Displacement of the epiglottis during intubation with the Pentax-AWS Airway Scope

被引:5
|
作者
Suzuki, Akihiro [1 ]
Katsumi, Norifumi [1 ]
Honda, Takashi [1 ]
Sasakawa, Tomoki [1 ]
Kunisawa, Takayuki [1 ]
Henderson, John J. [2 ]
Iwasaki, Hiroshi [1 ]
机构
[1] Asahikawa Med Coll, Dept Anesthesiol & Crit Care Med, Asahikawa, Hokkaido 0788510, Japan
[2] Gartnavel Royal Hosp, Dept Anaesthesia, Glasgow, Lanark, Scotland
关键词
Pentax-AWS; Airway Scope; Folding epiglottis; Epiglottis displacement; TRACHEAL INTUBATION; LARYNGOSCOPE;
D O I
10.1007/s00540-009-0826-4
中图分类号
R614 [麻醉学];
学科分类号
100217 ;
摘要
The Pentax-AWS system is a rigid indirect video laryngoscope with integrated tube guidance. Complications associated with this device are not well understood. We report two cases of epiglottis malposition during intubation with the Pentax-AWS. The standard technique of using the Pentax-AWS system involves direct elevation of the epiglottis for exposure of the vocal cords. The blade tip should be passed posterior to the epiglottis for laryngeal exposure, but pressure on the anterior surface of the epiglottis by the tip can rarely happen even during the correct maneuver. Although the Pentax-AWS provides clear images of the airway structures, it is sometimes difficult to observe the epiglottis continuously because the camera is located beneath the blade tip. Consequently, the view of the epiglottis from the camera may be impeded by the blade tip and may result in undiagnosed epiglottis malposition. The AWS's structural feature and its approach to the larynx can be associated with increased chance of unexpected epiglottis folding. It is particularly important to confirm normal position of the epiglottis during withdrawal of the device to prevent this complication.
引用
收藏
页码:124 / 127
页数:4
相关论文
共 50 条
  • [1] Displacement of the epiglottis during intubation with the Pentax-AWS Airway Scope
    Akihiro Suzuki
    Norifumi Katsumi
    Takashi Honda
    Tomoki Sasakawa
    Takayuki Kunisawa
    John J. Henderson
    Hiroshi Iwasaki
    Journal of Anesthesia, 2010, 24 : 124 - 127
  • [2] Pentax-AWS (Airway Scope®) for awake tracheal intubation
    Suzuki, Akihiro
    Kunisawa, Takayuki
    Takahata, Osamu
    Iwasaki, Hiroshi
    Nozaki, Koji
    Henderson, John J.
    JOURNAL OF CLINICAL ANESTHESIA, 2007, 19 (08) : 642 - 643
  • [3] Cardiovascular responses to tracheal intubation with the Airway Scope (Pentax-AWS)
    Akihiro Suzuki
    Yuki Toyama
    Norifumi Katsumi
    Takayuki Kunisawa
    John J. Henderson
    Hiroshi Iwasaki
    Journal of Anesthesia, 2008, 22 : 100 - 101
  • [4] Cardiovascular responses to tracheal intubation with the Airway Scope (Pentax-AWS)
    Suzuki, Akihiro
    Toyama, Yuki
    Katsumi, Norifumi
    Kunisawa, Takayuki
    Henderson, John J.
    Iwasaki, Hiroshi
    JOURNAL OF ANESTHESIA, 2008, 22 (01) : 100 - 101
  • [5] Efficacy of the Airway Scope (Pentax-AWS) for training in pediatric intubation
    Sakurai, Yoshio
    Tamura, Masanori
    PEDIATRICS INTERNATIONAL, 2015, 57 (02) : 217 - 221
  • [6] Airway management with pentax airway scope (pentax-aws®). Complications and incidents
    Garcia-Alvarez, Raquel
    Stolle, David
    Abad, Ane
    Oliden, Leticia
    Gonzalez, Lucio
    BRITISH JOURNAL OF ANAESTHESIA, 2012, 108 : 439 - 440
  • [7] Evaluation of tracheal intubation in several positions by the Pentax-AWS Airway Scope: a manikin study
    Nobuyasu Komasawa
    Ryusuke Ueki
    Motoi Itani
    Hajime Nomura
    Shin-ich Nishi
    Yoshiroh Kaminoh
    Journal of Anesthesia, 2010, 24 : 908 - 912
  • [8] Evaluation of tracheal intubation in several positions by the Pentax-AWS Airway Scope: a manikin study
    Komasawa, Nobuyasu
    Ueki, Ryusuke
    Itani, Motoi
    Nomura, Hajime
    Nishi, Shin-ich
    Kaminoh, Yoshiroh
    JOURNAL OF ANESTHESIA, 2010, 24 (06) : 908 - 912
  • [9] Comparison of tracheal intubation by the Macintosh laryngoscope and Pentax-AWS (Airway Scope) during chest compression: a manikin study
    Komasawa, Nobuyasu
    Ueki, Ryusuke
    Nomura, Hajime
    Itani, Motoi
    Kaminoh, Yoshiroh
    JOURNAL OF ANESTHESIA, 2010, 24 (02) : 306 - 308
  • [10] Comparison of tracheal intubation by the Macintosh laryngoscope and Pentax-AWS (Airway Scope) during chest compression: a manikin study
    Nobuyasu Komasawa
    Ryusuke Ueki
    Hajime Nomura
    Motoi Itani
    Yoshiroh Kaminoh
    Journal of Anesthesia, 2010, 24 : 306 - 308