Impact of dose reduction and the use of an advanced model-based iterative reconstruction algorithm on spectral performance of a dual-source CT system: A task-based image quality assessment

被引:24
|
作者
Greffier, Joel [1 ]
Dabli, Djamel [1 ]
Hamard, Aymeric [1 ]
Akessoul, Philippe [1 ]
Belaouni, Asmaa [1 ]
Beregi, Jean-Paul [1 ]
Frandon, Julien [1 ]
机构
[1] Univ Montpellier, Dept Med Imaging, Med Imaging Grp Nimes, EA 2992,CHU Nimes, Bd Pr Robert Debre, F-30029 Nimes 9, France
关键词
Multidetector computed tomography; Dual energy; Iterative reconstruction; Task-based image quality assessment; Phantom study; ENERGY CT; LIVER-LESIONS; OPTIMIZATION; TOMOGRAPHY; PULMONARY;
D O I
10.1016/j.diii.2021.03.002
中图分类号
R8 [特种医学]; R445 [影像诊断学];
学科分类号
1002 ; 100207 ; 1009 ;
摘要
Purpose: To assess the impact of dose reduction and the use of an advanced modeled iterative reconstruction algorithm (ADMIRE) on image quality in low-energy monochromatic images from a dual-source dual energy computed tomography CT (DSCT) platform. Materials and methods: Acquisitions on an image-quality phantom were performed using DSCT equipment with 100/Sn150 kVp for four dose levels (CTDvol: 20/11/8/5mGy). Raw data were reconstructed for six energy levels (40/50/60/70/80/100 keV) using filtered back projection and two levels of ADMIRE (A3/A5). Noise power spectrum (NPS) and task-based transfer function (TTF) were calculated on virtual monoenergetic images (VMIs). Detectability index (d') was computed to model the detection task of two enhanced iodine lesions as function of keV. Results: Noise-magnitude was significantly reduced between 40 to 70 keV by -56 +/- 0% (SD) (range: -56%--55%) with FBP; 56 +/- 0% (SD) (-56%--56%) with A3; and -57 +/- 1% (SD) (range: -57%--56%) with A5. The average spatial frequency of the NPS peaked at 70 keV and decreased as ADMIRE level increased. TTF values at 50% were greatest at 40 keV and shifted towards lower frequencies as the keV increased. The detectability of both lesions increased with increasing dose level and ADMIRE level. For the simulated lesion with iodine at 2 mg/mL, d' values peaked at 70 keV for all reconstruction types, except for A3 at 20 mGy and A5 at 11 and 20 mGy, where d' peaked at 60 keV. For the other simulated lesion, d' values were highest at 40 keV and decreased beyond. Conclusion: At low keV on VMIs, this study confirms that iterative reconstruction reduces the noise magnitude, improves the spatial resolution and increases the detectability of enhanced iodine lesions. (C) 2021 Societe francaise de radiologie. Published by Elsevier Masson SAS. All rights reserved.
引用
收藏
页码:405 / 412
页数:8
相关论文
共 50 条
  • [31] Ultra-low-dose lung screening CT with model-based iterative reconstruction: an assessment of image quality and lesion conspicuity
    Ju, Yun Hye
    Lee, Geewon
    Lee, Jiwon
    Hong, Seung Baek
    Suh, Young Ju
    Jeong, Yeon Joo
    ACTA RADIOLOGICA, 2018, 59 (05) : 553 - 559
  • [32] Application of low-dose CT combined with model-based iterative reconstruction algorithm in oncologic patients during follow-up: dose reduction and image quality
    Ippolito, Davide
    Maino, Cesare
    Pecorelli, Anna
    Salemi, Ilaria
    Gandola, Davide
    Riva, Luca
    Franzesi, Cammillo Talei
    Sironi, Sandro
    BRITISH JOURNAL OF RADIOLOGY, 2021, 94 (1124):
  • [33] Model-based iterative reconstruction for reduction of radiation dose in abdominopelvic CT: comparison to adaptive statistical iterative reconstruction
    Yasaka, Koichiro
    Katsura, Masaki
    Akahane, Masaaki
    Sato, Jiro
    Matsuda, Izuru
    Ohtomo, Kuni
    SPRINGERPLUS, 2013, 2 : 1 - 9
  • [34] Clinical impact of model-based type iterative reconstruction with fast reconstruction time on image quality of low-dose screening chest CT
    Yuki, Hideaki
    Oda, Seitaro
    Utsunomiya, Daisuke
    Funama, Yoshinori
    Kidoh, Masafumi
    Namimoto, Tomohiro
    Katahira, Kazuhiro
    Honda, Keiichi
    Tokuyasu, Shinichi
    Yamashita, Yasuyuki
    ACTA RADIOLOGICA, 2016, 57 (03) : 295 - 302
  • [35] Radiation Dose Reduction in Paranasal Sinus CT Using Model-Based Iterative Reconstruction
    Hoxworth, J. M.
    Lal, D.
    Fletcher, G. P.
    Patel, A. C.
    He, M.
    Paden, R. G.
    Hara, A. K.
    AMERICAN JOURNAL OF NEURORADIOLOGY, 2014, 35 (04) : 644 - 649
  • [36] Hybrid iterative reconstruction algorithm in brain CT: a radiation dose reduction and image quality assessment study
    Love, Askell
    Siemund, Roger
    Hoglund, Peter
    Van Westen, Danielle
    Stenberg, Lars
    Petersen, Cecilia
    Bjorkman-Burtscher, Isabella M.
    ACTA RADIOLOGICA, 2014, 55 (02) : 208 - 217
  • [37] A Novel Phantom for CT Performance Assessment: Towards a Task-Based Measure of Image Quality
    Richard, S.
    Samei, E.
    MEDICAL PHYSICS, 2011, 38 (06)
  • [38] Noise performance studies of model-based iterative reconstruction (MBIR) as a function of kV, mA and exposure level: Impact on radiation dose reduction and image quality
    Gomez-Cardona, Daniel
    Li, Ke
    Lubner, Meghan G.
    Pickhardt, Perry J.
    Chen, Guang-Hong
    MEDICAL IMAGING 2015: PHYSICS OF MEDICAL IMAGING, 2015, 9412
  • [39] Modeling Nonstationary Noise and Task-Based Detectability in CT Images Computed by Filtered Backprojection and Model-Based Iterative Reconstruction
    Gang, G.
    Stayman, J.
    Zbijewski, W.
    Siewerdsen, J.
    MEDICAL PHYSICS, 2013, 40 (06)
  • [40] The Impact of Dose Reduction in Quantitative Kinematic CT of Ankle Joints Using a Full Model-Based Iterative Reconstruction Algorithm: A Cadaveric Study
    Lombard, Charles
    Gervaise, Alban
    Villani, Nicolas
    Louis, Mathias
    Raymond, Ariane
    Blum, Alain
    Teixeira, Pedro Augusto Gondim
    AMERICAN JOURNAL OF ROENTGENOLOGY, 2018, 210 (02) : 396 - 403