Calibration transfer of near infrared spectrometers for the assessment of plasma ethanol precipitation process

被引:10
|
作者
Sun, Zhongyu [1 ,2 ]
Wang, Jiayue [1 ,2 ]
Nie, Lei [3 ]
Li, Lian [4 ]
Cao, Dawei [5 ]
Fan, Jiajin [5 ]
Wang, Haiyan [1 ,2 ]
Liu, Ruichen [1 ,2 ]
Zhang, Yinran [6 ]
Zang, Hengchang [1 ,2 ]
机构
[1] Shandong Univ, Sch Pharmaceut Sci, Wenhuaxi Rd 44, Jinan 250012, Shandong, Peoples R China
[2] Shandong Univ, Natl Glycoengn Res Ctr, Wenhuaxi Rd 44, Jinan 250012, Shandong, Peoples R China
[3] Shandong Univ, Sch Pharmaceut Sci & Pharmaceut Anal, Wenhuaxi Rd 44, Jinan 250012, Shandong, Peoples R China
[4] Shandong Univ, Sch Basic Med Sci, Wenhuaxi Rd 44, Jinan 250012, Shandong, Peoples R China
[5] Shandong Taibang Biol Prod Ltd Co, East Hushan Rd 14, Tai An 271000, Shandong, Peoples R China
[6] Ohio State Univ, Dept Chem & Biochem, Newman & Wolfram Lab, 100 West 18th Ave, Columbus, OH 43210 USA
基金
美国国家科学基金会;
关键词
Calibration transfer; Plasma ethanol precipitation; Near infrared spectroscopy (NIRS); Piecewise reverse standardization (PRS); Rank sample set partitioning based on joint x-y distance (Rank-SPXY); Process analytical technology (PAT); MULTIVARIATE CALIBRATION; CORRELATION-COEFFICIENT; ALCOHOL PRECIPITATION; FOURIER-TRANSFORM; NIR SPECTRA; MICRO-NIR; STANDARDIZATION; QUALITY; SPECTROSCOPY; INSTRUMENTS;
D O I
10.1016/j.chemolab.2018.08.012
中图分类号
TP [自动化技术、计算机技术];
学科分类号
0812 ;
摘要
Plasma ethanol precipitation process is an important step in the biopharmacy production process, and near infrared spectroscopy (NIBS) is an effective process analytical technology (PAT) tool in detecting the key quality attribute: the total protein (TP) content. But establishing a robust model is pretty complex. The aim of this study was to explore the calibration transfer strategy between master and slave spectrometers with the biopharmacy. Firstly, 8 batches of ethanol precipitation process were experimented. The changing spectra were collected by two spectrometers, reference values were measured by biuret method. Then, calibration transfer based on the PRS algorithm was carried out. Finally, before establishment of the optimal TP content partial least squares regression (PLSR) model of calibration transfer, rank sample set partitioning based on joint x-y distance (Rank-SPXY) and spectral pretreatment methods were selected after comparison. The determination coefficients of prediction (R-p(2)) and root mean squares error of prediction (RMSEP) were 0.934 and 0.8395 g/L, respectively. The results showed that Rank-SPXY-PRS was an effective calibration transfer strategy with different NIBS spectrometers. To some extent, the predictive capability of PLSR model was improved after calibration transfer, and it could provide reference for the further research in biopharmacy production process.
引用
收藏
页码:64 / 71
页数:8
相关论文
共 50 条
  • [41] Monitoring of the cellulosic ethanol fermentation process by near-infrared spectroscopy
    Pinto, Ariane S. S.
    Pereira, Sandra C.
    Ribeiro, Marcelo P. A.
    Farinas, Cristiane S.
    BIORESOURCE TECHNOLOGY, 2016, 203 : 334 - 340
  • [42] TEMPERATURE-COMPENSATING CALIBRATION TRANSFER FOR NEAR-INFRARED FILTER INSTRUMENTS
    WANG, YD
    KOWALSKI, BR
    ANALYTICAL CHEMISTRY, 1993, 65 (09) : 1301 - 1303
  • [43] Improvement of a standard-free method for near-infrared calibration transfer
    Tan, HW
    Sum, ST
    Brown, SD
    APPLIED SPECTROSCOPY, 2002, 56 (08) : 1098 - 1106
  • [44] An evaluation of orthogonal signal correction applied to calibration transfer of near infrared spectra
    Sjöblom, J
    Svensson, O
    Josefson, M
    Kullberg, H
    Wold, S
    CHEMOMETRICS AND INTELLIGENT LABORATORY SYSTEMS, 1998, 44 (1-2) : 229 - 244
  • [45] Calibration transfer of near-infrared spectra based on compression of wavelet coefficients
    Yoon, JG
    Lee, BW
    Han, CH
    CHEMOMETRICS AND INTELLIGENT LABORATORY SYSTEMS, 2002, 64 (01) : 1 - 14
  • [46] A calibration transfer optimized single kernel near-infrared spectroscopic method
    Xu, Zhuopin
    Fan, Shuang
    Liu, Jing
    Liu, Binmei
    Tao, Liangzhi
    Wu, Jin
    Hu, Shupeng
    Zhao, Liping
    Wang, Qi
    Wu, Yuejin
    SPECTROCHIMICA ACTA PART A-MOLECULAR AND BIOMOLECULAR SPECTROSCOPY, 2019, 220
  • [47] Transfer of multivariate calibrations between four near-infrared spectrometers using orthogonal signal correction
    Woody, NA
    Feudale, RN
    Myles, AJ
    Brown, SD
    ANALYTICAL CHEMISTRY, 2004, 76 (09) : 2595 - 2600
  • [48] A rare-Earth-oxide glass for the wavelength calibration of near-infrared dispersive and fourier-transform spectrometers
    Choquette, SJ
    Travis, JC
    O'Neal, LE
    Zhu, CJ
    Duewer, DL
    SPECTROSCOPY, 2001, 16 (04) : 14 - 19
  • [49] Development of calibration models for rapid determination of moisture content in rubber sheets using portable near-infrared spectrometers
    Puttipipatkajorn, Amorndej
    Puttipipatkajorn, Amornrit
    JOURNAL OF INNOVATIVE OPTICAL HEALTH SCIENCES, 2020, 13 (02)
  • [50] cSRM 2035: a rare earth oxide glass for the wavelength calibration of near infrared dispersive and Fourier transform spectrometers.
    Choquette, SJ
    Travis, JC
    Duewer, DL
    OPTICAL DIAGNOSTIC METHODS FOR INORGANIC TRANSMISSIVE MATERIALS, 1998, 3425 : 94 - 102