Motion Compensation for Ultrasound Thermal Imaging Using Motion-Mapped Reference Model: An in vivo Mouse Study

被引:13
|
作者
Seo, Joonho [1 ]
Kim, Sun Kwon [1 ]
Kim, Young-sun [2 ,3 ]
Choi, Kiwan [4 ]
Kong, Dong Geon [4 ]
Bang, Won-Chul [1 ]
机构
[1] Samsung Adv Inst Technol Samsung Elect, Seoul, South Korea
[2] Sungkyunkwan Univ, Sch Med, Samsung Med Ctr, Dept Radiol, Seoul, South Korea
[3] Sungkyunkwan Univ, Sch Med, Samsung Med Ctr, Ctr Imaging Sci, Seoul, South Korea
[4] Hlth & Med Equipment Samsung Elect, Seoul, South Korea
关键词
In vivo mouse experiment; motion compensation; ultrasound based thermal imaging; INDUCED ECHO-STRAIN; TEMPERATURE ESTIMATION; BACKSCATTERED ENERGY; VITRO; DEPENDENCE;
D O I
10.1109/TBME.2014.2325070
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Ultrasound (US)-based thermal imaging is very sensitive to tissue motion, which is a major obstacle to apply US temperature monitoring to noninvasive thermal therapies of in vivo subjects. In this study, we aim to develop a motion compensation method for stable US thermal imaging in in vivo subjects. Based on the assumption that the major tissue motion is approximately periodic caused by respiration, we propose a motion compensation method for change in backscattered energy (CBE) with multiple reference frames. Among the reference frames, the most similar reference to the current frame is selected to subtract the respiratory-induced motions. Since exhaustive reference searching in all stored reference frames can impede real-time thermal imaging, we improve the reference searching by using amotion-mapped reference model. We tested our method in six tumor-bearing mice with high intensity focused ultrasound (HIFU) sonication in the tumor volume until the temperature had increased by 7 degrees C. The proposed motion compensation was evaluated by root-mean-square-error (RMSE) analysis between the estimated temperature by CBE and the measured temperature by thermocouple. As a result, the mean +/- SD RMSE in the heating range was 1.1 +/- 0.1 degrees C with the proposed method, while the corresponding result without motion compensation was 4.3 +/- 2.6 degrees C. In addition, with the idea of motion-mapped reference frame, total processing time to produce a frame of thermal image was reduced in comparison with the exhaustive reference searching, which enabled themotioncompensated thermal imaging in 15 frames per second with 150 reference frames under 50% HIFU duty ratio.
引用
收藏
页码:2669 / 2678
页数:10
相关论文
共 50 条
  • [1] Verification and compensation of respiratory motion using an ultrasound imaging system
    Chuang, Ho-Chiao
    Hsu, Hsiao-Yu
    Chiu, Wei-Hung
    Tien, Der-Chi
    Wu, Ren-Hong
    Hsu, Chung-Hsien
    MEDICAL PHYSICS, 2015, 42 (03) : 1193 - 1199
  • [2] Adaptive motion compensation for in vivo ultrasound temperature estimation
    Bayat, Mahdi
    Ballard, John R.
    Ebbini, Emad S.
    2013 IEEE INTERNATIONAL ULTRASONICS SYMPOSIUM (IUS), 2013, : 1797 - 1800
  • [3] MOTION COMPENSATION USING GEOMETRY AND AN ELASTIC MOTION MODEL
    Muhit, Abdullah A.
    Pickering, Mark R.
    Frater, Michael R.
    2009 16TH IEEE INTERNATIONAL CONFERENCE ON IMAGE PROCESSING, VOLS 1-6, 2009, : 621 - 624
  • [4] Iterative Motion Compensation Approach for Ultrasonic Thermal Imaging
    Fleming, Ioana
    Hager, Gregory
    Guo, Xiaoyu
    Kang, Hyun Jae
    Boctor, Emad
    MEDICAL IMAGING 2015: ULTRASONIC IMAGING AND TOMOGRAPHY, 2015, 9419
  • [5] IN-VIVO MEASUREMENT OF VITREOUS MOTION USING AN ULTRASOUND IMAGING VELOCIMETRY TECHNIQUE
    Rossi, Tommaso
    Querzoli, Giorgio
    Repetto, Rodolfo
    Stocchino, Alessandro
    Pasqualitto, Giacomo
    Lossa, Mario
    Placentino, Luca
    PROCEEDINGS OF THE ASME SUMMER BIOENGINEERING CONFERENCE, PTS A AND B, 2012, : 1179 - 1180
  • [6] Using a Motion Capture System as Reference for Motion Tracking in Photoplethysmography Imaging
    Yu, Xinchi
    Cruz, Sonia
    Batista, Joao P.
    Bollheimer, Cornelius
    Leonhardt, Steffen
    Antink, Christoph Hoog
    2019 41ST ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2019, : 3915 - 3918
  • [7] Motion compensation for intravascular ultrasound palpography for in vivo vulnerable plaque detection
    Leung, KYE
    Baldewsing, RA
    Mastik, F
    Schaar, JA
    2005 IEEE ULTRASONICS SYMPOSIUM, VOLS 1-4, 2005, : 253 - 256
  • [8] Feasibility of Harmonic Motion Imaging Using a Single Transducer: In Vivo Imaging of Breast Cancer in a Mouse Model and Human Subjects
    Hossain, Md Murad
    Saharkhiz, Niloufar
    Konofagou, Elisa E.
    IEEE TRANSACTIONS ON MEDICAL IMAGING, 2021, 40 (05) : 1390 - 1404
  • [9] Extended Motion Compensation using Larger Blocks and an Elastic Motion Model
    Muhit, Abdullah A.
    Pickering, Mark R.
    Frater, Michael R.
    Arnold, John F.
    2008 IEEE 10TH WORKSHOP ON MULTIMEDIA SIGNAL PROCESSING, VOLS 1 AND 2, 2008, : 257 - 260
  • [10] Deadzone Compensation in Motion Control Systems Using Model Reference Direct Inverse Control
    Betancor-Martin, Carlos S.
    Montiel-Nelson, Juan A.
    Vega-Martinez, Aurelio
    2014 IEEE 57TH INTERNATIONAL MIDWEST SYMPOSIUM ON CIRCUITS AND SYSTEMS (MWSCAS), 2014, : 165 - 168