Evaluation of the long-term stability for the cylindrical ionization chambers

被引:0
|
作者
Lah, J.
Lim, C.
Suh, T.
机构
[1] Catholic Univ, Seoul, South Korea
[2] Korea Food & Drug Admin, Seoul, South Korea
关键词
D O I
10.1118/1.2760880
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To analyze the long‐term stability of Farmer‐type cylindrical ionization chambers using the calibration factor provided by the KFDA (Korea Food and Drug Administration). KFDA is the national secondary standard laboratory (SSDL) in Korea and approve as a member of the IAEA SSDL network. Materials and Methods: The cylindrical ionization chambers used in this study include PTW 30001(30006), 30013, 30002, 30004, 23333, Capintec PR06C, NE 2571, Exradin A12 and Wellhofer FC65G(IC70). The Nk and ND,W calibration factors for the cylindrical chambers were analyzed and the measured ND,W was compared with the calculated ND,W calibration factor. Results: The long‐term stability of PTW 30013(30006), Wellhofer FC65G(IC70) and NE 2571 varied within 0.2 %, meaning that they are more stable than other ionization chambers. This result is different from previous ones showing that the long term stability of ionization chambers with graphite wall material was superior to ionization chambers made of PMMA material. It turns out that almost the same levels of the stability of calibration factor were shown between PTW 30013(30006) ionization chamber made of PMMA wall material and Wellhofer FC65G(IC70) ionization chambers made of graphite material, and thus the former is shown to be superior from the aspect of long‐term safety, and PMMA wall material is firmer, more convenient to use, and it is more effective from the financial aspect. Therefore, it may be considered to be more appropriate to use for the regular dose measurements in comparison with graphite type ionization chamber. The measured ND,W calibration factor was approximately 1.0% higher than the calculated ND,W that was determined using the Nk calibration factor. Conclusion: The long‐term stability of the cylindrical chambers was evaluated using the Nk and ND,W calibration factor. It should help to improve clinical electron dosimetry in radiotherapy. © 2007, American Association of Physicists in Medicine. All rights reserved.
引用
收藏
页码:2452 / 2452
页数:1
相关论文
共 50 条
  • [41] Functional evaluation in orthodontic surgical treatment: long-term stability and predictability
    Giampietro Farronato
    Lucia Giannini
    Guido Galbiati
    Santo Andrea Stabilini
    Michele Sarcina
    Cinzia Maspero
    Progress in Orthodontics, 16
  • [42] In vitro evaluation of the long-term stability of polyimide as a material for neural implants
    Rubehn, Birthe
    Stieglitz, Thomas
    BIOMATERIALS, 2010, 31 (13) : 3449 - 3458
  • [43] Functional evaluation in orthodontic surgical treatment: long-term stability and predictability
    Farronato, Giampietro
    Giannini, Lucia
    Galbiati, Guido
    Stabilini, Santo Andrea
    Sarcina, Michele
    Maspero, Cinzia
    PROGRESS IN ORTHODONTICS, 2015, 16
  • [44] Neural Recording Chambers for Long-Term Access to Brain Tissue
    Pattadkal, Jagruti J.
    Priebe, Nicholas J.
    JOURNAL OF NEUROSCIENCE, 2025, 45 (07):
  • [45] Long-term stability of metal-envelope enclosed Bayard-Alpert ionization gauges
    Fedchak, James A.
    Defibaugh, Dana R.
    JOURNAL OF VACUUM SCIENCE & TECHNOLOGY A, 2012, 30 (06):
  • [46] Evaluation of ionization chambers for quality assurance
    Kim, CS
    Kim, CY
    Choi, MS
    PROCEEDINGS OF THE 22ND ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-4, 2000, 22 : 508 - 510
  • [47] LONG-TERM EVALUATION OF BSCS
    TAMIR, P
    AMERICAN BIOLOGY TEACHER, 1970, 32 (06): : 354 - &
  • [48] Long-term forecasting and evaluation
    Granger, Clive W. J.
    Jeon, Yongil
    INTERNATIONAL JOURNAL OF FORECASTING, 2007, 23 (04) : 539 - 551
  • [49] LONG-TERM EVALUATION OF TIMOLOL
    UUSITALO, RJ
    PALKAMA, A
    ACTA OPHTHALMOLOGICA, 1989, 67 (05): : 573 - 581
  • [50] A LONG-TERM EVALUATION OF CIMETIDINE
    WADE, AG
    JABEROO, D
    WALLACE, A
    PRACTITIONER, 1983, 227 (1378) : 642 - 645