AMS 800 sphincter in urinary incontinence after prostatic surgery, based on a series of 16 patients.

被引:0
|
作者
Koutani, A [1 ]
Lechevallier, E [1 ]
Coulange, C [1 ]
机构
[1] Hop Salvator, Serv Urol & Transplantat Renale, F-13274 Marseille 9, France
关键词
prostatic surgery; urinary incontinence; artificial sphincter;
D O I
暂无
中图分类号
R5 [内科学]; R69 [泌尿科学(泌尿生殖系疾病)];
学科分类号
1002 ; 100201 ;
摘要
Objective: to present the results of a series of artificial sphincters implanted for urinary incontinence after prostatic surgery, Patients and methods: from 1987 to 1996, artificial sphincters were implanted for urinary incontinence after prostatic surgery, consisting of one transvesical prostatectomy, 11 prostatic resections and 4 radical prostatectomies. All these patients were evaluate by urine culture, cystourethrography and urethrocystoscopy. Eleven patients underwent urodynamic assessment. The AMS 800 artificial sphincter was used. Results: the mean interval between onset of incontinence and implantation was 18 months. The mean follow-up after implantation of the sphincter was 6 years. An operative wound of the bulbar urethra did not prevent implantation of the sphincter, but required repair and prolonged catheterization (12 days). The mean duration of catheterization was 6 days. The overall functional result was 81 % (13 case), and 75 % of sphincters were functional at 5 years. Five revisions were necessary to maintain functioning of the sphincter in 2 cases of rupture and 1 displacement of the balloon, 1 case of urethral atrophy and 1 rupture of the pump. Three explantations, 2 total and 1 partial, were performed. Two reimplantations were performed after 12 months and 26 months respectively. Conclusion: the AMS 800 artificial sphincter currently represents an effective treatment for urinary incontinence due to sphincter insufficiency after prostatic surgery. However, an old artificial sphincter may require revision to restore function.
引用
收藏
页码:382 / 385
页数:4
相关论文
共 50 条
  • [21] DEVICE SURVIVAL FOLLOWING PRIMARY IMPLANTATION OF THE AMS 800 ARTIFICIAL URINARY SPHINCTER FOR MALE STRESS URINARY INCONTINENCE
    Yafi, F.
    DeLay, K. J.
    Stewart, C.
    Chiang, J.
    Sangkum, P.
    Hellstrom, W. J. G.
    JOURNAL OF SEXUAL MEDICINE, 2017, 14 (01): : S52 - S52
  • [22] DEVICE SURVIVAL FOLLOWING PRIMARY IMPLANTATION OF THE AMS 800 ARTIFICIAL URINARY SPHINCTER FOR MALE STRESS URINARY INCONTINENCE
    Yafi, Faysal A.
    Stewart, Carrie
    Chiang, Jason
    Sangkum, Premsant
    Hellstrom, Wayne J. G.
    JOURNAL OF UROLOGY, 2016, 195 (04): : E1182 - E1182
  • [23] Stress urinary incontinence in female neurological patients: long-term functional outcomes after artificial urinary sphincter (AMS 800TM) implantation
    Phe, Veronique
    Leon, Priscilla
    Granger, Benjamin
    Denys, Pierre
    Bitker, Marc-Olivier
    Mozer, Pierre
    Chartier-Kastler, Emmanuel
    NEUROUROLOGY AND URODYNAMICS, 2017, 36 (03) : 764 - 769
  • [24] Robot-assisted AMS-800 Artificial Urinary Sphincter Bladder Neck Implantation in Female Patients with Stress Urinary Incontinence
    Peyronnet, Benoit
    Capon, Gregoire
    Belas, Olivier
    Manunta, Andrea
    Allenet, Clement
    Hascoet, Juliette
    Calves, Jehanne
    Belas, Michel
    Callerot, Pierre
    Robert, Gregoire
    Descazeaud, Aurelien
    Fournier, Georges
    EUROPEAN UROLOGY, 2019, 75 (01) : 169 - 175
  • [25] PERI-PROSTATIC ARTIFICIAL URINARY SPHINCTER AMS 800 IN ADULT NEUROGENIC PATIENTS: A MULTI-CENTRIC STUDY
    Ruffion, A.
    Genevois, S.
    Game, X.
    Saramon, J. P.
    Karsenty, G.
    Le Normand, L.
    Leriche, A.
    Denys, P.
    Richard, F.
    Chartier-Kastler, E.
    EUROPEAN UROLOGY SUPPLEMENTS, 2009, 8 (04) : 244 - 244
  • [26] Double cuff AMS 800 artificial urinary sphincter and sacral neuromodulation in complex situation of male incontinence
    Spinelli, M.
    Citeri, M.
    Zanollo, L.
    Scroppo, F.
    Malaguti, S.
    EUROPEAN UROLOGY SUPPLEMENTS, 2007, 6 (02) : 140 - 140
  • [27] The artificial urinary sphincter AMS 800 in the treatment of complex female urinary incontinence: indications, results, complications and risk factors
    Heitz, M
    Olianas, R
    Schreiter, F
    UROLOGE-AUSGABE A, 1997, 36 (05): : 426 - 431
  • [28] MALE URINARY INCONTINENCE AND ITS CORRECTION WITH DOUBLE CUFF AMS 800 URINARY SPHINCTER AND PENO-SCROTAL APPROACH
    Claudia, De Maria
    Luigi, Rolle
    Carlo, Ceruti
    Beatrice, Lillaz
    Augusto, Negro Carlo Luigi
    Mariateresa, Carchedi
    Dario, Fontana
    NEUROUROLOGY AND URODYNAMICS, 2011, 30 : 16 - 16
  • [29] Re: Robot-Assisted AMS-800 Artificial Urinary Sphincter Bladder Neck Implantation in Female Patients with Stress Urinary Incontinence
    Cadeddu, Jeffrey A.
    JOURNAL OF UROLOGY, 2019, 202 (03): : 441 - 442
  • [30] Urethral pressure profile assessment after artificial urinary sphincter implantation (FlowSecure™ and AMS-800™): A case series
    Haab, Alexander C.
    Tornic, Lure
    John, Hubert A.
    SAGE OPEN MEDICAL CASE REPORTS, 2019, 7