MULTIDIMENSIONAL ULTRASOUND DOPPLER SIGNAL ANALYSIS FOR FETAL ACTIVITY MONITORING

被引:4
|
作者
Ribes, Sophie [1 ]
Girault, Jean-Marc [2 ]
Perrotin, Franck [3 ]
Kouame, Denis [1 ]
机构
[1] Univ Toulouse 3, CNRS, IRIT, UMR 5505, F-31062 Toulouse, France
[2] Univ Tours, INSERM, U930, Tours, France
[3] INSERM, CHU Bretonneau, U930, Tours Serv Gynecol Obstet, Tours, France
来源
ULTRASOUND IN MEDICINE AND BIOLOGY | 2015年 / 41卷 / 12期
关键词
Fetal monitoring; Fetal heart rhythm; Fetal movement; Multidimensional signals; Support vector machine; HEART-RATE; SYSTEM;
D O I
10.1016/j.ultrasmedbio.2015.07.026
中图分类号
O42 [声学];
学科分类号
070206 ; 082403 ;
摘要
Fetal activity parameters such as movements, heart rate and the related parameters are essential indicators of fetal wellbeing, and no device provides simultaneous access to and sufficient estimation of all of these parameters to evaluate fetal health. This work was aimed at collecting these parameters to automatically separate healthy from compromised fetuses. To achieve this goal, we first developed a multi-sensor-multi-gate Doppler system. Then we recorded multidimensional Doppler signals and estimated the fetal activity parameters via dedicated signal processing techniques. Finally, we combined these parameters into four sets of parameters (or four hyperparameters) to determine the set of parameters that is able to separate healthy from other fetuses. To validate our system, a data set consisting of two groups of fetal signals (normal and compromised) was established and provided by physicians. From the estimated parameters, an instantaneous Manning-like score, referred to as the ultrasonic score, was calculated and was used together with movements, heart rate and the associated parameters in a classification process employing the support vector machine method. We investigated the influence of the sets of parameters and evaluated the performance of the support vector machine using the computation of sensibility, specificity, percentage of support vectors and total classification error. The sensitivity of the four sets ranged from 79% to 100%. Specificity was 100% for all sets. The total classification error ranged from 0% to 20%. The percentage of support vectors ranged from 33% to 49%. Overall, the best results were obtained with the set of parameters consisting of fetal movement, short-term variability, long-term variability, deceleration and ultrasound score. The sensitivity, specificity, percentage of support vectors and total classification error of this set were respectively 100%, 100%, 35% and 0%. This indicated our ability to separate the data into two sets (normal fetuses and pathologic fetuses), and the results highlight the excellent match with the clinical classification performed by the physicians. This work indicates the feasibility of detecting compromised fetuses and also represents an interesting method of close fetal monitoring during the entire pregnancy. (C) 2015 World Federation for Ultrasound in Medicine & Biology.
引用
收藏
页码:3172 / 3181
页数:10
相关论文
共 50 条
  • [1] Compressive-Sensing-based multidimensional Doppler signal analysis for fetal activity monitoring
    Basarab, Adrian
    Quinsac, Celine
    Girault, Jean-Marc
    Kouame, Denis
    2014 IEEE INTERNATIONAL ULTRASONICS SYMPOSIUM (IUS), 2014, : 1073 - 1076
  • [2] Timing events in Doppler ultrasound signal of fetal heart activity
    Kupka, T
    Jezewski, J
    Matonia, A
    Horoba, K
    Wróbel, J
    PROCEEDINGS OF THE 26TH ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY, VOLS 1-7, 2004, 26 : 337 - 340
  • [3] DOPPLER ULTRASOUND AND FETAL ACTIVITY
    DAVID, H
    WEAVER, JB
    PEARSON, JF
    BRITISH MEDICAL JOURNAL, 1975, 2 (5962): : 62 - 64
  • [4] Template-based Quality Assessment of the Doppler Ultrasound Signal for Fetal Monitoring
    Valderrama, Camilo E.
    Marzbanrad, Faezeh
    Stroux, Lisa
    Clifford, Gari D.
    FRONTIERS IN PHYSIOLOGY, 2017, 8
  • [5] INFLUENCE OF DOPPLER ULTRASOUND ON FETAL ACTIVITY
    MURRILLS, AJ
    BARRINGTON, P
    HARRIS, PD
    WHEELER, T
    BRITISH MEDICAL JOURNAL, 1983, 286 (6370): : 1009 - 1012
  • [6] NEW TECHNIQUE FOR IMPROVING DOPPLER ULTRASOUND SIGNAL FOR FETAL HEART-RATE MONITORING
    LAUERSEN, NH
    HOCHBERG, HM
    GEORGE, MED
    TEGGE, CS
    MEIGHAN, JJ
    AMERICAN JOURNAL OF OBSTETRICS AND GYNECOLOGY, 1977, 128 (03) : 300 - 302
  • [7] DOPPLER ULTRASOUND AND SUBJECTIVE ASSESSMENT OF FETAL ACTIVITY
    PHILLIPS, WDP
    TOWELL, ME
    BRITISH MEDICAL JOURNAL, 1979, 2 (6182): : 101 - 102
  • [8] Fetal Cardiac Timing Events Estimation from Doppler Ultrasound Signal Cepstrum Analysis
    Alnuaimi, Saeed
    Jimaa, Shihab
    Kimura, Yoshitaka
    Hadjileontiadis, Leontios J.
    Khandoker, Ahsan H.
    2019 41ST ANNUAL INTERNATIONAL CONFERENCE OF THE IEEE ENGINEERING IN MEDICINE AND BIOLOGY SOCIETY (EMBC), 2019, : 4677 - 4681
  • [9] Fetal activity parameters classification from multidimensional Doppler signals
    Ribes, S.
    Voicu, I.
    Quinsac, C.
    Girault, J. -M.
    Fournier-Massignan, M.
    Perrotin, F.
    Kouame, D.
    IRBM, 2011, 32 (02) : 152 - 154
  • [10] Doppler Ultrasound Technology for Fetal Heart Rate Monitoring: A Review
    Hamelmann, Paul
    Vullings, Rik
    Kolen, Alexander F.
    Bergmans, Jan W. M.
    van Laar, Judith O. E. H.
    Tortoli, Piero
    Mischi, Massimo
    IEEE TRANSACTIONS ON ULTRASONICS FERROELECTRICS AND FREQUENCY CONTROL, 2020, 67 (02) : 226 - 238