Direct Current Combined With Bipolar Radiofrequency Ablation: An Ex Vivo Feasibility Study

被引:0
|
作者
Toshihiro Tanaka
Tobias Penzkofer
Peter Isfort
Philipp Bruners
Catherine Disselhorst-Klug
Elmar Junker
Kimihiko Kichikawa
Thomas Schmitz-Rode
Andreas H. Mahnken
机构
[1] RWTH Aachen University,Applied Medical Engineering, Helmholtz
[2] RWTH Aachen University,Institute Aachen
[3] Nara Medical University,Department of Diagnostic Radiology, Aachen University Hospital
来源
CardioVascular and Interventional Radiology | 2011年 / 34卷
关键词
Radiofrequency ablation; Liver tumor; Direct current;
D O I
暂无
中图分类号
学科分类号
摘要
The combination of radiofrequency ablation (RFA) with direct current (DC) is a promising strategy to improve the efficiency of RFA. However, DC-enhanced monopolar RFA is limited by electrolytic injury at the positive-electrode site. The aim of this study was to investigate the feasibility of the DC-enhanced bipolar RFA. To obviate the need for the subcutaneous positive electrode, the DC circuit was combined with a commercially available bipolar RFA system, in which both poles of the DC circuit are connected to a single RF probe. DC was applied for 15 min and followed by RFA in bovine livers using the following various DC currents: (1) no DC (control), (2) 3V continued until the end of RFA, (3) 5V continued until the end of RFA, (4) 10V continued until the end of RFA, (5) 5V continued in the circuit with reversed pole, (6) 3V stopped after initiation of RFA, and (7) 5V stopped. Coagulation volume, temperatures at a distance of 5, 10, and 15 mm from the RF probe, mean amperage, ablation duration, applied energy, minimum impedance, and degree of tissue charring were assessed and compared (analysis of variance, Student–Newman–Keuls test). All combined DC and RFA groups did increase coagulation volume. The 10V continued group showed significantly lower applied energy, shortest ablation duration, highest minimum impedance, and highest degree of charring with the lowest coagulation volume (p < 0.05). DC-enhanced bipolar RFA with both poles of the DC circuit on a single probe appears to be ineffective.
引用
收藏
页码:631 / 636
页数:5
相关论文
共 50 条
  • [21] An ex-vivo experimental study on optimization of bipolar radiofrequency liver ablation using perfusion-cooled electrodes
    Lee, JM
    Han, JK
    Kim, SH
    Lee, JY
    Shin, KS
    Choi, BI
    ACTA RADIOLOGICA, 2005, 46 (05) : 443 - 451
  • [22] Hepatic bipolar radiofrequency ablation using perfused-cooled electrodes:: a comparative study in the ex vivo bovine liver
    Lee, JM
    Han, JK
    Kim, SH
    Lee, JY
    Choi, SH
    Choi, BI
    BRITISH JOURNAL OF RADIOLOGY, 2004, 77 (923): : 944 - 949
  • [23] Ex Vivo Experimental Study of the Ablation Area of Bovine Liver Using STARmed Radiofrequency Ablation
    Ishikawa, Toru
    Hasegawa, Iori
    Hirosawa, Hiroshi
    Honmou, Tsubasa
    Sakai, Nobuyuki
    Igarashi, Takanori
    Yamazaki, Shun
    Kobayashi, Takamasa
    Sato, Toshifumi
    Iwanaga, Akito
    Sano, Tomoe
    Yokoyama, Junji
    Honma, Terasu
    IN VIVO, 2023, 37 (03): : 1047 - 1051
  • [24] Bipolar Radiofrequency Ablation for Nodular Thyroid Disease-Ex Vivo and In Vivo Evaluation of a Dose-Response Relationship
    Holmer, Christoph
    Lehmann, Kai S.
    Knappe, Verena
    Zurbuchen, Urte
    Frericks, Bernd
    Schumann, Thomas
    Buhr, Heinz J.
    Ritz, Joerg-Peter
    JOURNAL OF SURGICAL RESEARCH, 2011, 169 (02) : 234 - 240
  • [25] Ex vivo feasibility study of endoscopic intraductal laser ablation of the breast
    de Boorder, Tjeerd
    Waaijer, Laurien
    van Diest, Paul J.
    Witkamp, Arjen J.
    LASERS IN SURGERY AND MEDICINE, 2018, 50 (02) : 137 - 142
  • [26] Algorithm Optimization for Multitined Radiofrequency Ablation: Comparative Study in ex Vivo and in Vivo Bovine Liver
    Appelbaum, Liat
    Sosna, Jacob
    Pearson, Robert
    Perez, Sarah
    Nissenbaum, Yizhak
    Mertyna, Pawel
    Libson, Eugene
    Goldberg, S. Nahum
    RADIOLOGY, 2010, 254 (02) : 430 - 440
  • [27] Impedance, power, and current in radiofrequency ablation: Insights from technical, ex vivo, and clinical studies
    Bourier, Felix
    Ramirez, Francisco Daniel
    Martin, Claire A.
    Vlachos, Konstantinos
    Frontera, Antonio
    Takigawa, Masateru
    Kitamura, Takeshi
    Lam, Anna
    Duchateau, Josselin
    Pambrun, Thomas
    Cheniti, Ghassen
    Derval, Nicolas
    Denis, Arnaud
    Sacher, Frederic
    Hocini, Meleze
    Haissaguerre, Michel
    Jais, Pierre
    JOURNAL OF CARDIOVASCULAR ELECTROPHYSIOLOGY, 2020, 31 (11) : 2836 - 2845
  • [28] Bipolar radiofrequency ablation with 2 x 2 electrodes as a building block for matrix radiofrequency ablation: Ex vivo liver experiments and finite element method modelling
    Mulier, Stefaan
    Jiang, Yansheng
    Jamart, Jacques
    Wang, Chong
    Feng, Yuanbo
    Marchal, Guy
    Michel, Luc
    Ni, Yicheng
    INTERNATIONAL JOURNAL OF HYPERTHERMIA, 2015, 31 (06) : 649 - 665
  • [29] Comparison of wet radiofrequency ablation with dry radiofrequency ablation and radiofrequency ablation using hypertonic saline preinjection:: Ex vivo bovine liver
    Lee, JM
    Han, JK
    Kim, SH
    Shin, KS
    Lee, JY
    Park, HS
    Hur, H
    Choi, BI
    KOREAN JOURNAL OF RADIOLOGY, 2004, 5 (04) : 258 - 265
  • [30] Intrahepatic Radiofrequency Ablation Versus Electrochemical Treatment Ex Vivo
    Czymek, Ralf
    Loeffler, Stefan
    Dinter, Dorothea
    Gebhard, Maximilian
    Schmidt, Andreas
    Jungbluth, Thomas
    Kleemann, Markus
    Bruch, Hans-Peter
    Lubienski, Andreas
    JOURNAL OF SURGICAL RESEARCH, 2012, 174 (01) : 106 - 113