Skin and build up dose determination for a 2.5 MV medical linear accelerator imaging beam

被引:0
|
作者
Martin Butson
Ethan Butson
Johnny Morales
Robin Hill
机构
[1] Chris O’Brien Lifehouse,Department of Radiation Oncology
[2] The University of Sydney,Institute of Medical Physics, School of Physics
[3] The University of Sydney,School of Physics
[4] Queensland University of Technology,School of Chemistry, Physics and Mechanical Engineering
来源
Australasian Physical & Engineering Sciences in Medicine | 2019年 / 42卷
关键词
Radiotherapy; Skin dose; Build up; Dosimetry; Radiation; Surface dose; Radiochromic film;
D O I
暂无
中图分类号
学科分类号
摘要
The 2.5 MV Imaging beam produced by a Varian TrueBeam linear accelerator produces a dose build up effect at the beam entrance similar to other high energy photon beams. The surface dose values were found to range from 39% of maximum dose at a 5 cm × 5 cm field size up to 69% of maximum at a 40 cm × 40 cm field. The depth of maximum dose deposition was found to range from 5 mm at smaller field sizes to 4 mm at larger field sizes. Whilst large absorbed doses will not be delivered utilizing these beams, the data provided will allow the medical physics community to assess and estimate doses to patient’s skin and subcutaneous tissue from low energy MV imaging beams.