摘要
The use of time-of-flight (TOF) information during positron emission tomography (PET) reconstruction has been found to improve the image quality. In this work we quantified this improvement using two existing methods: 1) a very simple analytical expression only valid for a central point in a large uniform disk source and 2) efficient analytical approximations for postfiltered maximum likelihood expectation maximization (MLEM) reconstruction with a fixed target resolution, predicting the image quality in a pixel or in a small region of interest based on the Fisher information matrix. Using this latter method the weighting function for filtered backprojection reconstruction of TOF PET data proposed by C. Watson can be derived. The image quality was investigated at different locations in various software phantoms. Simplified as well as realistic phantoms, measured both with TOF PET systems and with a conventional PET system, were simulated. Since the time resolution of the system is not always accurately known, the effect on the image quality of using an inaccurate kernel during reconstruction was also examined with the Fisher information-based method. First, we confirmed with this method that the variance improvement in the center of a large uniform disk source is proportional to the disk diameter and inversely proportional to the time resolution. Next, image quality improvement was observed in all pixels, but in eccentric and high-count regions the contrast-to-noise ratio (CNR) increased less than in central and low- or medium-count regions. Finally, the CNR was seen to decrease when the time resolution was inaccurately modeled (too narrow or too wide) during reconstruction. Although the maximum CNR is not very sensitive to the time resolution error, using an inaccurate TOF kernel tends to introduce artifacts in the reconstructed image.
引用
收藏
页码:311 / 321
页数:11
相关论文
共 50 条
  • [1] Fisher information-based evaluation of image quality for time-of-flight PET
    Vunckx, Kathleen
    Zhou, Lin
    Matej, Samuel
    Defrise, Michel
    Nuyts, Johan
    2007 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-11, 2007, : 4129 - +
  • [2] Image quality evaluation on a new time-of-flight PET
    Turkington, Timothy
    Williams, John
    Wollenweber, Scott
    Stearns, Charles
    Ganin, Alexander
    Wilson, Joshua
    JOURNAL OF NUCLEAR MEDICINE, 2009, 50
  • [3] Evaluation of Image Quality on a Time-Of-Flight PET/CT Scanner
    Chang, T.
    Clark, J.
    Mawlawi, O.
    MEDICAL PHYSICS, 2011, 38 (06) : 3787 - +
  • [4] Evaluation of Windowed Image Reconstruction for Time-Of-Flight PET
    Guo, Jinxia
    Kao, Chien-Min
    Xie, Qinguo
    2011 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2011, : 4162 - 4166
  • [5] An evaluation of image noise variance for time-of-flight PET
    Watson, Charles C.
    2005 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD, VOLS 1-5, 2005, : 2041 - 2045
  • [6] An evaluation of image noise variance for time-of-flight PET
    Watson, Charles C.
    IEEE TRANSACTIONS ON NUCLEAR SCIENCE, 2007, 54 (05) : 1639 - 1647
  • [7] Fisher Information Guidance for Learned Time-of-Flight Imaging
    Li, Jiaqu
    Yue, Tao
    Zhao, Sijie
    Hu, Xuemei
    2022 IEEE/CVF CONFERENCE ON COMPUTER VISION AND PATTERN RECOGNITION (CVPR 2022), 2022, : 16313 - 16322
  • [8] Image quality characterization of time-of-flight information and point-spread-function recovery in PET imaging
    Gnesin, S.
    Delacoste, J.
    Martinez, P.
    Praz, C.
    Pappon, M.
    Prior, J. O.
    Baechler, S.
    Verdun, F. R.
    EUROPEAN JOURNAL OF NUCLEAR MEDICINE AND MOLECULAR IMAGING, 2013, 40 : S122 - S123
  • [9] Evaluation of Image Signal-to-Noise Ratio in Time-of-flight PET
    Clementel, Enrico
    Vandenberghe, Stefaan
    Karp, Joel S.
    Surti, Suleman
    2011 IEEE NUCLEAR SCIENCE SYMPOSIUM AND MEDICAL IMAGING CONFERENCE (NSS/MIC), 2011, : 3917 - 3921
  • [10] Time-of-flight precision and PET image accuracy
    Kolthammer, Jeffrey A.
    Tang, Jing
    Perkins, Amy E.
    Muzic, Raymond F., Jr.
    2010 IEEE NUCLEAR SCIENCE SYMPOSIUM CONFERENCE RECORD (NSS/MIC), 2010, : 3657 - 3660