Impact of dose reduction and the use of an advanced model-based iterative reconstruction algorithm on spectral performance of a dual-source CT system: A task-based image quality assessment

被引:24
|
作者
Greffier, Joel [1 ]
Dabli, Djamel [1 ]
Hamard, Aymeric [1 ]
Akessoul, Philippe [1 ]
Belaouni, Asmaa [1 ]
Beregi, Jean-Paul [1 ]
Frandon, Julien [1 ]
机构
[1] Univ Montpellier, Dept Med Imaging, Med Imaging Grp Nimes, EA 2992,CHU Nimes, Bd Pr Robert Debre, F-30029 Nimes 9, France
关键词
Multidetector computed tomography; Dual energy; Iterative reconstruction; Task-based image quality assessment; Phantom study; ENERGY CT; LIVER-LESIONS; OPTIMIZATION; TOMOGRAPHY; PULMONARY;
D O I
10.1016/j.diii.2021.03.002
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To assess the impact of dose reduction and the use of an advanced modeled iterative reconstruction algorithm (ADMIRE) on image quality in low-energy monochromatic images from a dual-source dual energy computed tomography CT (DSCT) platform. Materials and methods: Acquisitions on an image-quality phantom were performed using DSCT equipment with 100/Sn150 kVp for four dose levels (CTDvol: 20/11/8/5mGy). Raw data were reconstructed for six energy levels (40/50/60/70/80/100 keV) using filtered back projection and two levels of ADMIRE (A3/A5). Noise power spectrum (NPS) and task-based transfer function (TTF) were calculated on virtual monoenergetic images (VMIs). Detectability index (d') was computed to model the detection task of two enhanced iodine lesions as function of keV. Results: Noise-magnitude was significantly reduced between 40 to 70 keV by -56 +/- 0% (SD) (range: -56%--55%) with FBP; 56 +/- 0% (SD) (-56%--56%) with A3; and -57 +/- 1% (SD) (range: -57%--56%) with A5. The average spatial frequency of the NPS peaked at 70 keV and decreased as ADMIRE level increased. TTF values at 50% were greatest at 40 keV and shifted towards lower frequencies as the keV increased. The detectability of both lesions increased with increasing dose level and ADMIRE level. For the simulated lesion with iodine at 2 mg/mL, d' values peaked at 70 keV for all reconstruction types, except for A3 at 20 mGy and A5 at 11 and 20 mGy, where d' peaked at 60 keV. For the other simulated lesion, d' values were highest at 40 keV and decreased beyond. Conclusion: At low keV on VMIs, this study confirms that iterative reconstruction reduces the noise magnitude, improves the spatial resolution and increases the detectability of enhanced iodine lesions. (C) 2021 Societe francaise de radiologie. Published by Elsevier Masson SAS. All rights reserved.
引用
收藏
页码:405 / 412
页数:8
相关论文
共 50 条
  • [41] Image Quality and Dose Reduction Evaluation of a New CT Iterative Reconstruction Algorithm Using Model Observers
    Tseng, H.
    Fan, J.
    Kupinski, M.
    Sainath, P.
    Hsieh, J.
    MEDICAL PHYSICS, 2013, 40 (06)
  • [42] Task-Based Image Quality Assessment Comparing Classical and Iterative Cone Beam CT Images on Halcyon®
    Lassot-Buys, Marion
    Verstraet, Rodolfe
    Dabli, Djamel
    Moliner, Gilles
    Greffier, Joel
    DIAGNOSTICS, 2023, 13 (03)
  • [43] Improving image quality with model-based iterative reconstruction algorithm for chest CT in children with reduced contrast concentration
    Jihang Sun
    Di Hu
    Yun Shen
    Haiming Yang
    Chenghao Chen
    Jie Yin
    Yun Peng
    La radiologia medica, 2019, 124 : 595 - 601
  • [44] Improving image quality with model-based iterative reconstruction algorithm for chest CT in children with reduced contrast concentration
    Sun, Jihang
    Hu, Di
    Shen, Yun
    Yang, Haiming
    Chen, Chenghao
    Yin, Jie
    Peng, Yun
    RADIOLOGIA MEDICA, 2019, 124 (07): : 595 - 601
  • [45] Influence of model based iterative reconstruction algorithm on image quality of multiplanar reformations in reduced dose chest CT
    Barras, Heloise
    Dunet, Vincent
    Hachulla, Anne-Lise
    Grimm, Jochen
    Beigelman-Aubry, Catherine
    ACTA RADIOLOGICA OPEN, 2016, 5 (08)
  • [46] Task-based characterization of a deep learning image reconstruction and comparison with filtered back-projection and a partial model-based iterative reconstruction in abdominal CT: A phantom study
    Racine, Damien
    Becce, Fabio
    Viry, Anais
    Monnin, Pascal
    Thomsen, Brian
    Verdun, Francis R.
    Rotzinger, David C.
    PHYSICA MEDICA-EUROPEAN JOURNAL OF MEDICAL PHYSICS, 2020, 76 : 28 - 37
  • [47] Task-Based Optimal Dose Partitioning in Dual-Source Dual-Energy CT for Virtual Non-Contrast and Iodine Imaging
    Long, L.
    McCollough, C.
    Yu, L.
    MEDICAL PHYSICS, 2018, 45 (06) : E675 - E675
  • [48] Virtual monoenergetic images from dual-energy CT: systematic assessment of task-based image quality performance
    Cester, Davide
    Eberhard, Matthias
    Alkadhi, Hatem
    Euler, Andre
    QUANTITATIVE IMAGING IN MEDICINE AND SURGERY, 2022, 12 (01) : 726 - +
  • [49] Performance of four dual-energy CT platforms for abdominal imaging: a task-based image quality assessment based on phantom data
    J. Greffier
    S. Si-Mohamed
    D. Dabli
    H. de Forges
    A. Hamard
    P. Douek
    J. P. Beregi
    J. Frandon
    European Radiology, 2021, 31 : 5324 - 5334
  • [50] ACCURATE MODEL-BASED HIGH RESOLUTION CARDIAC IMAGE RECONSTRUCTION IN DUAL SOURCE CT
    Do, Synho
    Cho, Sanghee
    Karl, W. Clem
    Kalra, Mannudeep K.
    Brady, Thomas J.
    Pien, Homer
    2009 IEEE INTERNATIONAL SYMPOSIUM ON BIOMEDICAL IMAGING: FROM NANO TO MACRO, VOLS 1 AND 2, 2009, : 330 - +