Electrochemical sensor based on phospholipid modified glassy carbon electrode - determination of paraquat

被引:28
|
作者
Tomkova, H. [1 ]
Sokolova, R. [2 ]
Opletal, T. [1 ]
Kucerova, P. [1 ]
Kucera, L. [1 ]
Souckova, J. [1 ]
Skopalova, J. [1 ]
Bartak, P. [1 ]
机构
[1] Palacky Univ, Reg Ctr Adv Technol & Mat, Dept Analyt Chem, Fac Sci, 17 Listopadu 12, Olomouc 77146, Czech Republic
[2] Czech Acad Sci, J Heyrovsky Inst Phys Chem, Dolejskova 2155-3, Prague 18223 8, Czech Republic
关键词
Asolectin; Phospholipid; Paraquat; Sensor; Square wave voltammetry; BILAYER-LIPID MEMBRANES; WAVE VOLTAMMETRIC DETERMINATION; PERFORMANCE LIQUID-CHROMATOGRAPHY; ATOMIC-FORCE MICROSCOPY; DOPED DIAMOND ELECTRODE; PASTE ELECTRODE; MASS-SPECTROMETRY; SENSITIVE DETERMINATION; STRIPPING VOLTAMMETRY; CYCLIC VOLTAMMETRY;
D O I
10.1016/j.jelechem.2017.12.048
中图分类号
O65 [分析化学];
学科分类号
070302 ; 081704 ;
摘要
Electrochemical sensor based on glassy carbon electrode modified with asolectin (mixture of soybean phospholipids) was developed. Properties of phospholipid layer, such as thickness and compactness, were investigated using atomic force microscopy and X-ray photoelectron spectroscopy. Phospholipids completely cover whole surface and the thickness of the layer 7.1 +/- 1.2 nm corresponds to the value of the phospholipid bilayer of cell membranes. Due to the accumulation of paraquat in the phospholipid layer, square wave voltammetry with asolectin modified electrode exhibits three times higher sensitivity, measured as a slope of calibration dependence, and better linearity in comparison to unmodified glassy carbon electrode. Limit of detection 2.2 nmol L-1 and recoveries higher than 88% were achieved in model samples of water. The method was used for determination of paraquat in real sample of river water and results were compared with high performance liquid chromatography and differential pulse voltammetry on hanging mercury drop electrode giving 0.071 +/- 0.010, 0.050 +/- 0.010 and 0.061 +/- 0.008 mu mol L-1, respectively.
引用
收藏
页码:33 / 39
页数:7
相关论文
共 50 条
  • [41] Electrochemical determination of lead using bismuth modified glassy carbon electrode
    Tufa, Lemma Teshome
    Siraj, Khalid
    Soreta, Tesfaye Refera
    RUSSIAN JOURNAL OF ELECTROCHEMISTRY, 2013, 49 (01) : 59 - 66
  • [42] CuO nanoflowers modified glassy carbon electrode for the electrochemical determination of methionine
    Ebrahimi, Maryam
    Beitollahi, Hadi
    EURASIAN CHEMICAL COMMUNICATIONS, 2021, 3 (01): : 19 - 25
  • [43] Polymer modified glassy carbon electrode for the electrochemical determination of caffeine in coffee
    Amare, Meareg
    Admassie, Shimelis
    TALANTA, 2012, 93 : 122 - 128
  • [44] Electrochemical Determination of Malathion on an Acetylcholinesterase-Modified Glassy Carbon Electrode
    Shamgsumova, Rezeda V.
    Shurpik, Dmitry N.
    Evtugyn, Vladimir G.
    Stoikov, Ivan I.
    Evtugyn, Gennady A.
    ANALYTICAL LETTERS, 2018, 51 (12) : 1911 - 1926
  • [45] Electrochemical Sensor for Simple and Sensitive Determination of Hydroquinone in Water Samples Using Modified Glassy Carbon Electrode
    Karami-Kolmoti, Parisa
    Beitollahi, Hadi
    Modiri, Sina
    BIOMEDICINES, 2023, 11 (07)
  • [46] ELECTROCHEMICAL SENSOR OF NITRITE BASED ON AN INORGANIC FILM-MODIFIED GLASSY-CARBON ELECTRODE
    GAO, ZQ
    ZI, MX
    ZHANG, Y
    WANG, GQ
    ZHAO, ZF
    MIKROCHIMICA ACTA, 1993, 111 (1-3) : 63 - 70
  • [47] An electrochemical sensor based on octahedral composite modified glassy carbon electrode for voltammetric detection of cocaine
    Oveili, Elham
    Jahani, Shohreh
    Tayari, Omid
    Jahanara, Abbas
    Fazli, Zohreh
    Aramesh-Boroujeni, Zahra
    Rashidi, Soroush
    Baravati, Nadia Rezaei Zade
    Foroughi, Mohammad Mehdi
    ELECTROANALYSIS, 2023, 35 (07)
  • [48] Electrochemical L-Tyrosine Sensor Based on a Glassy Carbon Electrode Modified with Exfoliated Graphene
    Varodi, Codruta
    Pogacean, Florina
    Coro, Maria
    Ciori, Alexandra
    Pruneanu, Stela
    SENSORS, 2022, 22 (10)
  • [49] Electrochemical bisphenol A sensor based on nanoporous PtFe alloy and graphene modified glassy carbon electrode
    Tian, Chunhuan
    Chen, Dandan
    Lu, Nali
    Li, Yao
    Cui, Rongjing
    Han, Zhida
    Zhang, Genhua
    JOURNAL OF ELECTROANALYTICAL CHEMISTRY, 2018, 830 : 27 - 33
  • [50] Electrochemical sensor based on a poly(para-aminobenzoic acid) film modified glassy carbon electrode for the determination of melamine in milk
    Liu, Ya Ting
    Deng, Jian
    Xiao, Xi Lin
    Ding, Li
    Yuan, Ya Li
    Li, Hui
    Li, Xiu Ting
    Yan, Xiao Na
    Wang, Li Li
    ELECTROCHIMICA ACTA, 2011, 56 (12) : 4595 - 4602