High-voltage energy dispersive x-ray spectrometry using a low-energy primary beam

被引:1
|
作者
Wu, Ying [1 ]
Klyachko, Dimitri [1 ]
Davilla, Scott [1 ]
Spallas, James [1 ]
Indermuehle, Scott [1 ]
Muray, Lawrence P. [1 ]
机构
[1] Keysight Technol, Agilent Technol, Santa Clara, CA 95051 USA
来源
关键词
SCANNING-ELECTRON-MICROSCOPY;
D O I
10.1116/1.4901883
中图分类号
TM [电工技术]; TN [电子技术、通信技术];
学科分类号
0808 ; 0809 ;
摘要
This paper proposes a way to do energy dispersive x-ray spectroscopy (EDS) at high accelerating voltages using a low-voltage field emission scanning electron microscope (FESEM). By biasing the sample to a positive voltage, the electron landing energy (also known as the x-ray excitation energy) is increased, effectively extending the EDS capabilities of the low-voltage SEM. This positive biasing reduces the signal-to-noise ratio of the FESEM image, but it also introduces scaling and offset errors. A one-time calibration procedure is shown to compensate for these distortions allowing the user to view the unbiased image, and then do a precise superposition of the low energy image, high energy image, and EDS elemental map. The proposed solution is implemented on a commercial low-voltage FESEM with an EDS energy resolution of 130 eV. (C) 2014 American Vacuum Society.
引用
收藏
页数:8
相关论文
共 50 条
  • [21] Development of low-energy X-ray spectrometry at the Laboratoire National Henri Becquerel
    Lepy, M. C.
    Plagnard, J.
    REVISTA MEXICANA DE FISICA, 2007, 53 (03) : 68 - 73
  • [22] Measurements of the K X-ray intensity ratios by using energy-dispersive X-ray fluorescence spectrometry
    Guerol, A.
    APPLIED RADIATION AND ISOTOPES, 2008, 66 (03) : 372 - 376
  • [23] Dynamic response of CVD monocrystalline diamond to low-energy x-ray beam
    Trucchi, D. M.
    Allegrini, P.
    Spadaro, S.
    Conte, G.
    INTEGRATED PHOTONICS: MATERIALS, DEVICES, AND APPLICATIONS, 2011, 8069
  • [24] Cryogenic microcalorimeters and tunnel junctions for high resolution energy dispersive x-ray spectrometry
    Höhne, J
    Altmann, M
    Angloher, G
    Bühler, M
    von Feilitzsch, F
    Frank, T
    Hettl, P
    Hertrich, T
    Jochum, J
    Nüssle, T
    Pfnür, S
    Schnagl, J
    Wänninger, S
    IN-LINE CHARACTERIZATION, YIELD RELIABILITY, AND FAILURE ANALYSES IN MICROELECTRONIC MANUFACTURING, 1999, 3743 : 162 - 168
  • [25] Cryogenic microcalorimeters for high-resolution energy-dispersive x-ray spectrometry
    Angloher, G
    Altmann, M
    Bühler, M
    von Feilitzsch, F
    Hertrich, T
    Hettl, P
    Höhne, J
    Huber, M
    Jochum, J
    Mössbauer, R
    Schnagl, J
    Wänninger, S
    EUV, X-RAY, AND GAMMA-RAY INSTRUMENTATION FOR ASTRONOMY X, 1999, 3765 : 630 - 635
  • [26] USE OF SENSITIVITY AS A FUNCTION OF FLUORESCENT X-RAY-ENERGY IN ENERGY-DISPERSIVE X-RAY SPECTROMETRY
    WEGSCHEIDER, W
    ELLIS, AT
    GOLDBACH, K
    LEYDEN, DE
    MAHAN, KI
    ANALYTICA CHIMICA ACTA, 1986, 188 : 59 - 66
  • [27] The transistor and energy-dispersive x-ray spectrometry: roots and milestones in x-ray analysis
    Ryon, RW
    X-RAY SPECTROMETRY, 2001, 30 (06) : 361 - 372
  • [28] ENERGY DISPERSIVE-X-RAY SPECTROMETRY
    YAKOWITZ, H
    JOURNAL DE MICROSCOPIE ET DE BIOLOGIE CELLULAIRE, 1975, 22 (2-3): : 137 - 150
  • [29] A METHOD FOR MEASURING THE HIGH-VOLTAGE OF A LOW-ENERGY VANDEGRAAFF ACCELERATOR
    ANDRADE, E
    ZIRONI, EP
    NUCLEAR INSTRUMENTS & METHODS IN PHYSICS RESEARCH SECTION A-ACCELERATORS SPECTROMETERS DETECTORS AND ASSOCIATED EQUIPMENT, 1988, 273 (01): : 16 - 20
  • [30] Energy-dispersive X-ray spectroscopy in a low energy electron microscope
    Tromp, Rudolf M.
    ULTRAMICROSCOPY, 2024, 259