MULTIDIMENSIONAL ULTRASOUND DOPPLER SIGNAL ANALYSIS FOR FETAL ACTIVITY MONITORING

被引:4
|
作者
Ribes, Sophie [1 ]
Girault, Jean-Marc [2 ]
Perrotin, Franck [3 ]
Kouame, Denis [1 ]
机构
[1] Univ Toulouse 3, CNRS, IRIT, UMR 5505, F-31062 Toulouse, France
[2] Univ Tours, INSERM, U930, Tours, France
[3] INSERM, CHU Bretonneau, U930, Tours Serv Gynecol Obstet, Tours, France
来源
ULTRASOUND IN MEDICINE AND BIOLOGY | 2015年 / 41卷 / 12期
关键词
Fetal monitoring; Fetal heart rhythm; Fetal movement; Multidimensional signals; Support vector machine; HEART-RATE; SYSTEM;
D O I
10.1016/j.ultrasmedbio.2015.07.026
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
Fetal activity parameters such as movements, heart rate and the related parameters are essential indicators of fetal wellbeing, and no device provides simultaneous access to and sufficient estimation of all of these parameters to evaluate fetal health. This work was aimed at collecting these parameters to automatically separate healthy from compromised fetuses. To achieve this goal, we first developed a multi-sensor-multi-gate Doppler system. Then we recorded multidimensional Doppler signals and estimated the fetal activity parameters via dedicated signal processing techniques. Finally, we combined these parameters into four sets of parameters (or four hyperparameters) to determine the set of parameters that is able to separate healthy from other fetuses. To validate our system, a data set consisting of two groups of fetal signals (normal and compromised) was established and provided by physicians. From the estimated parameters, an instantaneous Manning-like score, referred to as the ultrasonic score, was calculated and was used together with movements, heart rate and the associated parameters in a classification process employing the support vector machine method. We investigated the influence of the sets of parameters and evaluated the performance of the support vector machine using the computation of sensibility, specificity, percentage of support vectors and total classification error. The sensitivity of the four sets ranged from 79% to 100%. Specificity was 100% for all sets. The total classification error ranged from 0% to 20%. The percentage of support vectors ranged from 33% to 49%. Overall, the best results were obtained with the set of parameters consisting of fetal movement, short-term variability, long-term variability, deceleration and ultrasound score. The sensitivity, specificity, percentage of support vectors and total classification error of this set were respectively 100%, 100%, 35% and 0%. This indicated our ability to separate the data into two sets (normal fetuses and pathologic fetuses), and the results highlight the excellent match with the clinical classification performed by the physicians. This work indicates the feasibility of detecting compromised fetuses and also represents an interesting method of close fetal monitoring during the entire pregnancy. (C) 2015 World Federation for Ultrasound in Medicine & Biology.
引用
收藏
页码:3172 / 3181
页数:10
相关论文
共 50 条
  • [21] A novel technique for fetal heart rate estimation from Doppler ultrasound signal
    Jezewski, Janusz
    Roj, Dawid
    Wrobel, Janusz
    Horoba, Krzysztof
    BIOMEDICAL ENGINEERING ONLINE, 2011, 10
  • [22] Study and Implementation of a Signal Processing Algorithm for Doppler Ultrasound Fetal Heart Signals
    Liu, Chuqi
    Ji, Chenpeng
    Lin, Xiaoqiang
    Yao, Jian
    Huang, Hai
    PROCEEDINGS OF THE 2019 14TH IEEE CONFERENCE ON INDUSTRIAL ELECTRONICS AND APPLICATIONS (ICIEA 2019), 2019, : 1402 - 1407
  • [23] ANALYSIS OF THE DOPPLER ULTRASOUND SIGNAL BY WAVELET PACKET TRANSFORM
    Meghnefi, F.
    Reguig, F. Bereksi
    Brahami, M.
    JOURNAL OF MECHANICS IN MEDICINE AND BIOLOGY, 2004, 4 (03) : 273 - 282
  • [24] DOPPLER ULTRASOUND AND FETAL HAZARD
    SMYTHE, GE
    MACRAE, DJ
    LANCET, 1975, 2 (7925): : 134 - 134
  • [25] On the wavelet analysis of the Doppler signals of fetal activity
    Kaluzynski, Krzysztof
    Kret, Tomasz
    Sienko, Jacek
    PROCEEDINGS OF THE FIFTH IASTED INTERNATIONAL CONFERENCE ON BIOMEDICAL ENGINEERING, 2007, : 180 - 185
  • [26] Physical activity and maternal-fetal circulation measured by Doppler ultrasound
    Nguyen, N. C.
    Evenson, K. R.
    Savitz, D. A.
    Chu, H.
    Thorp, J. M.
    Daniels, J. L.
    JOURNAL OF PERINATOLOGY, 2013, 33 (02) : 87 - 93
  • [27] Editorial: Recent Advances in Doppler Signal Processing and Modeling Techniques for Fetal Monitoring
    Khandoker, Ahsan H.
    Marzbanrad, Faezeh
    Kimura, Yoshitaka
    FRONTIERS IN PHYSIOLOGY, 2018, 9
  • [28] DOPPLER-ULTRASOUND ANALYSIS OF CARDIAC ACTIVITY
    SEIPEL, L
    GLEICHMANN, U
    ZEITSCHRIFT FUR KREISLAUFFORSCHUNG, 1969, 58 (12): : 1291 - +
  • [29] Blood and wall signal simulator for Doppler ultrasound signal analysis algorithm development
    Fish, PJ
    Bastos, CAC
    Vaz, F
    PROCEEDINGS OF THE 23RD ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-4: BUILDING NEW BRIDGES AT THE FRONTIERS OF ENGINEERING AND MEDICINE, 2001, 23 : 135 - 138
  • [30] DOPPLER ULTRASOUND SIGNAL ANALYSIS BASED ON THE TMS320 SIGNAL PROCESSOR
    MURPHY, C
    COATS, A
    CONWAY, J
    COLDITZ, P
    ROLFE, P
    JOURNAL OF BIOMEDICAL ENGINEERING, 1988, 10 (02): : 127 - 129