancer with pathological pelvic lymph node metastasis were included in this study. Two VMAT plans were made for each patient, one using conventional 6 MV beam with flattening filter and one with flattening filter free beam (FFF). The VMAT plans were made using two arcs, and then recalculated to give the planned dose distribution to the detectors in a Delta4 phantom. All thirty VMAT plans were irradiated on the Delta4 phantom using three Elekta linear accelerators, and planned and calculated doses to the phantom were compared. Measurements were done both before and after a routine linac optimization, and variation in radiation delivery between the linacs were studied. After machine optimization, there was an increase in the mean total gamma pass rate for both the 6 MV plans from 98.7 to 99.9% and the FFF plans also had an increase in the gamma pass rate from 91.7% to 98.4% for SB5 linac. There was significant statistical difference (p < 0.05) between the pass rates of the 6 MV plans and the FFF plans. 6 MV VMAT plans with conventional flattened beams are delivered more accurately and hence more beneficial compared to flattening filter free (FFF) VMAT plans for external radiation of cervical cancer with affected pelvic lymph nodes.
引用
收藏
页码:779 / 786
页数:7
相关论文
共 50 条
  • [31] Comparisons of volumetric modulated arc therapy (VMAT) quality assurance (QA) systems: sensitivity analysis to machine errors
    Liang, Bin
    Liu, Bo
    Zhou, Fugen
    Yin, Fang-Fang
    Wu, Qiuwen
    RADIATION ONCOLOGY, 2016, 11
  • [32] Comparisons of volumetric modulated arc therapy (VMAT) quality assurance (QA) systems: sensitivity analysis to machine errors
    Bin Liang
    Bo Liu
    Fugen Zhou
    Fang-fang Yin
    Qiuwen Wu
    Radiation Oncology, 11
  • [33] Establishing an optimized patient-specific verification program for volumetric modulated arc therapy
    Serna, Alfredo
    Mata, Fernando
    Puchades, Vicente
    MEDICAL DOSIMETRY, 2013, 38 (03) : 274 - 279
  • [34] Prospective Clinical Validation of Virtual Patient-Specific Quality Assurance of Volumetric Modulated Arc Therapy Radiation Therapy Plans
    Wall, Phillip D. H.
    Hirata, Emily
    Morin, Olivier
    Valdes, Gilmer
    Witztum, Alon
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2022, 113 (05): : 1091 - 1102
  • [35] Sensitivity of Three Patient-Specific Quality Assurance Systems to MLC Aperture Errors With Volumetric Modulated Arc Therapy
    Hu, Jinyou
    Gu, Shaoxian
    Wang, Ningyu
    Cui, Fengjie
    Zhang, Shengyuan
    Yin, Chuou
    Cai, Yunzhu
    Gou, Chengjun
    Zou, Lian
    Wu, Zhangwen
    TECHNOLOGY IN CANCER RESEARCH & TREATMENT, 2022, 21
  • [36] VOLUMETRIC-MODULATED ARC THERAPY: EFFECTIVE AND EFFICIENT END-TO-END PATIENT-SPECIFIC QUALITY ASSURANCE
    O'Daniel, Jennifer
    Das, Shiva
    Wu, Q. Jackie
    Yin, Fang-Fang
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2012, 82 (05): : 1567 - 1574
  • [37] Treatment Plan Complexity Quantification for Predicting Patient-Specific Quality Assurance of Stereotactic Volumetric Modulated Arc Therapy
    Xudong, X.
    Ding, Y.
    Wei, W.
    Ma, C.
    Wang, X.
    MEDICAL PHYSICS, 2022, 49 (06) : E286 - E287
  • [38] Cloud-Based Monte Carlo Patient-Specific Quality Assurance (QA) Method for Volumetric-Modulated Arc Therapy (VMAT): Clinical and Educational Impacts
    Chen, X.
    Xing, L.
    Mok, E.
    Luxton, G.
    Bush, K.
    MEDICAL PHYSICS, 2013, 40 (06)
  • [39] Effect of dose grid resolution on the results of patient-specific quality assurance for intensity-modulated radiation therapy and volumetric modulated arc therapy
    Chun, M.
    Kim, J., I
    Oh, G. H.
    Wu, H. G.
    Park, J. M.
    INTERNATIONAL JOURNAL OF RADIATION RESEARCH, 2020, 18 (03): : 521 - 530
  • [40] Novel dosimetric phantom for quality assurance of volumetric modulated arc therapy
    Letourneau, Daniel
    Publicover, Julia
    Kozelka, Jakub
    Moseley, Douglas J.
    Jaffray, David A.
    MEDICAL PHYSICS, 2009, 36 (05) : 1813 - 1821