Perfusion measurements by micro-CT using prior image constrained compressed sensing (PICCS): initial phantom results

被引:44
|
作者
Nett, Brian E. [1 ]
Brauweiler, Robert [2 ]
Kalender, Willi [2 ]
Rowley, Howard [3 ]
Chen, Guang-Hong [1 ,3 ]
机构
[1] Univ Wisconsin, Dept Med Phys, Madison, WI 53706 USA
[2] Univ Erlangen Nurnberg, Inst Med Phys, D-8520 Erlangen, Germany
[3] Univ Wisconsin, Dept Radiol, Madison, WI 53706 USA
来源
PHYSICS IN MEDICINE AND BIOLOGY | 2010年 / 55卷 / 08期
基金
美国国家卫生研究院;
关键词
HIGH-RESOLUTION MEASUREMENT; BRAIN TRANSFER CONSTANTS; TRACER BOLUS PASSAGES; QUANTITATIVE ASSESSMENT; COMPUTED-TOMOGRAPHY; TUMOR PERFUSION; DYNAMIC CT; PERFORMANCE EVALUATION; GRAPHICAL EVALUATION; CEREBRAL PERFUSION;
D O I
10.1088/0031-9155/55/8/014
中图分类号
R318 [生物医学工程];
学科分类号
0831 ;
摘要
Micro-CT scanning has become an accepted standard for anatomical imaging in small animal disease and genome mutation models. Concurrently, perfusion imaging via tracking contrast dynamics after injection of an iodinated contrast agent is a well-established tool for clinical CT scanners. However, perfusion imaging is not yet commercially available on the micro-CT platform due to limitations in both radiation dose and temporal resolution. Recent hardware developments in micro-CT scanners enable continuous imaging of a given volume through the use of a slip-ring gantry. Now that dynamic CT imaging is feasible, data may be acquired to measure tissue perfusion using a micro-CT scanner (CT Imaging, Erlangen, Germany). However, rapid imaging using micro-CT scanners leads to high image noise in individual time frames. Using the standard filtered backprojection (FBP) image reconstruction, images are prohibitively noisy for calculation of voxel-by-voxel perfusion maps. In this study, we apply prior image constrained compressed sensing (PICCS) to reconstruct images with significantly lower noise variance. In perfusion phantom experiments performed on a micro-CT scanner, the PICCS reconstruction enabled a reduction to 1/16 of the noise variance of standard FBP reconstruction, without compromising the spatial or temporal resolution. This enables a significant increase in dose efficiency, and thus, significantly less exposure time is needed to acquire images amenable to perfusion processing. This reduction in required irradiation time enables voxel-by-voxel perfusion maps to be generated on micro-CT scanners. Sample perfusion maps using a deconvolution-based perfusion analysis are included to demonstrate the improvement in image quality using the PICCS algorithm.
引用
收藏
页码:2333 / 2350
页数:18
相关论文
共 50 条
  • [41] Perfusion Measurements in Micro-CT: Increased Dose Efficiency by Use of HYPR LR Image Reconstruction
    Brauweiler, Robert
    Ertel, Dirk
    Keith, Lauren
    Speidel, Michael A.
    WORLD CONGRESS ON MEDICAL PHYSICS AND BIOMEDICAL ENGINEERING, VOL 25, PT 2 - DIAGNOSTIC IMAGING, 2009, 25 : 60 - 63
  • [42] Prior Image Constrained Compressed Sensing Metal Artifact Reduction (PICCS-MAR): 2D and 3D Image Quality Improvement with Hip Prostheses at CT Colonography
    Bannas, Peter
    Li, Yinsheng
    Motosugi, Utaroh
    Li, Ke
    Lubner, Meghan
    Chen, Guang-Hong
    Pickhardt, Perry J.
    EUROPEAN RADIOLOGY, 2016, 26 (07) : 2039 - 2046
  • [43] Prior Image Constrained Compressed Sensing Metal Artifact Reduction (PICCS-MAR): 2D and 3D Image Quality Improvement with Hip Prostheses at CT Colonography
    Peter Bannas
    Yinsheng Li
    Utaroh Motosugi
    Ke Li
    Meghan Lubner
    Guang-Hong Chen
    Perry J. Pickhardt
    European Radiology, 2016, 26 : 2039 - 2046
  • [44] Reduced Image Noise at Low-Dose Multidetector CT of the Abdomen with Prior Image Constrained Compressed Sensing Algorithm
    Lubner, Meghan G.
    Pickhardt, Perry J.
    Tang, Jie
    Chen, Guang-Hong
    RADIOLOGY, 2011, 260 (01) : 248 - 256
  • [45] Enhancement of 4D CBCT Image Quality Using An Adaptive Prior Image Constrained Compressed Sensing
    Lee, H.
    Yoon, J.
    Lee, E.
    Cho, S.
    Park, K.
    Choi, W.
    Keum, K.
    MEDICAL PHYSICS, 2015, 42 (06) : 3639 - 3639
  • [46] Radiation dose reduction in dual-energy CT using Prior Image Constrained Compressed Sensing: Image quality evaluation in virtual monochromatic imaging
    Brunner, Stephen
    Garrett, John
    Li, Yinsheng
    Tang, Tie
    Pulfer, Kari
    Hsieh, Jiang
    Rowley, Howard
    Chen, Guang-Hong
    MEDICAL IMAGING 2013: PHYSICS OF MEDICAL IMAGING, 2013, 8668
  • [47] Non-Coplanar Cone Beam CT Reconstruction with Limited Angle of Projections Using Rigid Image Registration Facilitated Prior Image Constrained Compressed Sensing
    Meng, B.
    Li, R.
    Han, B.
    Cheng, J.
    Xing, L.
    MEDICAL PHYSICS, 2013, 40 (06)
  • [48] Non-Coplanar Cone Beam CT Reconstruction With Limited Angle of Projections Using Rigid Image Registration Facilitated Prior Image Constrained Compressed Sensing
    Meng, B.
    Li, R.
    Han, B.
    Chen, J.
    Xing, L.
    INTERNATIONAL JOURNAL OF RADIATION ONCOLOGY BIOLOGY PHYSICS, 2013, 87 (02): : S146 - S146
  • [49] Tomosynthesis via total variation minimization reconstruction and prior image constrained compressed sensing (PICCS) on a C-arm system - art. no. 69132D
    Nett, Brian
    Tang, Jie
    Leng, Shuai
    Chen, Guang-Hong
    MEDICAL IMAGING 2008: PHYSICS OF MEDICAL IMAGING, PTS 1-3, 2008, 6913 : D9132 - D9132
  • [50] Spectral Prior Image Constrained Compressed Sensing Reconstruction for Photon-Counting Detector Based CT Using a Non-Local Means Filtered Prior (NLM-SPICCS)
    Rajendran, K.
    Tao, S.
    Zhou, W.
    Leng, S.
    McCollough, C.
    MEDICAL PHYSICS, 2018, 45 (06) : E675 - E675