Measurements of the K X-ray intensity ratios by using energy-dispersive X-ray fluorescence spectrometry

被引:7
|
作者
Guerol, A. [1 ]
机构
[1] Ataturk Univ, Dept Phys, Fac Art & Sci, TR-25240 Erzurum, Turkey
关键词
intensity ratio; EDXRF; radioisotope;
D O I
10.1016/j.apradiso.2007.09.017
中图分类号
O61 [无机化学];
学科分类号
070301 ; 081704 ;
摘要
K X-ray intensity ratios of some elements have been investigated by using radioisotope excited-energy-dispersive X-ray fluorescence spectrometer. The characteristic X-rays emitted from the samples have been recorded using a Si(Li) detector and they have been fitted to Gaussian functions by using a least-square method. The measured K X-ray intensity ratios have uncertainties of about 5%, and they are in good agreement with theoretical and experimental values available in the literature. (c) 2007 Elsevier Ltd. All rights reserved.
引用
收藏
页码:372 / 376
页数:5
相关论文